GE LOGIQ 200 Quality Assurance Manual page 143

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 2
2–3–8
System Clock Adjust
Procedure
Follow these steps to adjust the system clock of the LOGIQ
1.
Using the Arrow up/down keys, select the "System Clock Adjust" menu in the Top Menu. Refer to the
ILLUSTRATION 2–5.
2.
Press Set or Return key.
The System Clock Adjust Menu will appear on the monitor as shown in the ILLUSTRATION 2–31.
Service Utility MENU :
Current Software Version is 2.0
System Diagnosis
System Reset
Self Operation History
Option Set
Initialize :1 (I:America 2:Europe 3:Asia 4:Japan)
*System Clock Adjust: 98/3/27 15:10
Exit
3.
Input current date and time, and press Set or Return key.
a
200.
Top Menu
DATABASE INITIALIZE MENU
ILLUSTRATION 2–31
2–42
a
LOGIQ
200 PROPRIETARY MANUAL
96/03/12 09:16:37
2138854
DIAGNOSTICS