System Diagnosis (Continued)
TEST NAME: Sector Probe Path Test
PURPOSE
To check a Sector Probe Path.
Overviews
--.Before starting this test, Test pattern for OQCARD should be loaded in the R data area of the test
probe.
Diagnosis
Condition
Probe Type
Probe Name
Mode
Depth
Frame Rate
Freeze or Melt
Procedure
Diagnisis_SectorPathTest
Result code
Decision
Check Image Patterns.
The following Assemblies are necessary for this test
BPHV, PROBE, RBF, TBF, SCAN, MVP, MSTE, Monitor
Linear Probe
Test Probe
Full B
4cm (TX scan ranges = 0 to 510)
high
Melt
Save the current console parameters.
Change console parameters according to above.
outword (0x4c06, 0x42)
Wait for exit command (Ctrl+X).
Restore console parameters.
Checked
2--24