GE LOGIQ 200 Quality Assurance Manual page 163

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 0
2--3--2
System Diagnosis (Continued)
TEST NAME: BPHV TEST
PURPOSE
To check BPHV BOARD ASSY.
CRITERIA
Refer to result shown on the monitor.
Action if the test fails
If RBF board assy tests passed, change BPHV board assy.
The following Assemblies are necessary for this test
RBF, MVP, MSTE, BPHV, SCAN, Monitor
Diagnosis
Definition
# define PS_THVP_LOWEST
# define PS_THVP_HIGHEST
# define PS_THVM_LOWEST
# define PS_THVM_HIGHEST
Procedure
Diagnisis_ThvpTest
Diagnisis_ThvmTest
AC
return Diagnisis_IsPowerTestOkay(POWERNUMBER_THVP,
0x6040, PS_THVP_LOWEST, PS_THVP_HIGHEST) ?
NOVA_OK : NOVA_FALSE_SVC_THVP;
return Diagnisis_IsPowerTestOkay(POWERNUMBER_THVM,
0x6040, PS_THVM_LOWEST, PS_THVM_HIGHEST) ?
NOVA_OK : NOVA_FALSE_SVC_THVM;
BPHV
VOL
Regulator
ILLUSTRATION 2--11
LOGIQ 200 PRO Series PROPRIETARY MANUAL
Output
Protection
From RBF
BPHV test TEST
2--26
+SHV
SHV
--SHV
Gen.
THV
Gen.
--THV
+THV
2242594
DIAGNOSTICS