GE LOGIQ 200 Quality Assurance Manual page 165

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 0
2--3--2
System Diagnosis (Continued)
TEST NAME: MVP FRONT/END TEST
PURPOSE
To check Image softener and Average Control circuit with its circuit.
CRITERIA
Refer to result shown on the monitor.
Action if the test fails
Change MVP board assy.
The following Assemblies are necessary for this test
Probe, MVP, MSTE, RBF, TBF, SCAN, Monitor
MVP
FRONT_END
Image
Softener
From RBF
B/M
Front_end
Control
To NOVA_BUS
Buffer
Frame
Average
PEAK
HOLD
DSP BLOCK
USC BUS
Buffer
I/F
LOGIQ 200 PRO Series PROPRIETARY MANUAL
B MODE
ITP
ALM
CRDTF
XYADDR
Gen.
CINE
CINE
Control
SAP
FIFO
DSP
MVP TEST
ILLUSTRATION 2--13
2--28
POST
Image
Storage RAM
Image
Display Control
Video
XY
Lut
memory
Interlace
ZOOMER
Convert
Non--Int
Convert
Video
Decoder
Sync
CINE
Generator
Memory
XY
Memory
Program
Memory
2242594
Video
DAC
Video
DAC
DIAGNOSTICS