GE LOGIQ 200 Quality Assurance Manual page 151

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 0
2--3--2
System Diagnosis (Continued)
RESULT CODE
Passed
Checked
Skipped
No PIOP
212
213
321
322
323
325
326
327
328
330
331
341
342
343
344
345
350
351
352
353
354
355
356
DIAGNOSIS TEST DETAIL
Board
The DRAM test, at booting, is skipped.
ECG Module is absent.
BPHV
+THV of BPHV is failed.
--THV of BPHV is failed.
MVP
B Mode interpolation test is failed.
B Mode coordination test is failed.
B Mode ALM test is failed.
Image Capture test is failed.
Image Archive test is failed.
Image Recall test is failed.
Front End test is failed.
The first Cine memory test is failed.
The Second Cine memory test is failed.
M Mode Communication test is failed.
M Mode XY memory test is failed.
M Mode interpolation test is failed.
M Mode path test is failed.
M Cine test is failed.
Front End Line Memory test is failed.
Frame average memory test is failed.
B--Mode XY memory test is failed.
Image Capture Memory test is failed.
FPGA memory test is failed.
USC Bus generation test is failed.
DSP SRAM test is failed.
LOGIQ 200 PRO Series PROPRIETARY MANUAL
TABLE 2--2
Meaning
2--14
2242594
DIAGNOSTICS