Table of Contents

Advertisement

Quick Links

Technical
Publication
Direction 5184024-100
Revision 5
GE Healthcare
LOGIQworks Basic Service Manual
©
Copyright
2002-2007
by General Electric Company Inc.
All Rights Reserved
GE Healthcare

Advertisement

Table of Contents
loading

Summary of Contents for GE LOGIQworks

  • Page 1 GE Healthcare Technical Publication Direction 5184024-100 Revision 5 GE Healthcare LOGIQworks Basic Service Manual © Copyright 2002-2007 by General Electric Company Inc. All Rights Reserved...
  • Page 3 • WARTEN SIE DIESES GERÄT NUR, WENN SIE DIE ENTSPRECHENDEN ANWEISUNGEN IM KUNDENDIENST-HANDBUCH GELESEN HABEN UND WARNUNG NACHVOLLZIEHEN KÖNNEN. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
  • Page 4 GE H EALTHCARE 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL • ESTE MANUAL DE SERVICIO SÓLO ESTÁ DISPONIBLE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, LA TRADUCCIÓN ES RESPONSABILIDAD DEL CLIENTE.
  • Page 5 ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝΔΥΝΟΥΣ. • EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ • HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT, AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE. • NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A FIGYELMEZTETÉS KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
  • Page 6 GE H EALTHCARE 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL • TENTO SERVISNÍ NÁVOD EXISTUJE POUZE V ANGLICKÉM JAZYCE. • V PŘÍPADĚ, ŽE POSKYTOVATEL SLUŽEB ZÁKAZNÍKŮM POTŘEBUJE NÁVOD V JINÉM JAZYCE, JE ZAJIŠTĚNÍ PŘEKLADU DO ODPOVÍDAJÍCÍHO JAZYKA ÚKOLEM ZÁKAZNÍKA.
  • Page 7 GE H EALTHCARE 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL • ŠIS EKSPLOATAVIMO VADOVAS YRA IŠLEISTAS TIK ANGLŲ KALBA. • JEI KLIENTO PASLAUGŲ TEIKĖJUI REIKIA VADOVO KITA KALBA – NE ANGLŲ, VERTIMU PASIRŪPINTI TURI KLIENTAS. • NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS DARBŲ, NEBENT ĮSPĖJIMAS...
  • Page 8 GE H EALTHCARE 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL • ДАННОЕ РУКОВОДСТВО ПО ОБСЛУЖИВАНИЮ ПРЕДОСТАВЛЯЕТСЯ ТОЛЬКО НА АНГЛИЙСКОМ ЯЗЫКЕ. • ЕСЛИ СЕРВИСНОМУ ПЕРСОНАЛУ КЛИЕНТА НЕОБХОДИМО РУКОВОДСТВО НЕ НА АНГЛИЙСКОМ ЯЗЫКЕ, КЛИЕНТУ СЛЕДУЕТ САМОСТОЯТЕЛЬНО ОБЕСПЕЧИТЬ ПЕРЕВОД. • ПЕРЕД ОБСЛУЖИВАНИЕМ ОБОРУДОВАНИЯ ОБЯЗАТЕЛЬНО ОБРАТИТЕСЬ...
  • Page 9 GE H EALTHCARE 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL...
  • Page 10 GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
  • Page 11 The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Healthcare. GE Healthcare may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
  • Page 12 GE H EALTHCARE 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL This page was intentionally left blank.
  • Page 13 ASIC ERVICE ANUAL Revision History Revision Date Reason for change 2006-07-25 Initial Version for LOGIQworks 3.0 2006-08-07 Released Version for LOGIQworks 3.0 2007-01-09 Added Dell Precision 390 and description about TripleMonitor 2007-03-21 Released Version for LOGIQworks 3.0.4 2007-09-04 Version for LOGIQworks 3.0.5...
  • Page 14 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL This page was intentionally left blank.
  • Page 15: Table Of Contents

    Typical Users of the Basic Service Manual ......1 - 2 LOGIQworks Models covered by this Manual ..... . . 1 - 2 How to Turn the LOGIQworks workstation on/off .
  • Page 16 Electrical Requirements ........2 - 7 LOGIQworks Power Requirements ......2 - 7 Inrush Current .
  • Page 17 Installing LOGIQworks at the Site ....... . 3 - 29...
  • Page 18 TripleMonitors for LOGIQworks FX and LOGIQworks SoftwareOnly 3 - 121 Connection Service Properties up to LOGIQworks 1.3 ....3 - 133 General ..........3 - 133 Module .
  • Page 19 Local ..........3 - 172 Print Services up to LOGIQworks 1.3 ......3 - 173 General .
  • Page 20 Checking iLinq Connection ....... . 4 - 20 Modem Check (For LOGIQworks 3.0 optional) ..... . 4 - 21 Power-On .
  • Page 21 Healthpage ............4 - 26 Healthpage for LOGIQworks up to Rev. 1.3 ......4 - 26 Healthpage for LOGIQworks up to Rev.
  • Page 22 Contact GE ........
  • Page 23 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL PM ............5 - 42...
  • Page 24 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL Diagnostics ............7 - 4 Global Service User Interface (GSUI) Diagnostic .
  • Page 25 General abbreviations ......... 9 - 2 LOGIQworks PC versions covered in this manual ....9 - 2 Renewal parts list .
  • Page 26 LOGIQworks units Outer casing ....... . . 10 - 6...
  • Page 27: Overview

    Safety Considerations 1-19 Customer Assistance 1-1-2 Purpose of Service Manual This Service Manual provides installation and service information for the LOGIQworks and contains the following chapters: Chapter 1 - Introduction: Contains a content summary and warnings. Chapter 2 - Pre-Installation: Contains any pre-installation requirements for the LOGIQworks.
  • Page 28: Typical Users Of The Basic Service Manual

    Total BTU/h H49021LA 1520 H49021LB 2015 H49021LC 1220 H49021LD 1415 Table 1-3 LOGIQworks Standard Hardware of System based on Dell Precision 360 (delivered since May 2004 until December 2004) PART LCD1 LCD2 Modem BTU/h Power (W) Total BTU/h H49021AA 1520...
  • Page 29: How To Turn The Logiqworks Workstation On/Off

    Section 3-5-13 on page 3-203. 1-1-5 Purpose of Operator Manual(s) The Operator Manual(s), Part Numbers available in chapter 9, should be fully read and understood before operating the LOGIQworks workstation and also kept near the unit for quick reference. Chapter 1 Introduction...
  • Page 30: Important Conventions

    Conventions Used in Book Model Designations This manual covers the LOGIQworks PC. (See Chapter 1 - LOGIQworks Models covered by this Manual on page 1-2.) Icons Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter.
  • Page 31: Standard Hazard Icons

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
  • Page 32: Product Icons

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Table 1-8 Product Icons...
  • Page 33: Safety Considerations

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL Table 1-8 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION "OFF/Standby" Indicates the power off/ standby position of the power switch. CAUTION ON/OFF switches This Power Switch DOES NOT ISOLATE Mains Supply "Mains ON"...
  • Page 34 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 1-3-1 Safety Hints (cont’d) Danger of Electrical Shock! WARNING WARNING Danger of Electrical Shock! A connection between the patient and a personal computer can result in risks for the patient from electric current.
  • Page 35 The user has to ensure that the LOGIQworks system is protected from viruses, e.g. by using antivirus programs, firewalls and regular updates for antivirus programs, firewalls and the operation system.
  • Page 36: Label Locations Of System Based On Dell Precision 340

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 1-3-2 Label Locations of System based on Dell Precision 340 Sony Labels only. Guarantee agreement Figure 1-2 Labeling 1.) CE-Label 2.) VDE-Label 3.) UL-Label 4.) Identification plate 5.) Homologation label (for Japan only)
  • Page 37 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 1-3-2 Label Locations of System based on Dell Precision 340 (cont’d) Power Input Information Tower Rear Tower Top Side (Seen from front) Front Serial Number...
  • Page 38 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 1-3-2 Label Locations of System based on Dell Precision 340 (cont’d) CRT Monitor, seen from the rear Serial Number Figure 1-4 CRT Monitor Labels LCD Monitor seen...
  • Page 39 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 1-3-2 Label Locations of System based on Dell Precision 340 (cont’d) Figure 1-6 Recovery CD Label with Rating Plate Label Modem seen from below Figure 1-7 Modem Labels...
  • Page 40: Label Locations Of System Based On Dell Precision 360 And Dell Precision 370

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 1-3-3 Label Locations of System based on Dell Precision 360 and Dell Precision 370 Power Input Information Tower Rear Serial Number Tower Top Side (Seen from front)
  • Page 41 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 1-3-3 Label Locations of System based on Dell Precision 360 and Dell Precision 370 (cont’d) Tower Front Figure 1-9 Additional Computer Labeling for Dell Precision 370 only...
  • Page 42 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 1-3-3 Label Locations of System based on Dell Precision 360 and Dell Precision 370 (cont’d) LCD Monitor seen from the rear Figure 1-10 LCD DELL 2001 Monitor Labels...
  • Page 43: Label Locations Of System Based On Dell Precision 380 And Dell Precision 390

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 1-3-4 Label Locations of System based on Dell Precision 380 and Dell Precision 390 Tower Rear Tower Bottom Side Tower Top Side Figure 1-12 Computer Tower Labeling...
  • Page 44: Label Locations

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 1-3-5 Label Locations Figure 1-14 Recovery CD Label with Rating Plate Label Modem seen from below Figure 1-15 Modem Label 1-18 Section 1-3 - Safety Considerations...
  • Page 45: Customer Assistance

    1-800-682-5327 1-262-524-5698 Customer Answer Center Fax: +1-414-647-4125 Europe Tel: +33 130831300 (General Imaging & Cardiac) GE Ultraschall Deutschland GmbH & Co. KG Beethovenstraße 239 Postfach 11 05 50, D-42665 Solingen Fax: +49-212 2802 431 Germany Asia Singapore/Japan) Asia Support Center - Singapore...
  • Page 46: System Manufacturer

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 1-4-2 System Manufacturer Table 1-10 Phone Numbers for Customer Assistance Manufacturer Phone Number ViewPoint Bildverarbeitung GmbH Argelsrieder Feld 12 (+49) 8153 9311 0 D-82234 Wessling...
  • Page 47: Pre-Installation

    2-1-1 Purpose of Chapter 2 This chapter provides the information required to plan and prepare for the installation of a LOGIQworks. Included are descriptions of the facility and electrical needs to be met by the purchaser of the unit. A checklist is also provided at the end of this section to help determine whether the proper planning and preparation is accomplished before the actual equipment installation is scheduled.
  • Page 48: General Console Requirements

    2-2-1-1 Cooling The cooling requirement for the LOGIQworks is within the range 1220-2100 BTU/hr. This is depending on whether a CRT Monitor or LCD Monitor is used, and on the hardware options installed. For detailed BTU/hr. see table 2-7 and table 2-8. The BTU/hr. value relates to the maximum operational Power consumption of the LOGIQworks Workstation.
  • Page 49: Lighting

    5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 2-2-1-1 Cooling (cont’d) Table 2-5 LOGIQworks Standard Hardware of System based on Dell Precision 340 PART CRT1 CRT2 LCD1 LCD2 Modem BTU/HR Power (W) Total BTU/HR H49021LA 1520 H49021LB 2015...
  • Page 50: Cooling

    2-2-2-1 Cooling The cooling requirement for the LOGIQworks is within the range 1220-2100 BTU/hr. This is depended on which hardware options that have been installed. For detailed BTU/hr. see table 2-7 and table 2-8. The BTU/hr. value relates to the maximum operational Power consumption of the LOGIQworks Workstation.
  • Page 51 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 2-2-2-1 Cooling (cont’d) Table 2-9 LOGIQworks Standard Hardware of DELL Precision 360 and DELL Precision 370 PART LCD1 LCD2 Modem BTU/h Power (W) Total BTU/h H49021AC 1220 H49021AD 1415...
  • Page 52: Lighting

    IRECTION EVISION WORKS ASIC ERVICE ANUAL 2-2-2-1 Cooling (cont’d) Table 2-11 LOGIQworks Optional Hardware PART Modem BTU/h Power (W) Total BTU/h 2-2-2-2 Lighting Bright light is needed for system installation, updates and repairs. However, operator and patient comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting system (dim/bright) is recommended.
  • Page 53: Electrical Requirements

    It is recommended that the branch circuit breaker for the machine be readily accessible. 2-2-4-4 Site Power Outlets Three to five AC power outlet must be less than 1.2 meters (3.9feet) from the LOGIQworks Workstation. Electrical installation must meet all current local, state, and national electrical codes. Chapter 2 Pre-Installation...
  • Page 54: Unit Power Plug

    2-2-4-5 Unit Power Plug If the unit arrives without a power plug, or with the wrong plug, you must contact GE Service; they must supply what is locally required. Section 2-2 - General Console Requirements...
  • Page 55: Emi Limitations

    Computers are susceptible to Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transients in the air or wiring. They also generate EMI. The LOGIQworks complies with limits as stated on the EMC label. However there is no guarantee that interference will not occur in a particular installation.
  • Page 56 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL Table 2-16 EMI Prevention/Abatement EMI Rule Details Take care with cellular Cellular phones may transmit a 5 V/m signal; that could cause image artifacts. phones 2-10 Section 2-2 - General Console Requirements...
  • Page 57: Facility Needs

    • Power outlets for test equipment within 1 m (3.3 ft.) of PC. • Either an Analog Phone Line, or a GE Approved Broadband/VPN connection, is required for iLinq. A functional LAN is required for Network connectivity. • An Ethernet network wall jack is required.
  • Page 58 Hot & Cold water A dedicated Ethernet network wall outlet 1.5m 243,84 cm A dedicated Power outlet A dedicated Analog Telephone jack 105cm 609.6 cm Figure 2-2 Suggested Maximum Room with LOGIQworks and Ultrasound Scanner 2-12 Section 2-3 - Facility Needs...
  • Page 59: Patient Environment Iec 60601-1-1

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 2-3-3-1 Patient environment IEC 60601-1-1 An area in which patients are normally cared for, the patient vicinity is the space with surfaces likely to be contacted by the patient or attendant who can touch the patient.
  • Page 60: Networking Pre-Installation Requirements

    Stand Alone Network or Integrated into Hospital’s Network If LOGIQworks is to be used in a separate network together with a LOGIQ 9 or LOGIQ 7 and a network printer with no connection to the Hospital’s network or LOGIQworks is to be on the Hospital network either the predefined factory settings should be used.
  • Page 61: Insite 2

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 2-3-4-4 Insite 2 (cont’d) Figure 2-4 Data Collection Document Page 1 Chapter 2 Pre-Installation 2-15...
  • Page 62 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 2-3-4-4 Insite 2 (cont’d) Figure 2-5 Data Collection Document Page 2 2-16 Section 2-3 - Facility Needs...
  • Page 63 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 2-3-4-4 Insite 2 (cont’d) Figure 2-6 Data Collection Document Page 3 Chapter 2 Pre-Installation 2-17...
  • Page 64 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 2-3-4-4 Insite 2 (cont’d) Figure 2-7 Data Collection Document Page 4 2-18 Section 2-3 - Facility Needs...
  • Page 65 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 2-3-4-4 Insite 2 (cont’d) Figure 2-8 Data Collection Document Page 5 Chapter 2 Pre-Installation 2-19...
  • Page 66 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL Your Notes: 2-20 Section 2-3 - Facility Needs...
  • Page 67: Installation

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL Chapter 3 Installation Section 3-1 Overview 3-1-1 The Purpose of Chapter 3 This chapter contains information NEEDED to install the UNIT, a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim.
  • Page 68: Safety Reminders

    WARNING WARNING three to two prong adapter. This defeats safety grounding. OPERATOR MANUAL(S) WARNING WARNING The User Manual(s) should be fully read and understood before operating the LOGIQworks and kept near the unit for quick reference. Section 3-1 - Overview...
  • Page 69: Receiving And Unpacking The Equipment

    Section 3-2 Receiving and Unpacking the Equipment 3-2-1 Shipment Description From the source factory all LOGIQworks parts are packed in the original packaging and put into a card- board box. 3-2-1-1 Check of Delivered Box 1.) Check that cardboard box is undamaged as it was shipped.
  • Page 70 Transport only with forklift or stracker truck. During transport pay attention to the point of gravity! Have two people available to unpack the LOGIQworks. Attempts to move the unit considerable distances or on an incline by one person could result in injury or damage or both.
  • Page 71 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-2-1-1Check of Delivered Box (cont’d) Figure 3-2 The LOGIQworks H49021LC delivery Table 3-3 The LOGIQworks H49021LC parts Name Dell Precision 340 Monitor NEC LCD 1880 Sx Modem Chapter 3 Installation...
  • Page 72 EVISION WORKS ASIC ERVICE ANUAL 3-2-1-1Check of Delivered Box (cont’d) Figure 3-3 The LOGIQworks H49021LD delivery Table 3-4 The LOGIQworks H49021LD parts Name Dell Precision 340 Monitor NEC LCD 1880 Sx Modem Section 3-2 - Receiving and Unpacking the Equipment...
  • Page 73 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-2-1-1Check of Delivered Box (cont’d) Figure 3-4 The LOGIQworks H49021LA delivery Table 3-5 The LOGIQworks H49021LA parts Item Name Dell Precision 340 Monitor Sony GMD-F520 Modem Chapter 3 Installation...
  • Page 74 LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-2-1-1Check of Delivered Box (cont’d) Figure 3-5 The LOGIQworks H49021LB delivery Table 3-6 The LOGIQworks H49021LB parts Name Dell Precision 340 Monitor Sony GMD-F520 Modem Section 3-2 - Receiving and Unpacking the Equipment...
  • Page 75 EVISION WORKS ASIC ERVICE ANUAL 3-2-1-1Check of Delivered Box (cont’d) Figure 3-6 The LOGIQworks H49021AC/H49021AA or H49021AK/H49021AM delivery. Table 3-7 The LOGIQworks 1FP H49021AA/H49021AC parts Name Dell Precision 360 Dell Precision 370 Monitor Dell LCD 1901 Europe Monitor Dell LCD 2001 US/ASIA...
  • Page 76 EVISION WORKS ASIC ERVICE ANUAL 3-2-1-1Check of Delivered Box (cont’d) Figure 3-7 The LOGIQworks H49021AD/H49021AB or H49021AL/H49021AN delivery Table 3-9 The LOGIQworks H49021AD/H49021AB parts Name Dell Precision 360 Dell Precision 370 Monitor Dell LCD 1901 Europe Monitor Dell LCD 2001 US/Asia...
  • Page 77: Contents Of Cpu Box

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-2-1-2 Contents of CPU Box 3-2-1-2-1 Contents of CPU Box of System based on Dell Precision 340 Figure 3-8 Contents of CPU Box Table 3-11...
  • Page 78 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-2-1-2-2 Contents of CPU Box of System based on Dell Precision 360 or on Dell Precision 370 Figure 3-9 Contents of CPU Box of System based on Dell Precision 360 or on Dell Precision 370...
  • Page 79 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-2-1-2-3 Contents of CPU Box of System based on Dell Precision 380 and Dell Precision 390 Figure 3-10 Contents of CPU Box of System based on Dell Precision 380 and Dell Precision 390...
  • Page 80: Additional Parts In Cpu Box

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-2-1-3 Additional Parts in CPU Box 3-2-1-3-1 Additional Parts in CPU Box of System based on Dell Precision 340 Figure 3-11 Additional Parts in CPU Box Table 3-14...
  • Page 81 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-2-1-3-2 Additional Parts in CPU Box of System based on Dell Precision 360, Dell Precision 370, Dell Precision 380 and Dell Precision 390 Figure 3-12 Additional Parts in CPU Box of System based on Dell Precision 360, Dell Precision 370, Dell...
  • Page 82 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL Table 3-15 CPU Box parts list Item Name Qty. Optical USB Mouse Software Option Dongle USB Harddisk Instant CD/DVD Powercord FP2001US for US only Y-Monitor cable 3-16...
  • Page 83: Lcd Monitor

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-2-1-4 LCD Monitor 3-2-1-4-1 LCD Monitor of System based on Dell Precision 340 Figure 3-13 Parts in the LCD delivery of System based on Dell Precision 340...
  • Page 84 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-2-1-4-2 LCD Monitor of System based on Dell Precision 360, Dell Precision 370, Dell Precision 380 or Dell Precision 390 Figure 3-14 Parts in the LCD delivery of System based on Dell Precision 360, Dell Precision 370, Dell Precision...
  • Page 85: Sony Crt Monitor

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-2-1-5 Sony CRT Monitor Figure 3-15 Parts in the CRT delivery Table 3-18 Contents of CRT Monitor Box Item Name Monitor CRT Signal Cable analog...
  • Page 86: Modem Kit Parts Delivery Inspection

    Table 3-19 Modem Kit 066E0796 Parts Table Name Part Number Qty. Read Me First Document Part of GE Part Number 2245794 Modem GE Part Number 2245794 Telephone cable Part of GE Part Number 2245794 Telephone Wire Part of GE PartNumber 2245794...
  • Page 87 Figure 3-17 Modem Parts overview. Table 3-20 Modem Parts Table Name Part Number Qty. Installation CD Part of GE Part Number 2413338 Modem GE Part Number 2413338 Telephone cable Part of GE Part Number 2413338 Quick Start Guide Part of GE Part Number 2413338...
  • Page 88: Preparing For Installation

    ANUAL Section 3-3 Preparing for Installation 3-3-1 Physical Inspection 3-3-1-1 LOGIQworks Voltage Settings • Verify that the LOGIQworks has the correct voltage delivery. Switch 110/220V Figure 3-18 Correct Voltage Settings Table 3-21 International Voltage and Frequencies settings Country Argentina Australia...
  • Page 89 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL Table 3-21 International Voltage and Frequencies settings Country France Germany Great Britain Greece Hong Kong Hungary Iceland India Israel Italy Japan 50/60 Korea, South Luxembourg...
  • Page 90: Power Cables

    Power Cord German 2369734 Power Cord Italian 2369735 Power Cord Spanish 2369736 WARNING Connecting The LOGIQworks to a wrong voltage level destroys the LOGIQworks. WARNING Table 3-23 Power Cable for DELL Monitor 2001 Name Part Number Power Cord US 2409670...
  • Page 91: Completing The Installation

    Section 3-4 Completing the Installation 3-4-1 System Specifications 3-4-1-1 Physical Dimensions of LOGIQworks The physical dimensions of the LOGIQworks main parts are summarized in Table 3-25. Table 3-27 lists the weight of LOGIQworks, with monitor and without on-board peripherals. Table 3-24...
  • Page 92: Required Space For Logiqworks Installation

    5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-1-2 Required Space for LOGIQworks Installation min. 100 70 cm CRT 1 min. 80cm recommended 100 cm Keyboard Mouse Figure 3-19 Required Space for System with 1 CRT Monitor min. 140cm...
  • Page 93 LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-1-2 Required Space for LOGIQworks Installation (cont’d) min. 80 60 cm LCD 1 min. 80cm recommended 100 cm Keyboard Mouse Figure 3-21 Required Space for System with 1 LCD Monitor min. 120 cm...
  • Page 94: Weight

    3-4-1-4 Acoustic Noise Output max. 57dB(A) 3-4-2 Electrical Specifications To see the Electrical Specifications for the LOGIQworks see Table 2-14 on page 2-7. Power Consumption nominal 550 VA including all options. WARNING Modification of voltage setting only by an authorized service person! Output power: 550VA per mains WARNING outlet, max.
  • Page 95: Installing Logiqworks At The Site

    EVISION WORKS ASIC ERVICE ANUAL 3-4-3 Installing LOGIQworks at the Site 3-4-3-1 Place the Parts ergonomically correct. Tower PC mounted on the floor Figure 3-23 Work Station w/ Monitor, Keyboard, PC, Mouse 1.) Place the Monitor, Keyboard, Mouse, Peripherals, Desk and stool ergonomically correct and place the Tower Computer either on the floor below the desk or on the desk top.
  • Page 96: Single Lcd Monitor Installation Of System Based On Dell Precision

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-2 Single LCD Monitor Installation of System based on Dell Precision 340 Items needed: Figure 3-24 Items needed for Single LCD Monitor Installation of System based on Dell Precision 340...
  • Page 97: Dell Precision 370

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-2 Single LCD Monitor Installation of System based on Dell Precision 340 (cont’d) Table 3-28 Items needed for Single LCD Monitor Installation Item Name Dell Precision 340...
  • Page 98: Precision 390

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-3 Single LCD Monitor Installation of System based on Dell Precision 360 or on Dell Precision 370 Items needed: Figure 3-25 Items needed for Single LCD Monitor Installation of System based on Dell Precision 360 or on Dell...
  • Page 99: Dell Precision 370

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-3 Single LCD Monitor Installation of System based on Dell Precision 360 or on Dell Precision 370 (cont’d) Table 3-29 Items needed for Single LCD Monitor Installation...
  • Page 100 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-4 Single LCD Monitor Installation of System based on Dell Precision 380 and Dell Precision 390 Items needed: Figure 3-26 Items needed for Single LCD Monitor Installation of System based on Dell Precision 380 and Dell...
  • Page 101 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-4 Single LCD Monitor Installation of System based on Dell Precision 380 and Dell Precision 390 (cont’d) Table 3-30 Items needed for Single LCD Monitor Installation...
  • Page 102 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-5 Double LCD Monitor Installation of System based on Dell Precision 340 Figure 3-27 Items needed for Double LCD Monitor Installation of System based on Dell Precision 340...
  • Page 103 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-5 Double LCD Monitor Installation of System based on Dell Precision 340 (cont’d) Table 3-31 Double LCD Monitor Installation Item Name Dell Precision 340 Monitor NEC LCD 1880 Sx...
  • Page 104 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-6 Double LCD Monitor Installation of System based on Dell Precision 360 or on Dell Precision 370 Figure 3-28 Items needed for Double LCD Monitor Installation of System based on Dell Precision 360 or Dell...
  • Page 105 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-6 Double LCD Monitor Installation of System based on Dell Precision 360 or on Dell Precision 370 (cont’d) Table 3-32 Items needed for Double LCD Monitor Installation...
  • Page 106 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL Table 3-33 Double LCD Monitor Installation of System based on Dell Precision 380 and Dell Precision 390 Figure 3-29 Items needed for Double LCD Monitor Installation of System based on Dell Precision 380 and Dell...
  • Page 107 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-6 Double LCD Monitor Installation of System based on Dell Precision 360 or on Dell Precision 370 (cont’d) Table 3-34 Items needed for Double LCD Monitor Installation...
  • Page 108: Single Crt Monitor Installation

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-7 Single CRT Monitor Installation Figure 3-30 Items for Single CRT Monitor Installation 3-42 Section 3-4 - Completing the Installation...
  • Page 109 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-7 Single CRT Monitor Installation (cont’d) Table 3-35 Items for Single CRT Monitor Installation Item Name Dell Precision 380/Dell Precision 390 Monitor Sony GDM F520...
  • Page 110: Double Crt Monitor Installation

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-8 Double CRT Monitor Installation Items needed: Figure 3-31 Items needed for Double CRT Monitor Installation 3-44 Section 3-4 - Completing the Installation...
  • Page 111 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-8 Double CRT Monitor Installation (cont’d) Table 3-36 Items for Double CRT Monitor Installation Item Name Part Number Dell Precision 340 2365947 Monitor Sony GDM F520...
  • Page 112: Connections At Rear Of Pc Of System Based On Dell Precision 340

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-9 Connections at rear of PC of System based on Dell Precision 340 Computer Power Cable Connection Dongle Modem in 1 Keyboard Connection Network Mouse USB HDD...
  • Page 113: Mouse And Usb Harddisk Connection Of Dell Precision 360

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-10 Mouse and USB Harddisk Connection of Dell Precision 360 1.) Connect the Mouse to one of the USB Connectors. 2.) Connect the USB Harddisk to the other USB Connector.
  • Page 114: Keyboard And Network Connection Of Dell Precision 360

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-11 Keyboard and Network Connection of Dell Precision 360 (cont’d) Figure 3-34 Keyboard and PC Rear. 3-48 Section 3-4 - Completing the Installation...
  • Page 115: Connections At Rear Of Pc Of System Based On Dell Precision 370

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-12 Connections at Rear of PC of System based on Dell Precision 370 Figure 3-35 Connections at rear of PC of System based on Dell Precision 370...
  • Page 116 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-13 Monitor Connections of System based on Dell Precision 340 Rear of LCD Monitor Figure 3-36 LCD Monitor Power Connection 1.) Connect the Monitor Power Cable to the rear of the Monitor.
  • Page 117: Monitor Connections Of System Based On Dell Precision 340

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL .3-4-3-13 Monitor Connections of System based on Dell Precision 340 (cont’d) LCD Rear Figure 3-37 PC and LCD Rear Chapter 3 Installation 3-51...
  • Page 118: Monitor Connection

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-14 Monitor Connection 1.) Connect the Monitor Power Cable to the rear of the Monitor. 2.) Connect the other end to a Wall Power Outlet. Figure 3-38 CRT Monitor power connection 3.) Connect the Y-Monitor Signal Cable to the display Card Socket at the rear of the PC.
  • Page 119 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-15 Monitor Connections of System based on Dell Precision 360 1.) Connect the Monitor Power Cable to the rear of the Monitor. 2.) Connect the other end to a Wall Power Outlet.
  • Page 120: Monitor Connections Of System Based On Dell Precision 360

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-15 Monitor Connections of System based on Dell Precision 360 (cont’d) 3.) Connect the Y-Monitor Signal Cable to the display Card Socket at the rear of the PC.
  • Page 121 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-16 Monitor Connections of System based on Dell Precision 370 1.) Connect the Monitor Power Cable to the rear of the Monitor. 2.) Connect the other end to a Wall Power Outlet.
  • Page 122: Monitor Connections Of System Based On Dell Precision 370

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-16 Monitor Connections of System based on Dell Precision 370 (cont’d) 6.) In case of dual Monitor installation connect the second Monitor to the Y-cable tagged with 2.
  • Page 123: Pc Connections

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-17 PC Connections 1.) Connect the female end of the PC Power Cable to the Power Socket on the PC and the other end to a wall Power Outlet.
  • Page 124: Global Modem Connections

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-18 Global Modem Connections Modem Power Cable, male connector Power Phone line wall jack Cable, female connector RS232 Power Cable Cable 076D2610 Telephone Wire AC/DC Convertor To PC Com 13 Figure 3-45 Global Modem Signal and Power connections for Dell Precision 340 1.) To do Modem RS232 Connection, plug one end of the RS232 cable into the RS232 connector on...
  • Page 125 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-3-18 Global Modem Connections (cont’d) Figure 3-46 Global Modem Signal Connection for Dell Precision 360, Dell Precision 370, Dell Precision 380 or Dell Precision 390...
  • Page 126: Power On / Boot Up

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-4 Power On / Boot Up Table 3-37 Bootup Step Check Expected Result Turn ON the Power switch Drag down the menu Administrator/Admin: On the appearing dialog type: - lwuser...
  • Page 127: Third Party Products

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-5 Third Party Products The third party product Nero Burning ROM needs to be installed from the vendor setup CDs. Please use the default installation settings. They are no longer part of the default system configuration.
  • Page 128: Initialize Service Platform For Logiqworks Up To Version 2.0

    3-4-6 Initialize Service Platform for LOGIQworks up to Version 2.0 The Checkout process initiates a communication between the customer and GE. Before you begin with the checkout you need to have an IP address to connect to the GE network.
  • Page 129: Modem Configuration

    EVISION WORKS ASIC ERVICE ANUAL 3-4-6 Initialize Service Platform for LOGIQworks up to Version 2.0 (cont’d) Figure 3-47 Configuring the “Device connection” tab 3-4-6-1 Modem Configuration 6.) Select “Modem” as device connection type. 7.) Select one of the following items corresponding to your telephone connection:...
  • Page 130 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-6-1 Modem Configuration (cont’d) Figure 3-48 The “Location Information” window 11.)The “Phone and Modem Options” window pops up. Click on “Remove” and then “Add”. 12.)Follow the instructions of the “Add/Remove Hardware Wizard”.
  • Page 131: Vpn Configuration

    17.)Click “Checkout now” and reboot the system. The PC is ready for OLC checkout now. 3-4-6-2 VPN Configuration 1.) Open the “InSite Interactive Platform Configuration”. See Initialize Service Platform for LOGIQworks up to Version 2.0 on page 3-62. Chapter 3 Installation 3-65...
  • Page 132 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-4-6-2 VPN Configuration (cont’d) Figure 3-52 The InSite Interactive Platform Configuration” window 2.) Select “network” on the “Device Connection Type” field. 3.) Enter the local gateway IP address.
  • Page 133: Initialize Service Platform For Logiqworks 3.0

    ERVICE ANUAL 3-4-7 Initialize Service Platform for LOGIQworks 3.0 For Installation Inside II see Technical Publication, Installation Manual: InSite II ViewPoint/LOGIQworks (Direction 5181936-100). 3-4-8 Filling out and sending the Product Locator Card To finish the installation procedure the product locator card has to be filled out and send to the given address.
  • Page 134: Configuration

    ANUAL Section 3-5 Configuration This chapter provides you with information on how to configure your LOGIQworks at an operating system and application level. When you receive your LOGIQworks system all these configuration settings are already pre configured according to the information you provided with the Data Collection Document (please refer to Section "Recommendation: A permanent broadband to the internet.
  • Page 135: 5184024-100, Revision

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-1-1 Set System Time (cont’d) Figure 3-56 Windows Time Zone-Table 4.) Click on OK. 5.) Select the Date & Time tab and set the time and date appropriately.
  • Page 136: Select Units

    Select here U.S. or Metric Figure 3-58 Windows Control Panel If you wish to overrule Windows’ Regional Settings you can select either Metric or U.S.from the drop down list in the LOGIQworks Generic Properties page (see Section 3-5-2-5 "System Properties" 3-70...
  • Page 137: Monitor Setup

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-1-3 Monitor Setup 1.) Select Control Panel from the Windows Start Menu. 2.) Select Display, the Display Properties page opens. 3.) Select the Settings tab and click Advanced...
  • Page 138 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-1-3-1 Monitor Setup based on Dell 340 Second monitor not Monitor connected connected Figure 3-60 ATI Radeon Display Properties - Dual Monitor Set up your system for dual monitor use: 4.) On the ATI Displays Tab make sure that both monitors are checked and activated (green light...
  • Page 139 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-1-3 Monitor Setup (cont’d) 4: Green light = active 6: Monitor Settings 5: Radio Button Figure 3-61 ATI Radeon Display Properties - Dual Monitor 5.) Select the radio button for the left monitor. For the right monitor select the to create an extended desktop to the right.
  • Page 140 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-1-3-2 Set up your system for true dual graphic card use (based on Dell 360) Figure 3-62 Dual Monitor Configuration for True Dual Graphic Card Port two...
  • Page 141 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-1-3 Monitor Setup (cont’d) Figure 3-64 Dual Monitor connection for Dell Precision 370 Chapter 3 Installation 3-75...
  • Page 142: Configure Tcp/Ip Settings

    3-5-1-4 Configure TCP/IP Settings Connect your LOGIQworks system to the network. To achieve communication between Scanner and LOGIQworks the following steps must be completed successfully. Configure the TCP/IP settings using the Windows Control Panel: If not already configured according to the information you provided with the Data Collection Document (see chapter ) or in case of configuration changes enter the correct data in the fields described in the figure below.
  • Page 143: Keyboard Setup

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-1-5 Keyboard setup This Procedure is not required if the keyboard is a US keyboard, since the default setting is set to US, NOTE English keyboards.
  • Page 144 3.) Use the scroll bar to find the correct language keyboard layout for your keyboard. 4.) Click on the Set as Default button. 5.) Click on Apply button. 6.) Click on OK. 7.) Reboot LOGIQworks to get the full effect of the Keyboard Layout change. 3-78 Section 3-5 - Configuration...
  • Page 145: General Application Level Configuration For Logiqworks Up To Rev. 1.3

    WORKS ASIC ERVICE ANUAL 3-5-2 General Application Level Configuration for LOGIQworks up to Rev. 1.3 3-5-2-1 TruAccess Plugin Definition Verify that the TruAccess Raw Data Processing plugin is already defined and activate the plugin if not yet done. 1.) Select Configuration -> Plugin Definitions... from the Data Selector menu.
  • Page 146 YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-2-1 TruAccess Plugin Definition (cont’d) Resulting list of installed plugins after TruAccess Raw Data Processing has been added: Figure 3-69 LOGIQworks- Plugin Definition Properties (2) 3-80 Section 3-5 - Configuration...
  • Page 147: General Settings

    Work Directory: Directory for temporary files created by RA 600. (Default: D:\AMI_61\WORK) • • Main Window: Initial View: Check from the Initial View drop-down list the Worklist View you want LOGIQworks to display by default when starting. (Default: nothing selected, Local Studies will be default then) Chapter 3 Installation 3-81...
  • Page 148: Logging

    Compare Patients on: Indicate how you want to view patient studies in a patient worklist: by name, ID or both. LOGIQworks will collapse all the studies into one folder for viewing on the Local Patient Worklist, provided that the studies have the same patient name and/or patient ID.(Default: Both name and ID)
  • Page 149: Popup Menu

    Always Flush File after Write: If checked, all logging actions will be written immediately to the log • file. This option will have a serious impact on the performance of LOGIQworks, so it should not normally be checked. (Default: not checked) Log Pipe: Name of the pipe where logging information is written.
  • Page 150: System Properties

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-2-4 Popup Menu (cont’d) Figure 3-72 General Settings - Popup Menu Check the boxes next to the menu items to select the ones you want to appear on the pop-up menu that appears when right-clicking in the Data Selector.
  • Page 151 3-70). (Default: METRIC) If you wish to overrule Windows' regional settings, you can select either METRIC (Default setting) or US from the drop-down list. LOGIQworks will then use millimeters or inches irrespective of what is specified in the Windows Control Panel.
  • Page 152 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-2-5 System Properties (cont’d) Figure 3-74 Generic Properties - System Settings (2) Person name representation: • Here you can indicate how you want patient names to appear on work lists.
  • Page 153: Dicom Settings

    ANUAL 3-5-2-6 DICOM settings To set the DICOM properties for your LOGIQworks system: 1.) Select Configuration > Generic... from the Data Selector menu. 2.) Select the DICOM tab from the Generic Properties dialog. These steps describe how to enter the correct Application Entity Title and Port number.
  • Page 154 ANUAL 3-5-2-7 Monitors of System Dell Precision 340 To set up LOGIQworks for use with single monitor log on as system administrator: Administrator only Figure 3-76 Single Monitor Setup 1.) Select Configuration > Generic... from the Data Selector menu and click the monitor tab.
  • Page 155: Monitors Of System Dell Precision 340

    ANUAL 3-5-2-7 Monitors of System Dell Precision 340 (cont’d) Setting up LOGIQworks for use with dual monitors log on as system administrator: Administrator only Figure 3-77 Dual Monitor Setup 1.) Select Configuration -> Generic... from the menu bar in the Data Selector and click the Monitor tab.
  • Page 156 3-5-2-8 Monitors of System Dell Precision 360 and of Dell Precision 370 To set up LOGIQworks for use with single monitor log on as system administrator: Administrator only Figure 3-78 Single Monitor Setup for Dell 360 and for Dell 370 1.) Select Configuration >...
  • Page 157: Monitors Of System Dell Precision 360 And Of Dell Precision 370

    ANUAL 3-5-2-8 Monitors of System Dell Precision 360 and of Dell Precision 370 (cont’d) Setting up LOGIQworks for use with dual monitors for true dual graphic card log on as system administrator. Administrator only Figure 3-79 Simulated Dual Monitor Setup for Dell 360 and Dell 370...
  • Page 158 Figure 3-80 Real Dual Monitor Setup for Dell 360 only Setting up LOGIQworks for use with true dual monitors log on as system administrator: 1.) Select Configuration -> Generic... from the menu bar in the Data Selector and click the Monitor tab.
  • Page 159: Study Statuses

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-2-9 Study Statuses To add/remove/modify study statuses: 1.) Select Configuration > Generic... from the Data Selector menu. 2.) Select the Study Statuses tab from the Generic Properties dialog.
  • Page 160 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-2-9 Study Statuses (cont’d) Figure 3-82 Study Statuses including user defined status The backslash separates the DICOM description with your own description of that particular status. For this you may use more than one word separated by spaces.
  • Page 161: Study Status Permission

    After adding the necessary statuses, the System Administrator has to configure the status permissions for each LOGIQworks user. The Allow (other) users to configure their status permissions check box must be checked (default after installation) in order to do this. Each user of this particular workstation should be present while doing so.
  • Page 162 Figure 3-85 Generic Properties - Study Status Permission (with user added study status) Set which study statuses the user will be able to select from when presented with the opportunity to change study status while using LOGIQworks. Check the check boxes of the statuses the user will be able to select.
  • Page 163: His-Ris

    1.) Select Configuration > Generic... from the Data Selector menu. Figure 3-87 Generic Properties - HIS-RIS In order to be able to use work lists you must configure or add HIS-RIS implementations. LOGIQworks comes with several Dynamic Link Libraries (DLL) which enable the system to connect to a HIS-RIS system, providing there is an external program that can make use of the supplied DLLs.
  • Page 164 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-2-11 HIS-RIS (cont’d) Functionality • The Functionality list shows the various functionalities which are available in the HIS-RIS Implementations. Worklist: The Worklist which will be retrieved from the HIS-RIS and shown in the Data Selector.
  • Page 165: Viewing Properties

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-2-12 Viewing Properties To set general viewing properties: 1.) Select Configuration > Viewing... from the Data Selector menu. 2.) Select the General tab from the Viewing Properties dialog.
  • Page 166 Select Series if you want to display all the series of the first study, and then all the series of any remaining studies. Show Menu Bar: Check this check box if you want LOGIQworks to display in the Viewing Section a menu bar with commands for viewing images.
  • Page 167 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-2-12 Viewing Properties (cont’d) Open Priors for viewing Opens for viewing priors associated with a study in the relevant worklist. Please refer to the RadWorks 6.1 Installation guide, page 70 for more information on the viewing properties and the other configuration tabs on the Viewing Properties page.
  • Page 168: General Application Level Configuration For Logiqworks Rev. 2.0 And Rev. 3.0

    WORKS ASIC ERVICE ANUAL 3-5-3 General Application Level Configuration for LOGIQworks Rev. 2.0 and Rev. 3.0 3-5-3-1 TruAccess Plugin Definition Verify that the TruAccess Raw Data Processing plugin is already defined and activate the plugin if not yet done. 1.) Select Configuration -> Plugin Definitions... from the Data Selector menu.
  • Page 169 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-3-1 TruAccess Plugin Definition (cont’d) Resulting list of installed plugins after TruAccess Raw Data Processing has been added: Figure 3-90 LOGIQworks- Plugin Definition Properties (2) Chapter 3 Installation 3-103...
  • Page 170: General Settings

    Figure 3-91 Generic Properties - General Settings • Main Window: Initial View: Check from the Initial View drop-down list the Worklist View you want LOGIQworks to display by default when starting. (Default: nothing selected, Local Studies will be default then) 3-104 Section 3-5 - Configuration...
  • Page 171: Logging

    Compare Patients on: Indicate how you want to view patient studies in a patient worklist: by name, ID or both. LOGIQworks will collapse all the studies into one folder for viewing on the Local Patient Worklist, provided that the studies have the same patient name and/or patient ID.(Default: Both name and ID)
  • Page 172: Popup Menu

    Always Flush File after Write: If checked, all logging actions will be written immediately to the log • file. This option will have a serious impact on the performance of LOGIQworks, so it should not normally be checked. (Default: not checked)
  • Page 173: System Properties

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-3-4 Popup Menu (cont’d) Figure 3-93 General Settings - Popup Menu Check the boxes next to the menu items to select the ones you want to appear on the pop-up menu that appears when right-clicking in the Data Selector.
  • Page 174 3-70). (Default: METRIC) If you wish to overrule Windows' regional settings, you can select either METRIC (Default setting) or US from the drop-down list. LOGIQworks will then use millimeters or inches irrespective of what is specified in the Windows Control Panel.
  • Page 175 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-3-5 System Properties (cont’d) Figure 3-95 Generic Properties - System Settings (2) Person name representation: • Here you can indicate how you want patient names to appear on work lists.
  • Page 176: Dicom Settings

    ANUAL 3-5-3-6 DICOM settings To set the DICOM properties for your LOGIQworks system: 1.) Select Configuration > Generic... from the Data Selector menu. 2.) Select the DICOM tab from the Generic Properties dialog. These steps describe how to enter the correct Application Entity Title and Port number.
  • Page 177 ANUAL 3-5-3-7 Monitors of System Dell Precision 380 and Dell Precision 390 To set up LOGIQworks for use with single monitor log on as system administrator: Administrator only Figure 3-97 Single Monitor Setup for Dell 380 and for Dell 390 1.) Select Configuration >...
  • Page 178: Monitors Of System Dell Precision 380 And Dell Precision 390

    ANUAL 3-5-3-7 Monitors of System Dell Precision 380 and Dell Precision 390 (cont’d) Setting up LOGIQworks for use with dual monitors for true dual graphic card log on as system administrator. Administrator only Figure 3-98 Simulated Dual Monitor Setup for Dell 380 and Dell 390...
  • Page 179: Study Statuses

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-3-8 Study Statuses To add/remove/modify study statuses: 1.) Select Configuration > User... from the Data Selector menu. 2.) Select the Study Statuses tab from the User Properties dialog.
  • Page 180 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-3-8 Study Statuses (cont’d) Figure 3-100 Study Statuses including user defined status The backslash separates the DICOM description with your own description of that particular status. For this you may use more than one word separated by spaces.
  • Page 181: Study Status Permission

    After adding the necessary statuses, the System Administrator has to configure the status permissions for each LOGIQworks user. The Allow (other) users to configure their status permissions check box must be checked (default after installation) in order to do this. Each user of this particular workstation should be present while doing so.
  • Page 182 Figure 3-103 User Properties - Study Status Permission (with user added study status) Set which study statuses the user will be able to select from when presented with the opportunity to change study status while using LOGIQworks. Check the check boxes of the statuses the user will be able to select.
  • Page 183: His-Ris

    1.) Select Configuration > Generic... from the Data Selector menu. Figure 3-105 Generic Properties - HIS-RIS In order to be able to use work lists you must configure or add HIS-RIS implementations. LOGIQworks comes with several Dynamic Link Libraries (DLL) which enable the system to connect to a HIS-RIS system, providing there is an external program that can make use of the supplied DLLs.
  • Page 184 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-3-10 HIS-RIS (cont’d) Functionality • The Functionality list shows the various functionalities which are available in the HIS-RIS Implementations. Worklist: The Worklist which will be retrieved from the HIS-RIS and shown in the Data Selector.
  • Page 185: Viewing Properties

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-3-11 Viewing Properties To set general viewing properties: 1.) Select Configuration > Viewing... from the Data Selector menu. 2.) Select the General tab from the Viewing Properties dialog.
  • Page 186 Select Series if you want to display all the series of the first study, and then all the series of any remaining studies. Show Menu Bar: Check this check box if you want LOGIQworks to display in the Viewing Section a menu bar with commands for viewing images.
  • Page 187 The following configuration steps assume, that you already have two monitors configured for your RA600 / CA1000 system, and that you have now purchased a 3rd monitor and the LOGIQworks plugin, and would like to extend your system to a TripleMonitor system.
  • Page 188 ERVICE ANUAL 3-5-3-12 TripleMonitors for LOGIQworks FX and LOGIQworks SoftwareOnly (cont’d) Your Adapters section may have three entries with "1 x 1". This is fine. NOTE Your Monitor groups section may have two entries, each with "1 x 1". This is fine.
  • Page 189 ASIC ERVICE ANUAL 3-5-3-12 TripleMonitors for LOGIQworks FX and LOGIQworks SoftwareOnly (cont’d) 8.) Set the number in the "Hanging protocol group" text field to 2. Figure 3-109 TripleMonitor Viewing Properties 9.) Quit and restart RA600 / CA1000 as prompted. 3-5-3-12-3 Configuring Your TripleMonitor System for Optimal Usage.
  • Page 190 ERVICE ANUAL 3-5-3-12 TripleMonitors for LOGIQworks FX and LOGIQworks SoftwareOnly (cont’d) 3.) You should also see that your left most monitor essentially remains empty (it will most likely show a black empty viewport). It should look roughly like this: Figure 3-110 TripleMonitor Black Empty Viewport Your general UI configuration will most likely differ from the above screenshot.
  • Page 191: Triplemonitors For Logiqworks Fx And Logiqworks Softwareonly

    ASIC ERVICE ANUAL 3-5-3-12 TripleMonitors for LOGIQworks FX and LOGIQworks SoftwareOnly (cont’d) 5.) You should now see your data selector on the left-most monitor. Your left-most monitor should look something like this: Figure 3-111 TripleMonitor Data Selector Chapter 3 Installation...
  • Page 192 ANUAL 3-5-3-12 TripleMonitors for LOGIQworks FX and LOGIQworks SoftwareOnly (cont’d) 6.) If the data selector doesn't expand to take up the entire workspace of the left-most monitor (as shown in the above screenshot), then click-grab the top edge of the data selector window, and drag...
  • Page 193 3-5-3-12-6 Using the LOGIQworks plugin in the TripleMonitor environment When you wish to review an image in the LOGIQworks plugin, we strongly recommend using the following workflow: 1.) Double click the study you wish to review an image from.
  • Page 194 TripleMonitors for LOGIQworks FX and LOGIQworks SoftwareOnly (cont’d) 3.) Once you have located the image you wish to open in LOGIQworks, set focus in the left monitor, and hit the "2" key on your num lock pad (make sure your Num Lock key is activated) 4.) This will hide the data selector on the left monitor, and display a black empty viewport.
  • Page 195 ASIC ERVICE ANUAL 3-5-3-12 TripleMonitors for LOGIQworks FX and LOGIQworks SoftwareOnly (cont’d) 6.) Click the image in the pictorial index, and you should now have that image visible in the viewport of the left monitor. Figure 3-116 TripleMonitor Image Visible...
  • Page 196 ANUAL 3-5-3-12 TripleMonitors for LOGIQworks FX and LOGIQworks SoftwareOnly (cont’d) 7.) To activate the LOGIQworks plugin, right click in the viewport, and select Series Area Plugin -> TruAccess Raw Data Processing. Figure 3-117 TripleMonitor TruAccess Raw Data Processing 8.) The selected image should now be open in the LOGIQworks plugin.
  • Page 197 ANUAL 3-5-3-12 TripleMonitors for LOGIQworks FX and LOGIQworks SoftwareOnly (cont’d) 10.)This will display the RA600 / CA1000 toolbars on the left monitor. This is where all the LOGIQworks related controls are located. Figure 3-118 TripleMonitor LOGIQworks 11.)To review another image of the same study in the LOGIQworks plugin, ensure that the left monitor viewport is selected, and select any other image in the left monitor pictorial index.
  • Page 198 Summary: Using the LOGIQworks Plugin in the TripleMonitor Environment • The LOGIQworks plugin is only available on the left (color LCD) monitor. • At any time can you review your images on the middle and right monitors, just as you are used to.
  • Page 199: Connection Service Properties Up To Logiqworks 1.3

    EVISION WORKS ASIC ERVICE ANUAL 3-5-4 Connection Service Properties up to LOGIQworks 1.3 ONLY CHANGE THE DEFAULT SETTINGS IN THE CONNECTION SERVICE PROPERTIES IF YOU NOTE ARE AN EXPERT! 3-5-4-1 General The General Tab contains a collection of generic settings for the Connection Service.
  • Page 200: Module

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-4-2 Module To view the JPEG Options: 1.) Select Configuration > Connection Service... from the Data Selector menu. 2.) Select the Module tab from the Connection Service Properties page.
  • Page 201 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-4-3 Logging (cont’d) D:\RA600\LOG Figure 3-121 Connection Service Properties - Logging (Default Settings) Logging Level: Set the level of logging messages. If set to ALL, all messages will be logged. If set •...
  • Page 202: Database

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-4-3 Logging (cont’d) Set the Maximum Size of the log file to 0 to disable log file truncation. This can be very handy for troubleshooting. Always Flush File after Write: If checked, all logging actions will be written immediately to the log •...
  • Page 203 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-4-4 Database (cont’d) Administrator only Figure 3-122 Connection Service Properties - Database (Default Settings) Chapter 3 Installation 3-137...
  • Page 204: Maintenance

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-4-4 Database (cont’d) Connect DSN: Name of the ODBC database source (DSN) for the database used by the • Connection Service. Destinations: Name of the database table containing pre-configured destinations.
  • Page 205 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-4-5 Maintenance (cont’d) Administrator only Administrator only Figure 3-123 Connection Service Properties - Maintenance (Default Settings) Startup Delay (sec.): Specifies the delay after start-up when the first maintenance check will be •...
  • Page 206: Receive

    Thread Priority: Priority of the maintenance thread within the Connection Service. Changing this • setting can influence the system performance. (Default: IDLE) Time Out DICOM Storage Commitment (h): Number of hours LOGIQworks will wait for a DICOM • Storage Commitment response before timing out.
  • Page 207 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-4-6 Receive (cont’d) Figure 3-124 Connection Service Properties - Receive (Default settings) • Generic Receive Verify Origin: Check this check box if you want the system to perform 'origin checking'. This option is available for DICOM transfers made to this system.
  • Page 208: Dicom

    This will save disk space, but performance will be affected to some extent when studies are opened for viewing, since LOGIQworks will have to decompress them before they are displayed. Select the type of compression you want RadWorks to use from the Format drop-down list.
  • Page 209 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-4-7 DICOM (cont’d) Figure 3-125 Connection Service Properties - DICOM (Default settings) The settings are divided into a section for DICOM Receive, followed by a section for DICOM send. The...
  • Page 210: Transmit

    LOGIQworks should be able to handle. (Default: 4) Receive time-out (s): The time after which LOGIQworks will report an error if no command is sent to the system after another system has initiated a connection with it. (Default: 60) DICOM Send •...
  • Page 211 (LOGIQworksor RadWorks) systems. Data received on other LOGIQworks machines (and sent from this machine) will have their origin set to this name. The name is best kept short and concise.
  • Page 212 RAS is actually installed on the system. Also make sure the RAS service dependency is enabled for the Connection Service. Send prior studies: Check this check box if you want LOGIQworks to automatically send to other • workstations prior studies associated with an active study.
  • Page 213: Connection Service Properties Of Logiqworks 2.0 And 3.0

    IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-5 Connection Service Properties of LOGIQworks 2.0 and 3.0 ONLY CHANGE THE DEFAULT SETTINGS IN THE CONNECTION SERVICE PROPERTIES IF YOU NOTE ARE AN EXPERT! 3-5-5-1 Module To view the JPEG Options: 1.) Select Configuration > Connection Service... from the Data Selector menu.
  • Page 214: Logging

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-5-1 Module (cont’d) Standard JPEG compression options: Each option is represented by one line. Before the semicolon a textual description can be entered, that appears in a list-box containing JPEG compression options.
  • Page 215: Database

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-5-2 Logging (cont’d) Logging Level: Set the level of logging messages. If set to ALL, all messages will be logged. If set • to NORMAL (the typical setting) only error messages will be logged.
  • Page 216 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-5-3 Database (cont’d) Administrator only Figure 3-129 Connection Service Properties - Database (Default Settings) 3-150 Section 3-5 - Configuration...
  • Page 217: Maintenance

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-5-3 Database (cont’d) Connect DSN: Name of the ODBC database source (DSN) for the database used by the • Connection Service. Destinations: Name of the database table containing pre-configured destinations.
  • Page 218 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-5-4 Maintenance (cont’d) Administrator only Administrator only Figure 3-130 Connection Service Properties - Maintenance (Default Settings) Startup Delay (sec.): Specifies the delay after start-up when the first maintenance check will be •...
  • Page 219: Receive

    Thread Priority: Priority of the maintenance thread within the Connection Service. Changing this • setting can influence the system performance. (Default: IDLE) Time Out DICOM Storage Commitment (h): Number of hours LOGIQworks will wait for a DICOM • Storage Commitment response before timing out.
  • Page 220 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-5-5 Receive (cont’d) Figure 3-131 Connection Service Properties - Receive (Default settings) • Generic Receive Verify Origin: Check this check box if you want the system to perform 'origin checking'. This option is available for DICOM transfers made to this system.
  • Page 221: Dicom

    This will save disk space, but performance will be affected to some extent when studies are opened for viewing, since LOGIQworks will have to decompress them before they are displayed. Select the type of compression you want RadWorks to use from the Format drop-down list.
  • Page 222 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-5-6 DICOM (cont’d) Figure 3-132 Connection Service Properties - DICOM (Default settings) The settings are divided into a section for DICOM Receive, followed by a section for DICOM send. The...
  • Page 223: Transmit

    LOGIQworks should be able to handle. (Default: 4) Receive time-out (s): The time after which LOGIQworks will report an error if no command is sent to the system after another system has initiated a connection with it. (Default: 60) DICOM Send •...
  • Page 224 (LOGIQworksor RadWorks) systems. Data received on other LOGIQworks machines (and sent from this machine) will have their origin set to this name. The name is best kept short and concise.
  • Page 225: Database Service Properties Up To Logiqworks 1.3

    RAS is actually installed on the system. Also make sure the RAS service dependency is enabled for the Connection Service. 3-5-6 Database Service Properties up to LOGIQworks 1.3 WARNING WARNING ONLY CHANGE THE DEFAULT SETTINGS IF YOU ARE AN EXPERT!
  • Page 226: Logging

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-6-1 General (cont’d) Pipe Name: Name of the pipe on which clients can connect with the Database Service. • Import priority: The priority with which incoming items are imported into the database.
  • Page 227: Auto Delete

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-6-2 Logging (cont’d) Truncation Size (kB): Set the truncation size of the logging file. The truncation size is the size the • log file will be shortened to when its maximum size has been exceeded. (Default: 800 kB) Maximum Size (kB): Set the maximum size of the log file.
  • Page 228: Input

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-6-3 Auto Delete (cont’d) Age deletion • Time to Wait: Specify the age of image data (in days / hours) before the deletion rules will apply. If set to 0, no deletion will take place; if set to 2 DAYS for example, image files 2 days old or older will be removed.
  • Page 229: Maintenance

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-6-4 Input (cont’d) Administrator only Figure 3-137 Database Service Properties - Input (Default settings) Input Directory: Directory which will be watched (monitored) for new data. Data converted to AMI •...
  • Page 230 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-6-5 Maintenance (cont’d) Administrator only Administrator only Figure 3-138 Database Service Properties - Maintenance (Default Settings) Startup Delay: Specify the delay (in seconds) after start-up before the first maintenance check will •...
  • Page 231: Local

    ODBC Parameters: Parameters required to create a DSN (Data Source Name) in ODBC. • Local Query Limit: The maximum number of studies or patients LOGIQworks will display as a • result of a local query. You can specify up to a maximum of 100,000. Leave the setting at 0 if you want LOGIQworks to be able to show the maximum.
  • Page 232: Logging

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-7-1 General (cont’d) 2.) Select the General tab from the Database Service Properties page. Administrator only Figure 3-140 Database Service Properties - General (Default Settings) Performance monitoring: Creates a log (DBS_Performance.log) of RadWorks' performance. The •...
  • Page 233 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-7-2 Logging (cont’d) Administrator only D:\RA600\LOG Figure 3-141 Database Service Properties - Logging (Default Settings) Logging Level: Set the level of logging messages. If set to ALL, all messages will be logged. If set •...
  • Page 234: Auto Delete

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-7-2 Logging (cont’d) Maximum Size (kB): Set the maximum size of the log file. (Default: 1024 kB) • Set the Maximum Size of the log file to 0 to disable log file truncation. This can be very handy for troubleshooting.
  • Page 235: Input

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-7-3 Auto Delete (cont’d) Deletion Rules: Rules which determine whether data will be deleted if it is older than the specified time frame. These rules are SQL (Structured Query Language) rules. Once one or more rules are entered, Auto Delete will be activated.
  • Page 236: Maintenance

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-7-4 Input (cont’d) Administrator only Figure 3-143 Database Service Properties - Input (Default settings) Input Directory: Directory which will be watched (monitored) for new data. Data converted to AMI •...
  • Page 237 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-7-5 Maintenance (cont’d) Administrator only Administrator only Figure 3-144 Database Service Properties - Maintenance (Default Settings) Startup Delay: Specify the delay (in seconds) after start-up before the first maintenance check will •...
  • Page 238: Local

    ODBC: Parameters required to create a DSN (Data Source Name) in ODBC. • Local Query Limit: The maximum number of studies or patients LOGIQworks will display as a • result of a local query. You can specify up to a maximum of 100,000. Leave the setting at 0 if you want LOGIQworks to be able to show the maximum.
  • Page 239: Print Services Up To Logiqworks 1.3

    Work Directory: Directory where temporary files generated by the Print Module/Service are • placed. Maximum Bitmap Size: Set the maximum size of bitmaps that LOGIQworks will export/print. • Maximum Number of Copies: Adjust the maximum number of copies of an image a printer can •...
  • Page 240: Logging

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-8-2 Logging To set and view RA600 logging properties: 1.) Select Configuration > Print Service... from the Data Selector menu. 2.) Select the Logging tab from the Print Service Properties page.
  • Page 241 • (Default: \\.\pipe\AMI2\Print\Logging) 3-5-8-3 Database To set database properties for the LOGIQworks Print Service: 1.) Select Configuration > Print Service... from the Data Selector menu. 2.) Select the Database tab from the Print Service Properties page. Administrator only Figure 3-148 Print Service Properties - Database (Default settings) Print DSN: The name of the (ODBC) Data Source Name for the Print Service.
  • Page 242 ERVICE ANUAL 3-5-8-4 Maintenance To specify maintenance tasks relating to the LOGIQworks Print Service: 1.) Select Configuration > Print Service... from the Data Selector menu. 2.) Select the Maintenance tab from the Print Service Properties page. Administrator only Administrator only...
  • Page 243 Medium Type for Preview: Select a medium size from the drop-down list of common sizes. • LOGIQworks will use this size when creating the previews. Click on the Change... button to change the selection of layouts to be shown on the Print tool tab •...
  • Page 244: General

    Administrator only Figure 3-151 Print Service Properties - General (Default settings) Maximum Bitmap Size: Set the maximum size of bitmaps that LOGIQworks will export/print. • Maximum Number of Copies: Adjust the maximum number of copies of an image a printer can •...
  • Page 245 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-9-2 Logging To set and view RA600 logging properties: 1.) Select Configuration > Print Service... from the Data Selector menu. 2.) Select the Logging tab from the Print Service Properties page.
  • Page 246: Logging

    (Default: not checked)) 3-5-9-3 Database To set database properties for the LOGIQworks Print Service: 1.) Select Configuration > Print Service... from the Data Selector menu. 2.) Select the Database tab from the Print Service Properties page. Administrator only Figure 3-153 Print Service Properties - Database (Default settings) Print DSN: The name of the (ODBC) Data Source Name for the Print Service.
  • Page 247: Maintenance

    ERVICE ANUAL 3-5-9-4 Maintenance To specify maintenance tasks relating to the LOGIQworks Print Service: 1.) Select Configuration > Print Service... from the Data Selector menu. 2.) Select the Maintenance tab from the Print Service Properties page. Administrator only Administrator only...
  • Page 248 Medium Type for Preview: Select a medium size from the drop-down list of common sizes. • LOGIQworks will use this size when creating the previews. Click on the Change... button to change the selection of layouts to be shown on the Print tool tab •...
  • Page 249: Connectivity (Application Level)

    Access to your system can be restricted using “origins”. When another system tries to communicate with your LOGIQworks, you can tell your system to only allow communication if this origin is one of a pre- configured set of permitted origins.
  • Page 250 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-10-1 Setting up and configuring origins (cont’d) Figure 3-157 Origin Properties Page 3.) Enter the appropriate information in the dialog box (Name, Host Name or IP address, DICOM AE Title, DICOM Port Number are mandatory fields), any additional fields may help for troubleshooting.
  • Page 251 Setting up and configuring origins (cont’d) Figure 3-158 Configure Origins (Example) Please refer also to RadWorks 6.1 Installation Guide, page 61 LOGIQworks up to Version 1.3 or to Radworks 7.0 Installation Guide for LOGIQworks Version 2.0 or 3.0. Chapter 3 Installation 3-185...
  • Page 252: Setting Up And Configuring Destinations

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-10-2 Setting up and configuring destinations To add a teleradiology destination 1.) Select Connection -> Destinations... from the menu bar in the Data Selector. 2.) In the Destinations dialog box, click Add. The Add Destinations dialog opens.
  • Page 253 6.) For more detail on the options and IODs to select, please refer to the RadWorks 6.1 Installation Guide, page 57 for LOGIQworks up to Version 1.3 or Radworks 7.0 Installation Guide for LOGIQworks Version 2.0 or 3.0. 7.) Click Next.
  • Page 254 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-10-2 Setting up and configuring destinations (cont’d) Figure 3-161 DICOM Destination 8.) Enter the DICOM AE Title for the remote system. 9.) Enter the DICOM Port Number for the remote system.
  • Page 255: Setting Up Auto-Routing

    3.) Change any of the items in the Destination Properties dialog box (refer to the online help for details on the tabs and all parameters) Connection Parameters: The Backup Destination is used to provide redundancy. Whenever sending studies to a destination fails, LOGIQworks will send the data to the backup. Please see RadWorks 6.1 Installation Guide, page 57, for LOGIQworks up to Version 1.3 or Radworks 7.0 Installation Guide for...
  • Page 256 Even if no protocols have yet been added you will see the Basic Protocol. This is a simple autoroute protocol installed with LOGIQworks that allows you to set up basic auto routing on your system. You can modify the Basic Protocol but you cannot delete it.
  • Page 257 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-10-3 Setting up auto-routing (cont’d) Figure 3-164 Basic Auto Route Parameters 2.) Select a destination from the drop-down list, in the Auto Route Basic Teleradiology box.
  • Page 258 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-10-3 Setting up auto-routing (cont’d) 2.) In the Auto Route dialog box, click <basic protocol> (in the Name column) and then Modify. Select <none> from the Destination drop-down list in the Auto Route Basic Teleradiology dialog box.
  • Page 259: Cd-R/Rw And Dvd Setup

    1.) Select Configuration -> Generic... from the Data Selector menu bar and click on the HIS-RIS tab in the Generic Properties dialog box. 2.) Select Default HIS/RIS Implementation in the Implementations section and then click on the Config... button. Figure 3-166 Generic Properties - HIS-RIS for LOGIQworks up to Version 1.3 Chapter 3 Installation 3-193...
  • Page 260 3-5-11-1 Reading DICOM volume from DVD or CD (cont’d) Figure 3-167 Generic Properties - HIS-RIS for LOGIQworks Version 2.0 and 3.0 3.) Select the Study/Patient list tab of the Configure RadWorks Default HIS/RIS Implementation dialog box. Click the Add button to add an additional list.
  • Page 261 3-5-11-1 Reading DICOM volume from DVD or CD (cont’d) Figure 3-168 Generic Properties - Add Study/Patient List for LOGIQworks up to Version 1.3 Figure 3-169 Generic Properties - Add Study/Patient List for LOGIQworks Version 2.0 or 3.0 Chapter 3 Installation...
  • Page 262 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-11-1 Reading DICOM volume from DVD or CD (cont’d) Figure 3-170 Study/Patient List 4.) In the Patient/Study List dialog box type in a Description of the Patient or Study list you want to create.
  • Page 263 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-11-1 Reading DICOM volume from DVD or CD (cont’d) Figure 3-171 Source Dialog Box 8.) Click the Test button to see whether your connection will work. If you get a message stating that the connection has failed, go back and correct any mistakes you have made.
  • Page 264 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-11-1 Reading DICOM volume from DVD or CD (cont’d) 1.) In the Data Selector, select Database -> Add view from the menu bar. Figure 3-173 Database - Add View (1) 2.) In the Add View dialog, specify the Description field by typing in a description for your new study...
  • Page 265: Writing A Dicom Volume To Dvd Or Cd-R/Rw (Sme)

    12 "Viewing Properties" on page 3-99). 7.) If you want to import the study, select it, right-click it and then left-click Import from the shortcut menu. LOGIQworks will return the file to the worklist view where it came from. Figure 3-177 Import Study 3-5-11-2...
  • Page 266 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-11-2 Writing a DICOM volume to DVD or CD-R/RW (SME) (cont’d) WARNING After writing a CD/DVD, eject the media to finalize the session and read the media.
  • Page 267: Writing A Dicom Volume To Dvd Or Cd-R/Rw

    7.) In the “Image directory” field, type in the path and directory on your hard disk where you want LOGIQworks to temporarily store images before burning them to CD or DVD. You can also use the Browse... button to navigate to this location.
  • Page 268: Optional Peripherals/Peripheral Connections

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-12 Optional Peripherals/Peripheral Connections 3-5-12-1 Network Printer (Option)- Connections and Configurations Refer to RadWorks 6.1 Installation Guide, page 140, for further details on DICOM, Windows and File Printer configuration.
  • Page 269: Software Option Configuration

    WORKS ASIC ERVICE ANUAL 3-5-13 Software Option Configuration 3-5-13-1 Available Software options for LOGIQworks up to Version 1.3 The following software options are available for LOGIQworks. • Single Media Exchange Part Number H49021LE. • DICOM Print Part Number H49021LF. •...
  • Page 270 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-5-13-3 Software Option Installation (cont’d) 3.) Press the Test Button to run the License Configuration test. 4.) The test results will be displayed, listing all tested components with Licensed or Not Licensed flag.
  • Page 271: Software Upgrades

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL Section 3-6 Software Upgrades To configure new software options for RA600 please see Software Option Configuration on page 3-203. For software upgrades of the LOGIQworksturnkey system please see the Update Installation Manual.
  • Page 272: Disable Raw Data Image Transfer For Le

    Figure 3-181 DICOM Image Storage Service While setting up the connectivity for LOGIQworks, make sure that the ’Allow Raw Data’ checkbox is NOT checked. This disables the RAW data image transfer. For detailed description, please also refer to the LOGIQ e Basic User Manual, Direction 5118586-100 Rev.2, page 16-71.
  • Page 273 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL This page was intentionally left blank. Chapter 3 Installation 3-207...
  • Page 274 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 3-208 Section 3-7 - Disable RAW data image transfer for Le...
  • Page 275: Functional Checks

    Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 The purpose of this section is to provide quick procedures to check the major Hardware & Software functions of the LOGIQworks. Table 4-1 Contents in Chapter 4 Section Description Page Number...
  • Page 276: System Check

    Check the Program start up and the Data Selector 4-3-2-1 on page 4-8 Check generic DICOM settings 3-5-2-6 on page 3-87 Check LOGIQworks connectivity settings: 3-5-10-2 on page 3-186 Receiving devices (destinations) Check LOGIQworks connectivity settings: 3-5-10-1 on page 3-183...
  • Page 277: Windows Level Check

    3-186) to ensure that the remote system and LOGIQworks have been correctly configured. There is a utility to test the DICOM association: DCMPing, which can be found in: C:\AMI_61\Bin for LOGIQworks up to 1.3 systems and in directory C:\Program Files\Centricity\RA600\7.0\Bin for LOGIQworks 2.0 and LOGIQworks 3.0.
  • Page 278 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 4-3-1-1 Check the DICOM Connection (cont’d) 5.Output Window Figure 4-1 DICOM Pinger 1.) Enter the Host name or IP address of the system the association shall be tested with.
  • Page 279 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 4-3-1-1 Check the DICOM Connection (cont’d) Figure 4-2 Calling Remote Device 5.) The results will be listed in the output window. Figure 4-3 Successful DICOM communication Chapter 4 Functional Checks...
  • Page 280 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 4-3-1-1 Check the DICOM Connection (cont’d) Figure 4-4 DICOM Ping failure If the DCM Ping fails please go to 4-3-1-2 on page 4-7 for further trouble shooting.
  • Page 281: Check Ethernet Connection

    2-14 for details. Check the status of the Ethernet connection to the devices your LOGIQworks system is set up to communicate with. Please look up all sending and receiving devices in the connections destination and origins list from the data selector menu. Refer to...
  • Page 282: Application Level Check

    If not already started automatically during system start up, go to the desktop and double-click the LOGIQworks icon When the application is up and running the initial view should be the following: Figure 4-6 LOGIQworks Data Selector Section 4-3 - System Check...
  • Page 283: Confirm Images Can Be Displayed On The Graphical User Interface

    Confirm images can be displayed on the graphical user interface 1.) If not already done, start the LOGIQworks application. 2.) Make sure you have GE Ultrasound Raw Data studies available in the local view or you have a CD or DVD work list view configured, see Section 3-5-11-1 "Reading DICOM volume from DVD or CD"...
  • Page 284 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 4-3-2-3 Confirm images can be sent to a configured destination device. Only applicable if your system is connected to a network. 1.) Select the data set you want to send to a remote device listed in the destinations list (see 3-5-10-2 "Setting up and configuring destinations"...
  • Page 285: Confirm Images Can Be Sent To A Configured Destination Device

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 4-3-2-3 Confirm images can be sent to a configured destination device. (cont’d) 5.) In the Send to... dialog box specify the format you would like to use from the Format drop-down list.
  • Page 286: Connectivity

    Teleradiology allows you to send series, studies, images, or image-related data to another LOGIQworks™ or other DICOM 3.0 compliant systems. Also the system is able to query and retrieve images from other DICOM 3.0 compliant systems in the network. The steps to prepare the system for sending and query retrieve are described in the following sections.
  • Page 287 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 4-3-3-4 Configuring Destinations (cont’d) Figure 4-13 Destinations 1.) Enter the name of the new destination. The name will be displayed in the list box on the Teleradiology Tooltab. The name field must be unique. The system will check this when you enter the name.
  • Page 288 Figure 4-14 Origins 4-3-3-5 Configuring Origins To increase system security, you can restrict communication with your LOGIQworks™ PC to a pre- configured set of permitted origins. To turn verification on: 1.) In the Data Selector, select Configuration -> Connection Service from the menu and click on the Receive tab.
  • Page 289 While the patient is open and the plugin not active you can send entire studies, selected images or series from your system to another LOGIQworks™ or DICOM 3.0 compliant system. 1.) In the Data Selector select an entire or partial study in the Study or Patient list.
  • Page 290 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 4-3-3-9 Monitoring the Send Process (cont’d) Figure 4-16 Send Queue 4-3-3-10 Remote View The filing tooltab lets you select which remote views will appear in the center of your screen.
  • Page 291 3.) Choose the "Device" tab. 1.) Enter the "Name" and the "IP Address". 4.) Choose the "Service" tab. 1.) Choose in the "Destination Device" field the "LOGIQworks" entry from the pull-down menu. 2.) Choose the "Dicom Image Storage" from the text field below.
  • Page 292 4.) Choose a "Dataflow View" from the list and add with <<. 6.) Choose the "Button" tab. 7.) Do as described below: Assume Select LOGIQWorks entries and assume it with >> in the Printflow View Figure 4-18 Configuring the scanner...
  • Page 293 ASIC ERVICE ANUAL 4-3-5 Service Platform Check for LOGIQworks up to Revision 2.0 4-3-5-1 Checking Global Service Platform connections 1.) Click on the Wrench icon on the desktop or in the Quick Launch bar. Figure 4-19 Service selection and Logon 2.) Select the logon level and enter the password for it to connect as shown below.
  • Page 294 ERVICE ANUAL 4-3-5-2 Checking iLinq Connection 1.) Click On the iLinq button on the Desktop. Wait until the system is connected to GE to check for new messages. You will see the update screen. Figure 4-21 iLinq Home screen 4-20...
  • Page 295 EVISION WORKS ASIC ERVICE ANUAL 4-3-6 Modem Check (For LOGIQworks 3.0 optional) On/Off Switch Indicators flash here Figure 4-22 Modem front 4-3-6-1 Power-On 1.) Test the modem by turning it on (a power on/off switch is located on the right side). When you apply power, the modem performs a diagnostic self-test, indicated by several indicators flashing for a second or two, after which the 56 indicator should light.
  • Page 296 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 4-3-7 DVD Writer Check 4-3-7-1 Check reading from DVD Eject Figure 4-23 DVD-Drive Front Table 4-4 Using DVD Disk Drive Step Task Expected Result(s) Press the Eject Button on the DVD drive front to eject the disk tray The DVD drive tray is open.
  • Page 297 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 4-3-8 Archiving data with the Single Media Exchange Software Option To test the SME functionality first select some studies in the Data Selector for writing to the SME.
  • Page 298 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 4-3-8 Archiving data with the Single Media Exchange Software Option (cont’d) Figure 4-25 Select for Archiving (using the Right Click menu) 2.) When all the desired studies have been selected for archiving, first ensure a writable CD or DVD is in the drive.
  • Page 299 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 4-3-8 Archiving data with the Single Media Exchange Software Option (cont’d) Figure 4-26 Create Volume (Tool tab) 3.) Start the write operation either a.) by clicking the Local Database > Create Volume button on the Archive tab to the right of the...
  • Page 300: Healthpage

    Section 4-4 Healthpage 4-4-1 Healthpage for LOGIQworks up to Rev. 1.3 The Healthpage tool indicates some details around LOGIQworks.The user can view the license, the running services and log files. All log files are also available under 'Service Browser'. NOTE 1.) To start 'Healthpage', click on the desktop icon:...
  • Page 301: Healthpage For Logiqworks Up To Rev. 2.0 And 3.0

    4-4-2 Healthpage for LOGIQworks up to Rev. 2.0 and 3.0 The Healthpage tool indicates some details around LOGIQworks.The user can view the license, the running services and log files. All log files are also available under 'Service Browser'. NOTE 1.) To start 'Healthpage', click on the desktop icon:...
  • Page 302 WORKS ASIC ERVICE ANUAL 4-4-1 Healthpage for LOGIQworks up to Rev. 1.3 (cont’d) 2.) The main window of 'Healthpage' appears. Figure 4-31 Healthpage Window 3.) Start or stop a service 4.) The green light for each service shows that the service is running.
  • Page 303: Logfiles

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 4-4-3 Logfiles You can view the overall Logfile for RadWorks or the logfile for a single service. Figure 4-32 RadWorks Overall Logfile Chapter 4 Functional Checks 4-29...
  • Page 304: License Configuration

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 4-4-4 License Configuration 1.) Click 'Licenses', if you want to see what options are included in your license. NOTE: This is an example! Your output may differ...
  • Page 305: License Test

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 4-4-5 License Test Figure 4-34 Licence Test Chapter 4 Functional Checks 4-31...
  • Page 306: Site Log

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL Section 4-5 Site Log Date Service person Problem Comments Table 4-6 Site Log 4-32 Section 4-5 - Site Log...
  • Page 307: Components And Functions (Theory)

    ERVICE ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains LOGIQworks PC’s system concepts, component arrangement, and subsystem function. Table 5-1 Contents in Chapter 5 Section Description Page Number Overview Major Components...
  • Page 308: Major Components

    Section 5-2 Major Components LOGIQworks Work Station is a commercially available Dell PC which has Ethernet for communication with the local network or the hospital network. It has a modem for internet communications with on-line service. It has external read and write media for locally handled scan data.
  • Page 309 5-3-1-2 Theoretical Description of System based on Dell Precision 340 LOGIQworks runs on a conventional off the shelf PC with either 18.1” LCD monitor or 21” CRT monitor, an extended keyboard and a mouse. It communicates with Insight and iLinq via a modem interface which has dedicated power and phone line delivery.
  • Page 310: Block Diagrams And Theory

    Theoretical Description of System based on Dell Precision 360, Dell Precision 370, Dell Precision 380 and Dell Precision 390 LOGIQworks runs on a conventional off the shelf PC with either 19” LCD monitor or 20” LCD monitor, an extended keyboard and a mouse.
  • Page 311: Description Of Drives And Partition

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-3-1-8 Description of Drives and Partition Table 5-2 Description of Drives and Partition of System based on Dell Precision 340 Drive Partition Drive A 3.5” Floppy Disc...
  • Page 312 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-3-1 Work Station Overview (cont’d) Internet Access DELL MONITOR MONITOR Option Twisted Pairs Signal Cables Signal Cables Power Cables Mouse Keyboard Fire wire Cables Wall * Alternative a: a stand- alone network...
  • Page 313: Monitors

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-3-2 Monitors 5-3-2-1 Technical Specifications, CRT Monitor Manufacturer & Model Name: Sony, GDM-F520. 21" screen, (19,8" viewable image size) maximum resolution 2048 x 1536. Aperture Grille pitch 0.22 mm...
  • Page 314: Theoretical Description Of System Based On Dell Precision

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-3-2-4 Theoretical Description of System based on Dell Precision 340 The Monitors receive a video signal from the PC. Video Signal LCD (Analog) CRT(ANALOG) DELLPC Monitors (CRT...
  • Page 315 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-3-2-5 Theoretical Description of System based on Dell Precision 360, 370, 380, 390 The Monitors receive a video signal from the PC. Video Signal LCD (digital)
  • Page 316: Dell Pc

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-3-3 Dell PC 5-3-3-1 Technical Specifications, Dell Precision Workstation Main Processor - Pentium IV, 2.6 GHz Hard Disk Capacity 120 GB or 250GB DVD +R/RW, for S/W Upgrade 1024 MB RAM 3 1/2 “Floppy Disk, 1.44MB, for download data...
  • Page 317: Keyboard & Mouse

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-3-4 Keyboard & Mouse 5-3-4-1 Theoretical Description With the keyboard and mouse controls, functions and text are entered and sent to the PC. There they are interpreted and translated into digital or analog signal and presented on the monitor.
  • Page 318: Modem (Optional)

    5-3-5-2 Theoretical Description The Modem Receives power via a dedicated wall outlet. It is the telephonic link between the work station and the GE online centre. It communicates with the PC over an RS232 interface. Telephone Jack Wall Outlet Analog...
  • Page 319 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-3-5-2 Theoretical Description (cont’d) Telephone Jack Analog Signal MODEM DELL PC Figure 5-7 Modem Signal cabling information of the USB Modem Chapter 5 Components and Functions (Theory)
  • Page 320 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-3-6 USB HDD 5-3-6-1 Technical Specifications USB Manufacturer: DELL Interface: USB Capacity: 64MB 5-3-6-2 Theoretical descriptions The USB HDD receives power from the USB interface. It has read/write communications with the PC over a USB interface.
  • Page 321 5-4-1 Software Theoretics LOGIQworks is a standard PC application. It runs on the same operating system that is globally chosen by GE for all its cardiac and radiology products. LOGIQworks is a combination of the proven RadWorks (Centricity RA600 review workstation) multi-modality platform and GE’s TruAccess technology. In other...
  • Page 322 Double click on the selected study, series or images or right click on the selection and press View from the right-click-menu. When the data is opened for viewing you get to the viewing section of LOGIQworks. At this time you are still working in the RadWorks part of LOGIQworks, where only the DICOM standard image data is displayed.
  • Page 323 Display Window Pictorial Index Control Panel Figure 5-10 LOGIQworks Viewing Section When in this Viewing Section of LOGIQworks, there are many possible layouts or hanging protocols that can be used to show your information. Chapter 5 Components and Functions (Theory) 5-17...
  • Page 324 One of them is the TruAccess hanging protocol: Figure 5-11 TruAccess Hanging Protocol When you are viewing images from GE’s LOGIQ 7 and LOGIQ 9 ultrasound scanners that contain GE’s TruAccess Raw Data you can select this TruAccess hanging protocol to activate the TruAccess Raw Data Processing Plugin.
  • Page 325 (cont’d) Figure 5-12 TruAccess Plugin From the LOGIQworks TruAccess plugin images can be saved back to the local database in DICOM Secondary Capture format. They will be stored in a new series to the same study. As an alternative they can be also explicitly saved to disk either as AVI, JPEG, RawDICOM or DICOM.
  • Page 326: Software Theory

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-4-1 Software Theoretics (cont’d) New series containing Secondary Capture image stored from within the plugin Figure 5-13 Pictorial Index after Storing an image from the TruAccess plugin...
  • Page 327: Services

    Please note that the destination of imports is the LOGIQworks Connection Service. The LOGIQworks Print Service is responsible for managing print job queues for one or more printers or laser imagers. It supports several protocols of which the DICOM standard is one. Please note that the print service is an optional component of the LOGIQworks installation.
  • Page 328: Functional Definitions Of Application Entities

    The Application Entity of the LOGIQworks front-end application acts as an SCU for the query and retrieve SOP classes. The Application Entity of the LOGIQworks Print Service acts as an SCU for the basic gray scale and color print management meta SOP classes, and optionally the Print Job and Basic Annotation Box SOP classes.
  • Page 329: Query A Remote Database And Retrieve From A Remote System

    Print management The LOGIQworks user selects a study in the local database and submits it to the Print Service as a job for the desired laser imager. The user can choose several parameters such as the layout of the film and number of copies.
  • Page 330: Keyboard Shortcuts

    EVISION WORKS ASIC ERVICE ANUAL 5-4-10 Keyboard Shortcuts Table 5-6 Overview of LOGIQworks Shortcut checks Step Check Expected Result Press the CTRL + Q key when in Data Perform a query. Query Parameters dialog Selector. box pops up. Press the CTRL + I key when in Data View information Selector.
  • Page 331 IRECTION EVISION WORKS ASIC ERVICE ANUAL Table 5-6 Overview of LOGIQworks Shortcut checks Step Check Expected Result Load selected study for viewing (and Press ALT + S, L when in viewing. replace current loaded study) Add study (load selected study to compare Press ALT + S, A when in viewing.
  • Page 332 IRECTION EVISION WORKS ASIC ERVICE ANUAL Table 5-6 Overview of LOGIQworks Shortcut checks Step Check Expected Result Press ALT + I, C, O when in viewing. Start cine. Press ALT + I, C, M when in viewing. Create manual cine Press ALT + I, C, F when in viewing.
  • Page 333: Workstation Directory Tree Of Logiqworks Up To Revision 1.3

    WORKS ASIC ERVICE ANUAL 5-4-11 Workstation Directory Tree of LOGIQworks up to Revision 1.3 AMI_61 RA600 Root directory Archive Index Tool, containing the AIT database RA600 binaries (dlls, exe, manuals, tools like Healthpage, DCMPing etc.). RA600 CD Viewer, DICOM Viewer to display DICOM data from Media...
  • Page 334: Workstation Directory Tree Of Logiqworks 2.0 And 3.0

    ASIC ERVICE ANUAL 5-4-12 Workstation Directory Tree of LOGIQworks 2.0 and 3.0 RA600 Root directory RA600 binaries (dlls, exe, manuals, tools like Healthpage, DCMPing etc.). RA600 CD Viewer, DICOM Viewer to display DICOM data from Media TruAccess plugin root directory TruAccess plugin, executable Resources –...
  • Page 335: Common Service Interface

    The main application is displayed in the form of HTML pages whenever the Netscape browser starts. This is the entry point for any user to start any iLinq application. *These features that allow the customer to contact the GE OnLine Center are available for Warranty NOTE and Contract customers only.
  • Page 336: Global Service User Interface (Gsui)

    Global Service User Interface (GSUI) 5-5-3-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface. 5-5-3-2 Service Login Select the wrench icon in the Windows task bar at the bottom of the screen.
  • Page 337: Service Home Page

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-5-3-3 Access / Security (cont’d) Every access request, whether successful or not, will be logged into a service access log that is viewable to authorized users.
  • Page 338 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-5-4 Service Home Page (cont’d) Figure 5-18 Customer Service Home Page 5-32 Section 5-5 - Common Service Interface...
  • Page 339: Error Logs Page

    From the Error Logs tab the Log Viewer displays four categories with pull-down sub-menus and an Exit selection. The Service Interface allows scanner logs to be viewed by all service users. The Filter Error log is keyed to GE Service login access permissions and is not available to customer level analysis.
  • Page 340 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-5-5-2 Error Logs for TruAccess (cont’d) Table 5-8 Log Entry Key Severity Error Level Color Code Information Green Warning Blue Error The Service Interface supports the transfer of these logs to local destinations such as the MOD and CD- ROM drive.
  • Page 341 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-5-5-2 Error Logs for TruAccess (cont’d) • Informatics Log entry headings include TimeStamp; Revision; PtID; PtDOB; PtSex; PtWeight; PtHeight; ExamID; Exam Category; ExamCurDate and ExamStartTime.
  • Page 342: Error Logs For Truaccess

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-5-5-2 Error Logs for TruAccess (cont’d) 5-5-5-2-3 Exit The sub-menu, Exit Log Viewer, returns the user to the Service Desktop home page. 5-36 Section 5-5 - Common Service Interface...
  • Page 343 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-5-5-2 Error Logs for TruAccess (cont’d) Figure 5-22 Exit Log Sub-Menu Chapter 5 Components and Functions (Theory) 5-37...
  • Page 344: Diagnostics

    5-5-6 Diagnostics Detailed Diagnostic information is found in Section "Diagnostics/Troubleshooting" on page 7-1 Figure 5-23 GE Service Diagnostic Page 5-5-6-1 Diagnostics Execution Diagnostic tests are executable by both local and remote users. The Service Platform provides top-level diagnostic selection based on the user’s level and login access permissions. Remote access will require disruptive diagnostic permissions to run Acquisition diagnostics.
  • Page 345: Image Quality

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-5-7 Image Quality The Image Quality page is intended to contain tools for troubleshooting image quality issues. Figure 5-24 Image Quality Page 5-5-8 Calibration The Calibration page is intended to contain tools used to calibrate the system.
  • Page 346: Configuration

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-5-9 Configuration The Configuration page is intended to be used to setup various configuration files on the system. The Service Platform is the access and authorization control for remote access to the configuration subsystem.
  • Page 347: Utilities

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-5-10 Utilities The Utilities page contains several miscellaneous tools. Figure 5-27 Utilities/Tools Page These tools are available: • Windows utilities Disk usage: shows available and used disk space...
  • Page 348: Replacement

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 5-5-11 Replacement Not available for this product. 5-5-12 Not available for this product. 5-42 Section 5-5 - Common Service Interface...
  • Page 349: Service Adjustments

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of Chapter 6 This chapter explains how to do Monitor adjustments. Table 6-1 Chapter 6 Contents...
  • Page 350: Monitor Adjustment

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL Section 6-2 Monitor Adjustment 6-2-1 NEC 1880, 18,1" LCD The Quick Start Guide is delivered with the LCD Monitor. It is also available on the Internet at this location: http://www.plasmasync.com.
  • Page 351: Power-Save Setup

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 6-2-2-4 Color Adjustment 1.) See Dell LCD 1901/2001 Manual. 2.) To calibrate the monitor so that it is capable for diagnostic medical usage set the Color temperature to Native.
  • Page 352: Sony Triniton, Gdm-F520, 21" Crt

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 6-2-3 Sony Triniton, GDM-F520, 21" CRT The Sony Monitor Operating Instruction are delivered with the Monitor. WARNING WARNING Allow the monitor to warm up at least 20 minutes before making adjustment.
  • Page 353: Calibrate Crt Monitor

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 6-2-4 Calibrate CRT Monitor Before you will be able to view images at their true size, your monitor or monitors will have to be calibrated. Also, the picture on all display monitors will gradually change over time, so the image may become slightly smaller or larger.
  • Page 354 Type this length in millimeters in the box in the center right of the screen. 3.) Read the vertical length (in millimeters) automatically inserted by LOGIQworks in the other box in the center at the bottom of the screen and use the monitor's controls to set the lengths of the vertical lines equal to each other and to the value given by LOGIQworks.
  • Page 355: Diagnostics/Troubleshooting

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
  • Page 356: Collect Vital System Information

    Section 7-2 Collect Vital System Information LOGIQworks has a common interface for the user to collect critical system data in the event of a failure. This mechanism can be invoked by the globally available hotkey <Ctrl-Alt-D>. After entering the hotkey the default folder 'D:\Export' is suggested. To save the data on a different place enter a valid path without ending slash.
  • Page 357 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL Section 7-2Collect Vital System Information (cont’d) Figure 7-2 Save data Chapter 7 Diagnostics/Troubleshooting...
  • Page 358: Diagnostics

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL Section 7-3 Diagnostics The Diagnostics are part of the Service Browser. To enter the Service Browser follow Section 5-5-3-2 "Service Login" on page 5-30. 7-3-1 Global Service User Interface (GSUI) Diagnostic The Diagnostics are started from the Global Service User Interface Main Menu.
  • Page 359 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 7-3-1 Global Service User Interface (GSUI) Diagnostic (cont’d) Figure 7-4 GSUI Diagnostic Menu Chapter 7 Diagnostics/Troubleshooting...
  • Page 360: Virtual Console Observation

    Using the tool a remote service technician or the online center can access and modify all PC settings and programs on the customer's PC. Before a GE Service FE can access the customer’s PC remotely, first the customer has to enable this feature by choosing "Disruptive Mode" and confirming "Yes".
  • Page 361 GE M EDICAL YSTEMS 5184024-100, R LOGIQ ® B IRECTION EVISION WORKS ASIC ERVICE ANUAL 7-4-1 General (cont’d) Second the customer has to start the VCO: • Open Utilities. • Go to Common Utilities. • Open the Virtual Console Observation.
  • Page 362 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL This page was intentionally left blank. Section 7-4 - Virtual Console Observation...
  • Page 363: Replacement Procedures

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter contains replacement procedures for the following modules/subsystems: WARNING No covers or panels should be removed from the system (high-voltage risk).
  • Page 364: Fru Replacement Procedure

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL Section 8-2 FRU Replacement Procedure 8-2-1 PC Replacement Procedure 8-2-1-1 Preparations • Write down the Computer-ID of the PC. • If possible save all system presets (see Section 10-6-4-1 "Backup"...
  • Page 365: Crt Monitor Replacement Procedure

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-2-1-2 PC Removal (cont’d) Disconnect all cables connected on this partition. Remove the PC. Computer ID Disconnect all cables on the backend. Figure 8-2 PC disconnections and removal System based on Dell Precision 360 or on Dell Precision 370...
  • Page 366 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-2-2-2 Monitor Removal (cont’d) 4.) Remove the Monitor Disconnect Remove Figure 8-3 Monitor disconnection and removal Section 8-2 - FRU Replacement Procedure...
  • Page 367: Monitor Installation

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-2-2-3 Monitor Installation 1.) Position new Monitor where old one was positioned. 2.) Connect the Power cable to the wall outlet. 3.) Connect the Signal cables to the PC.
  • Page 368: Lcd Monitor Replacement Procedure

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-2-3 LCD Monitor Replacement Procedure 8-2-3-1 Preparations • Shut Down and turn Off the PC as described in chapters 3 and 4. 8-2-3-2 Monitor Removal 1.) Switch OFF the Monitor.
  • Page 369 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-2-3-3 Monitor Installation 1.) Position new Monitor where old one was positioned. 2.) Connect the Power cable to the wall outlet. 3.) Connect the Signal cables to the PC.
  • Page 370 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-2-4 LCD Monitor Replacement Procedure 8-2-4-1 Preparations • Shut Down and turn Off the PC as described in chapters 3 and 4. 8-2-4-2 Monitor Removal 1.) Switch OFF the Monitor.
  • Page 371 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-2-4-3 Monitor Installation 1.) Position new Monitor where old one was positioned. 2.) Connect the Power cable to the wall outlet. 3.) Connect the Signal cables to the PC.
  • Page 372: Modem Replacement Procedure

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-2-5 Modem Replacement Procedure 8-2-5-1 Modem of Dell 340 and Dell 360 8-2-5-1-1 Preparations • Shut Down and turn Off the Workstation and all its parts as described.
  • Page 373: Modem Of Dell Precision 370

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-2-5-2 Modem of Dell Precision 370 8-2-5-2-1 Preparations • Shut Down and turn Off the Workstation and all its parts as described. 8-2-5-2-2 Modem Removal 1.) Disconnect the phone cable from the wall jack.
  • Page 374: Modem Of Dell Precision 380 And Dell 390

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-2-5-4 Modem of Dell Precision 380 and Dell 390 8-2-5-4-1 Preparations • Shut Down and turn Off the Workstation and all its parts as described. 8-2-5-4-2 Modem Removal 1.) Disconnect the phone cable from the wall jack.
  • Page 375 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-2-6 Cable Replacement Procedure 8-2-6-1 Preparations • Shut Down and turn Off the Workstation and all its parts as described. This procedure is a general cable replacement procedure, and should be valid for Power, Fire wire, CAUTION Ethernet and RS-232 cables.
  • Page 376 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL Section 8-3 Recovery Procedure 8-3-1 Purpose of Software Recovery 8-3-1-1 Complete reformatting of a System WARNING All data will be lost. Your system will be set to factory settings.Store user settings on USB HDD and WARNING DICOM data on network store devices or CD-RW/DVD-RW if possible.
  • Page 377 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-3-1-1 Complete reformatting of a System (cont’d) Figure 8-12 Restore Menu 4.) Press 2. for restoring the system like Factory new condition. 5.) If you don’t want to proceed the Recovery Procedure, press 4. to exit DOS. Remove the Recovery...
  • Page 378 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-3-1-1 Complete reformatting of a System (cont’d) Figure 8-14 Ghost Surface 7.) While system restores the Ghost surface like on figure 8-15 is shown on the screen.
  • Page 379 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-3-1-1 Complete reformatting of a System (cont’d) Figure 8-15 Restore Complete 8.) After the restore has finished remove the CD from the CD drive and reboot the PC.
  • Page 380: Pc Is Working But System Is Corrupted

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-3-1-2 PC is working but System is corrupted WARNING All data will be lost. Your system will be set to factory settings. Store user settings on USB HDD and WARNING DICOM Data on network store devices or CD-RW/DVD-RW is possible.
  • Page 381 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-3-1-3 PC is working but Windows or application files are corrupted (cont’d) Figure 8-18 Restore Menu 3.) Press 3. for restoring your C: Partition like Factory new condition.
  • Page 382 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-3-1-3 PC is working but Windows or application files are corrupted (cont’d) Figure 8-20 Ghost surface partition restore 6.) While partition restore, the Ghost surface like on Figure 8-14 is shown on the screen.
  • Page 383: Pc Is Working But Software Is Corrupted

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-3-1-3 PC is working but Windows or application files are corrupted (cont’d) Figure 8-22 USB HDD check of System based on Dell Precision 340 Figure 8-23 USB HDD check of System based on Dell Precision 360 8.) Check the installation of USB HDD.
  • Page 384: Windows Message After Recovery Procedure

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-3-1-5 Windows Message after Recovery Procedure If the message shown in the picture below occurs click on the “Yes"-button and reboot the system. Figure 8-24 Windows Message...
  • Page 385: Recover With Spare Part Recovery Cd

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-3-1-6 Recover with Spare Part Recovery CD Recover the system with Spare Part recovery DVD P/N 2413251 for DELL 340 and 2366663-2 for DELL 360 and 2413250 for Dell 370 1.) Proceed the steps in...
  • Page 386 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-3-1-6 Recover with Spare Part Recovery CD (cont’d) Figure 8-26 Confirm License Agreement 8-24 Section 8-3 - Recovery Procedure...
  • Page 387 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-3-1-6 Recover with Spare Part Recovery CD (cont’d) 4.) Select “I accept this agreement” and click “Next”. Figure 8-27 Regional Settings 5.) To change the local settings for system and keyboard click “Customize”.
  • Page 388 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-3-1-6 Recover with Spare Part Recovery CD (cont’d) Figure 8-28 Product Key 7.) Type in the Product Key you will find on a label on top of the CPU box.
  • Page 389 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 8-3-1-6 Recover with Spare Part Recovery CD (cont’d) Figure 8-29 Date and Time Settings 9.) Verify date and time and if necessary change the Date & Time settings.
  • Page 390 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL This page was intentionally left blank. 8-28 Section 8-3 - Recovery Procedure...
  • Page 391: Renewal Parts

    ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Spare Parts available for the LOGIQworks. Table 9-1 Contents in Chapter 9 Section Description Page Number Overview List of Abbreviations...
  • Page 392: List Of Abbreviations

    FRU N - Field Replacement Unit - Non Stock Part. 9-2-3 General abbreviations QTY= Quantity 9-2-4 LOGIQworks PC versions covered in this manual Section 1-1-4 "LOGIQworks Models covered by this Manual" on page 1-2. Section 9-2 - List of Abbreviations...
  • Page 393: Renewal Parts List

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL Section 9-3 Renewal parts list 9-3-1 Stock Items FRU 1 and 2 Figure 9-1 LOGIQworks Parts identifications Chapter 9 Renewal Parts...
  • Page 394 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL Table 9-2 PC, Monitor, Keyboard and Mouse parts Item Part Name Part Number Qty. Description Dell Precision 340 2365947 Dell support on GPO 2366635 Spacesaver Keyboard US...
  • Page 395 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 9-3-2 Stock Items FRU 1 and 2 Figure 9-2 LOGIQworks Parts identifications Chapter 9 Renewal Parts...
  • Page 396 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL Table 9-3 PC, Monitor, Keyboard and Mouse parts Item Part Name Part Number Qty. Description Dell Precision 360 2365947-2 Dell Precision 370 2413249 Dell support not on GPO...
  • Page 397: Renewal Part Lists

    First update your system to LOGIQworks 1.3, after that to LOGIQworks 2.0 and in the end to LOGIQworks 3.0. For the update from LOGIQworks 1.3 to 3.0, two parts are required, item 2 and item 3. First update your system to LOGIQworks 2.0 and finally update your system to LOGIQworks 3.0.
  • Page 398: Product Manuals

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 9-3-4 Product Manuals Part Name Part Number Basic Service Manual English 5184024-100 Proprietary Service Manual English 5184027-100 Instruction for Use English 5177849-100 Release Notes English 5177928-100 Quick Guide English...
  • Page 399 5177986-144 Instruction for Use Danish 5177849-160 Release Notes Danish 5177928-160 Quick Guide Danish 5177986-160 Quick Card Danish 5177962-160 Quick Guide Norwegian 5177986-161 Quick Guide Finish 5177986-162 InSite II ViewPoint/LOGIQworks 5181936-100 Installation Manual Table 9-7 Product manuals Chapter 9 Renewal Parts...
  • Page 400 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL This page was intentionally left blank. 9-10 Section 9-3 - Renewal parts list...
  • Page 401: Quality Assurance

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL Chapter 10 Quality Assurance Section 10-1 Overview 10-1-1 Purpose of Chapter 10 This chapter describes Periodic Maintenance (PM) and customer performed tasks on the workstation and peripherals. These procedures are intended to maintain the quality of the computer system. Read this chapter completely and familiarize yourself with the procedures before performing a task.
  • Page 402: Why Perform Periodic Maintenance

    10-2-1 Keeping Records It is good business practice that ultrasound facilities maintain the records of periodic and corrective maintenance of the LOGIQworks which is a peripheral to radiology ultrasound scanners such as LOGIQ 7 and LOGIQ 9. 10-2-2 Quality Assurance...
  • Page 403: Periodic Maintenance Schedule

    10-3-1 How often should PMs be performed? The Maintenance Task Schedule specifies how often your LOGIQworks should be serviced and what items need attention. It is the customer’s responsibility to ensure the LOGIQworks periodic maintenance is performed as scheduled in order to retain its high level of safety, dependability and performance. Your GE Service Representative knows your LOGIQworks best and can provide competent, efficient service.
  • Page 404: Tools Required

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL Section 10-4 Tools Required 10-4-1 Necessary Tools 10-4-1-1 Specific Requirements for Periodic Maintenance Table 10-3 Overview of Requirements for Periodic Maintenance Tool Part Number Comments 46–194427P278 120V Anti Static Vacuum Cleaner 46–194427P279...
  • Page 405 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL Section 10-5 Safety Inspections 10-5-1 Electrical Safety Tests 10-5-1-1 Specifications to meet THE USER MUST ENSURE THAT SAFETY INSPECTIONS ARE PERFORMED AT LEAST WARNING WARNING EVERY 12 MONTHS ACCORDING TO THE REQUIREMENTS OF THE PATIENT SAFETY STANDARD IEC-EN 60601-1.
  • Page 406: Pc Maintenance

    10-6-4-1 Backup The system data (C: drive) has to be backed up every day in order to restore LOGIQworks to its latest state, in case of system failure. The recommended method to backup system data is to have a separate RW-DVD for each working day (e.g.
  • Page 407: Save

    10-6-5 Save and Restore Following settings are saved by the Save and Restore function: LOGIQworks user and system settings, as for example the Hanging Protocols or the Desktop-Window settings, Windows Network and Operating System settings as the IP-Address or the Host name.
  • Page 408 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 10-6-4-1 Backup (cont’d) Figure 10-2 RadWorks Settings Tool To save user settings: 1.) Click on 'RadWorks User'. Figure 10-3 RadWorks User Settings 2.) Click on 'Save'. 10-8...
  • Page 409 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 10-6-4-1 Backup (cont’d) Figure 10-4 RadWorks User Settings - Save 3.) Enter a valid device (use the USB mass storage device). 4.) Enter a file name for the user settings.This file name should reflect the date and the type of settings you are storing.
  • Page 410 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 10-6-4-1 Backup (cont’d) Figure 10-5 RadWorks System Settings 2.) Click on “Backup”. Figure 10-6 RadWorks System Settings - Save 3.) Enter a valid device (use the USB mass storage device).
  • Page 411 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 10-6-4-1 Backup (cont’d) Figure 10-7 Windows Script Host - Saved successfully 7.) Open Windows Explorer to check whether the Save operation has completed successfully. Select the drive letter that represents the USB HDD and verify that three files were saved to the USB HDD.
  • Page 412 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 10-6-5-2 Restore (cont’d) After that you have started the ’RA settings tool’, you can restore User and/or System Data. Figure 10-8 RadWorks Settings Tool To load user settings: 1.) Click on “RadWorks User”.
  • Page 413: Restore

    GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 10-6-5-2 Restore (cont’d) Figure 10-9 Restore RadWorks User Settings 2.) Select the user settings from the Available Settings List (see Figure 10-9). Make sure that you don’t select System settings.
  • Page 414 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 10-6-5-2 Restore (cont’d) Figure 10-10 Load user settings 4.) Select a previously saved file and click “Open”. Figure 10-11 Overwriting of User Settings 5.) Confirm with “OK”.
  • Page 415 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL 10-6-5-2 Restore (cont’d) Figure 10-12 Restore RadWorks System Settings. 2.) Select the system settings from the Available Settings list (see Figure 10-12). 3.) Click on “Restore”. Figure 10-13 Load System Settings 4.) Pick up a previously saved file and click “Open”.
  • Page 416 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL This page was intentionally left blank. 10-16 Section 10-6 - PC Maintenance...
  • Page 417 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL INDEX electrical, 2-7 Power-up Procedures Acoustic Noise Output, 3-28 Voltage Settings, 3-22 Adjustment Preparing the Unit for Installation, 3-22 Monitor, 6-2 Product Icons, 1-6 Purchaser Responsibilities, 2-11...
  • Page 418 GE M EDICAL YSTEMS 5184024-100, R LOGIQ IRECTION EVISION WORKS ASIC ERVICE ANUAL...
  • Page 420 GE Healthcare GE Medical Systems: Telex 3797371 P.O. Box 414; Milwaukee, Wisconsin 53201, U.S.A. (Asia, Pacific, Latin America, North America) ViewPoint Bildverarbeitung GmbH Tel. +49 8153 93 11-0 Argelsrieder Feld 12 82234 Wessling, Germany...

Table of Contents