GE LOGIQ 200 Quality Assurance Manual page 135

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 2
2–3–2
System Diagnosis (Continued)
TEST NAME: MONITOR TEST
PURPOSE
To check monitor gradation ability, video circuit and video characteristics.
This test assists to check latitude (capability of density) of B/W printer.
CRITERIA
Refer to test pattern and gray scale shown on the monitor as in ILLUSTRATION 2–24. 16 gray scale
on the upper side of monitor. 16 gray scale on the lower side of monitor. The source of test pattern and
gray scale is graphic controller in MST and B_mode display memory in DSC, respectively
DISPLAY FORMAT OF RESULTS
Action if the test fails
Change the Monitor.
The following Assemblies are necessary for this test
RTC, ESP, CONN, DSC, MST,Monitor
LOGIQ
191 mm
MONITOR TEST IMAGE
ILLUSTRATION 2–24
2–34
a
200 PROPRIETARY MANUAL
2138854
148.5 mm
DIAGNOSTICS