GE LOGIQ 200 Quality Assurance Manual page 180

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 0
2--3--5
Option Set
Procedure
Follow these steps to set the option.
1.
Using the Arrow up/down keys or Trackball, select the "Option Set" menu in the Top Menu. Refer to the
ILLUSTRATION 2--5.
2.
Press Set or Return key.
The Option Set Menu will appear on the monitor as shown in the ILLUSTRATION 2--25.
Service UtilitY MENU : Option Set
Current Software Version is 3.0
CINE
ATO
MOD Driver
*Biopsy Guideline
MGOB Calculation
Cardiac Calculation :1
Urology Calculation :l
Serial Printer
Paralled Printer
Data Management Center:l (1:Not Install 2:Install)
InSite
Exit
The status of each option will be indicated.
3.
Using the Arrow up/down keys, select the option item which will should be changed and then Input the number.
4.
To activate the Option, exit the menu and reset the System.
LOGIQ 200 PRO Series PROPRIETARY MANUAL
:1
(1:Not Install 2:Install)
:1
(1:Not Install 2:Install)
:1
(1:Not Install 2:Install)
:1
(1:Not Install 2:Install)
:1
(1:Not Install 2:Install)
(1:Not Install 2:Install)
(1:Not Install 2:Install)
:l
(1:Not Install 2:Install)
:1
(1:Not Install 2:Install)
:1
(1:Not Install 2:Install)
OPTION SET MENU
ILLUSTRATION 2--25
2--43
99/03/12 09:16:37