GE MEDICAL SYSTEMS
REV 2
2–3–2
System Diagnosis (Continued)
TEST NAME: HV TEST
PURPOSE
To check HV BOARD ASSY.
CRITERIA
Refer to result shown on the monitor.
Action if the test fails
If RTC board assy tests passed, change HV board assy.
The following Assemblies are necessary for this test
RTC, DSC, MST, HV, CONN, Monitor
H V
SHV
THV
HV
HV
Ref
Status
HV test TEST
ILLUSTRATION 2–12
2–21
a
LOGIQ
200 PROPRIETARY MANUAL
2138854
DIAGNOSTICS