System Diagnosis (Continued)
TEST NAME: SYSTEM DRAM TEST
PURPOSE
To check DRAM in MSTE assy.
CRITERIA
Refer to result shown on the monitor.
Action if the test fails
Change MSTE board assy.
The following Assemblies are necessary for this test
RBF, TBF, SCAN, MVP, MSTE, Monitor
MSTE
VRAM
Reset
ROM
RAM
AGDC
SYSTEM DRAM TEST
ILLUSTRATION 2--18
LOGIQ 200 PRO Series PROPRIETARY MANUAL
CPU
RTC
Super
I/O
IDE
CTL
From NOVA_BUS