System Diagnosis (Continued)
TEST NAME: SYSTEM TEST
PURPOSE
To check all the system functions sequentially except Beam former test, ADC test, keyboard test,
Monitor test and ROM version.
CRITERIA
Refer to result shown on the monitor.
DISPLAY FORMAT OF RESULTS
Refer to result shown on the monitor as in ILLUSTRATION 2–9.
Service Utility MENU :
Current Software Version is 2.0
NAME
*System test
Beam Former test
HV test
ADC test
DSC Front/end test
B mode Function test
B mode X–Y Memory test
B mode Cine test
M mode Function test
MST test
System DRAM test
Keyboard test
PIOP test
Monitor test
Power Supply test
ROM Version
Exit
Action if the test fails
Refer to fail code listed in TABLE 2–2 and take actions.
The following Assemblies are necessary for this test
HV, PROBE, RTC, ESP, CONN, DSC, MST, Monitor
System Diagnosis
Time
Status
## M
Checked
0 M
Checked
## M
Passed
O M
Checked
## M
Running
## M
Passed
## M
Passed
## M
## M
## M
Fail : ###
## M
Skipped
0 M
Checked
## M
No PIOP
0 M
Checked
## M
0 M
SYSTEM TEST MENU–RESULT
ILLUSTRATION 2–9
2–18
a
LOGIQ
200 PROPRIETARY MANUAL
96/03/12 09:16:37
2138854
DIAGNOSTICS