GE MEDICAL SYSTEMS
REV 0
2--3--3
System Reset
Procedure
Follow these steps to reset the system.
1.
Using the Arrow up/down keys or Trackball, select the "System Reset" menu in the Top Menu. Refer to the
ILLUSTRATION 2--5.
2.
Press Set or Return key.
The Massage <Press ' Set'to confirm reset, or press ' Clear' > will appear on the bottom of monitor as shown in the
ILLUSTRATION 2--23.
Service Utility MENU :
Current Software Version is 3.0
System Diagnosis
Board/Program Version
*System Reset
Key Operation History
Option Set
Initialize :1
System Clock Adjust
Software/Probe data upgrade
Exit
3.
Press Set to the system reset, or Press Clear to skip the system reset..
LOGIQ 200 PRO Series PROPRIETARY MANUAL
Top Menu
(1:America 2:Europe 3:Asia 4:Japan)
SYSTEM RESET MENU
ILLUSTRATION 2--23
2--41
99/03/12 09:16:37
2242594