GE LOGIQ 200 Quality Assurance Manual page 15

Table of Contents

Advertisement

g
2277615-100
GE Medical Systems
5/16
REV 0
GRAY SCALE TEST
Refer toChapter 2, Section 2-3-2 in the
LOGIQ a200
or
LOGIQ 200PRO
Proprietary manual for details on
entering the service software program, tests and control commands.
Select Monitor Test from the Service Utility Menu
The gray scale test is not intended to set up the camera, printer or peripherals to the test pattern. Ensure
that the monitor and photography are set up to the customer specifications.
Look at the gray bar test pattern on the left side of the viewing monitor.
Count the number of distinct gray bar steps on the viewing monitor. Then count the number of steps
visualized in the Gray bar on the hard copy image. Circle the number of steps, if any, that are missing on
the hard copy. Circle the number of steps, if any, that are missing on the hard copy.
Check the one box that applies:
Number of steps missing on the hard copy:
0
1
2
3
4+
Has the photography changed since the last QA Test?
Change
No Change (or First Test)
NOTE: Press <CTRL><X> to exit the Monitor Test
SECURE PAGE PRINT OF GRAY SCALE TEST HERE