GE LOGIQ 200 Quality Assurance Manual page 181

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 0
2--3--6
Initialize
Procedure
Follow these steps to initialize the data which is saved in the LOGIQ 200 PRO Series.
If this function is performed, the data which is saved in the LOGIQ 200 PRO Series will be initialized.
All preset parameters including customer presets and system presets will also be set to the factory
default.
1.
Using the Arrow up/down keys or Trackball, select the "Initialize" menu in the Top Menu. Refer to the
ILLUSTRATION 2--5.
2.
Press Set or Return key.
The Initialize Menu will appear on the monitor as shown in the ILLUSTRATION 2--26.
Service Utility MENU :
Current Software Version is 3.0
System Diagnosis
Board/Program Version
System Reset
Key Operation History
Option Set
*Initialize :1
System Clock Adjust
Software/Probe data upgrade
Exit
LOGIQ 200 PRO Series PROPRIETARY MANUAL
Note
Top Menu
(1:America 2:Europe 3:Asia 4:Japan)
INITIALIZE MENU
ILLUSTRATION 2--26
2--44
99/03/12 09:16:37
2242594
DIAGNOSTICS