GE LOGIQ 200 Quality Assurance Manual page 12

Table of Contents

Advertisement

g
GE Medical Systems
LINE VOLTAGE TEST
System OFF: ______________VAC__
LINE CURRENT TEST
System OFF: ______________Amps_
OUTLET WIRING TEST
Correct Wiring
Reversed Polarity
Hot & Gnd Reversed
GROUND CONTINUITY
QA Reference Manual Section 3-4. Use the Dale 600 Tester. Resistance should be less than 0.2
Ohms. If greater than 0.2 Ohms, perform detailed chassis leakage tests and troubleshoot.
TEST POINT
Console
Caster
Monitor
Ground Continuity Tests
CHASSIS SOURCE LEAKAGE TEST
QA Reference Manual Section 3-5. Max allowable is 300 uA.
IMPORTANT: Be sure to pause the polarity reverse switch in the OFF position for a minimum of 10
seconds before reversing polarity.
POWER
POLARITY
ON
NORM
ON
NORM
OFF
NORM
OFF
NORM
Chassis Source Leakage Tests Summary
TRANSDUCER LEAKAGE CURRENT TESTS
All readings in micro-amps. Maximum allowable leakage current is 100 uA for general purpose and
endocavitary or 50 uA for surgical probes. QA Reference Manual Section 3-9.
DALE 600
DALE 600
POLARITY
GND SW
NORM
Closed
NORM
Open
Transducer Leakage Current Test Summary:
Transducer 1
Transducer 2
(optional)
QA Reference Manual Section 3-3. If correct, proceed to the next step.
Resistance (ohms)
_______________
_______________
_______________
GROUND
CONSOLE
SWITCH
CHASSIS GND
Closed
Open
Closed
Open
PROBE 1
PASS
CHANGE
PASS
CHANGE
2/16
System ON: ________________VAC__
System ON: ________________Amps_
Open Ground Wire
Open Neutral Wire
Open Hot Wire
TEST POINT
Probe Screw
Foot Switch
Printer
VCR
PASS
FAIL
POWER
POLARITY
ON
ON
OFF
OFF
PASS
PROBE 2
Choose Pass or Fail and Change or No Change
FAIL
NO CHANGE
FAIL
NO CHANGE
2277615-100
REV
Resistance (ohms)
______________(any side screw)
________________
________________
________________
GROUND
CONSOLE
SWITCH
CHASSIS GND
REV
Closed
REV
Open
REV
Closed
REV
Open
FAIL
0