GE LOGIQ 200 Quality Assurance Manual page 142

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 2
2–3–7
Initialize (Continued)
3.
Input the number and press Set or Return key.
The Massage <Press 'Set' to confirm reset, or press 'Clear'> will appear on the bottom of monitor as shown in the
ILLUSTRATION 2–30.
Service Utility MENU :
Current Software Version is 2.0
System Diagnosis
System Reset
Self Operation History
Option Set
*Initialize :1 (1:America 2:Europe 3:Asia 4:Japan)
System Clock Adjust
Exit
Press 'Set' to confirm reset, or press 'Clear'
4.
Press Set to reload the factory default depending on the selected country/zone, or Press Clear to cancel the
initializations.
LOGIQ
Top Menu
96/03/12 09:16:37
INITIALIZE MENU
ILLUSTRATION 2–30
2–41
a
200 PROPRIETARY MANUAL
2138854
DIAGNOSTICS