Cable Cross-Sections And Cable End Sleeves - ProMinent DULCOMETER D1Cb Series Instructions For Assembly And Use

Hide thumbs Also See for DULCOMETER D1Cb Series:
Table of Contents

Advertisement

Programmierschnittstelle
Programming interface
Schirmklemme
Ground terminal
Klemmenanordnung Ausführung Wandgerät
Terminal order type wall mounting
Manufacturing coaxial cable for the connection at D1Cb
or prefabricates in the variants
Konfektionierung Koaxialkabel zum Anschluss an D1Cb
oder vorkonfektioniert in den Varianten
0.8m - TN 1024105
2.0m - TN 1024106
5.0m - TN 1024107
Fig. 9: Preparation of coaxial cable
3.4.5

Cable Cross-Sections and Cable End Sleeves

Without cable end
sleeve
Cable end sleeve
without insulation
Cable end sleeve with
insulation
26
16
8
Minimum cross-section Maximum cross-
0.25 mm
2
0.20 mm
2
0.20 mm
2
Detail Anschluss Koaxialkabel D1Cb
Detail connection coaxial wire D1Cb
section
1.5 mm
2
1.0 mm
2
1.0 mm
2
Assembly
optional
RC-Schutzbeschaltung D1Cb
RC-protection D1Cb
Sicherung 5x20
fuse 5x20
Stripped insulation
length
8 - 9 mm
10 - 11 mm
29