GE LOGIQ 200 Quality Assurance Manual page 112

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 2
2–3–2
System Diagnosis (Continued)
Procedure 3
Follow these steps to enter each test in the System Diagnosis menu.
1.
Using the Arrow up/down keys, select test which should be performed. Refer to ILLUSTRATION 2–6.
2.
Press Set or Return keys.
Refer to each section to perform the each test for System Diagnosis.
Procedure 4
The result of test will appear on the status column in the System Diagnosis menu. Refer to the ILLUSTRATION 2–7.
The AUTO type tests have three status as belows.
Running
It shows that this test item is doing now.
Pass
It shows that this test item finished without failure.
Fail : ###
It shows that this test item finished with failure. The "###" shows failure code. Refer to
TABLE 2–1 as detailed failure code.
Checked
It shows that this test item had been checked.
Skipped
It shows that DRAM test had not been executed at booting.
No PIOP
It shows that the option PIOP unit is not installed.
In case of CINE test, It shows the failure message when the option unit is not installed.
The MANUAL type tests have one status as belows.
Service Utility MENU :
Current Software Version is 2.0
NAME
System test
*Beam Former test
HV test
ADC test
DSC Front/end test
B mode Function test
B mode X–Y Memory test
B mode Cine test
M mode Function test
MST test
System DRAM test
Keyboard test
PIOP test
Monitor test
Power Supply test
ROM Version
Exit
System Diagnosis
Time
## M
0 M
## M
O M
## M
## M
## M
## M
## M
## M
## M
0 M
## M
0 M
## M
0 M
SERVICE DIAGNOSIS MENU–RESULT
ILLUSTRATION 2–7
2–11
a
LOGIQ
200 PROPRIETARY MANUAL
96/03/12 09:16:37
Status
Checked
Checked
Passed
Checked
Running
Passed
Passed
Fail : ###
Skipped
Checked
No PIOP
Cheeked
2138854
DIAGNOSTICS