GE LOGIQ 200 Quality Assurance Manual page 138

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 2
2–3–3
System Reset
Procedure
Follow these steps to reset the system.
1.
Using the Arrow up/down keys or Trackball, select the "System Reset" menu in the Top Menu. Refer to the
ILLUSTRATION 2–5.
2.
Press Set or Return key.
The Massage <Press 'Set' to confirm reset, or press 'Clear'> will appear on the bottom of monitor as shown in the
ILLUSTRATION 2–26.
Service Utility MENU : Top Menu
Current Software Version is 2.0
System Diagnosis
*System Reset
Key Operation History
Option Set
Initialize :i (1:America 2:Europe 3:Asia 4:Japan)
System Clock Adjust
Exit
Press 'Set' to confirm reset, or press 'Clear'
3.
Press Set to the system reset, or Press Clear to skip the system reset..
LOGIQ
96/03/12 09:16:37
SYSTEM RESET MENU
ILLUSTRATION 2–26
2–37
a
200 PROPRIETARY MANUAL
2138854
DIAGNOSTICS