GE LOGIQ 200 Quality Assurance Manual page 150

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 0
2--3--2
System Diagnosis (Continued)
TEST NAME
MSTE TEST
KEYBOARD TEST
PIOP TEST
MONITOR TEST
POWER SUPPLY TEST
TABLE 2--1
SYSTEM DIAGNOSIS TEST LIST
TEST DETAIL
Application ROM Check Sum
System DRAM Test
ROM Data Checksum Test
Probe Data Check Sum Test
500 Patient Data Test
Setup Data Test
386EX COM1 Port Test
Super I/O COM1 port Test
Super I/O COM2 port Test
Reserved flash memory test
Key History Data test
Real Time Clock test
386EX Timer test
Super I/O Basic Test
Super I/O Parallel Port Test
VRAM Test
AGDC Test
MOD Read/Write Test
Analog 5V
Analog --5V
Digital +5V
+12V for Monitor
+12V for ECG
--12V for ECG
2--13
LOGIQ 200 PRO Series PROPRIETARY MANUAL
TEST TYPE
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
MANUAL
AUTO
MANUAL
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
2242594
FAILURE CODE
430
438, Skipped
431
432
435
433
440
442
443
436
434
437
439
441
444
445
446
447
--
600~601
--
810
811
812
813
814
815
DIAGNOSTICS