GE LOGIQ 200 Quality Assurance Manual page 148

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 0
2--3--2
System Diagnosis (Continued)
Procedure 3
Follow these steps to enter each test in the System Diagnosis menu.
1.
Using the Arrow up/down keys, select test which should be performed. Refer to ILLUSTRATION 2--6.
2.
Press Set or Return keys.
Refer to each section to perform the each test for System Diagnosis.
Procedure 4
The result of test will appear on the status column in the System Diagnosis menu. Refer to the ILLUSTRATION 2--7.
The AUTO type tests have three status as belows.
Running
Pass
Fail : ###
Checked
Skipped
No PIOP
In case of CINE test, It shows the failure message when the option unit is not installed.
The MANUAL type tests have one status as belows.
Service Utility MENU :
Current Software Version is 3.0
NAME
*System test
Scan Test
TBF Test
Linear/Convex Path
Sector Probe Path
BPHV test
MVP Hardware Test
B mode Function
B mode Cine
M mode Function
MSTE test
Keyboard
PIOP
Monitor
Power Supply
Exit
Press 'ctrl'+'X'to
It shows that this test item is doing now.
It shows that this test item finished without failure.
It shows that this test item finished with failure. The "###" shows failure code. Refer to
TABLE 2--1 as detailed failure code.
It shows that this test item had been checked.
It shows that DRAM test had not been executed at booting.
It shows that the option PIOP unit is not installed.
SERVICE DIAGNOSIS MENU--RESULT
LOGIQ 200 PRO Series PROPRIETARY MANUAL
System Diagnosis
Time
Status
21 M
Passed
0 M
Checked
0 M
Checked
0 M
Checked
0 M
Checked
1 M
Checked
1 M
Running
1 M
Passed
8 M
2 M
4 M
Fail:###
0 M
Checked
1 M
0 M
Checked
1 M
ILLUSTRATION 2--7
2--11
99/03/12 09:16:37
2242594
DIAGNOSTICS