Cable Cross-Sections And Cable End Sleeves; Installation Of Coaxial Cable To Guard Terminal Xe1 - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Compact controller measured variable: ph / orp
Table of Contents

Advertisement

Assembly and installation
5.3.1

Cable Cross-Sections and Cable End Sleeves

Without cable end
sleeve
Cable end sleeve
without insulation
Cable end sleeve
with insulation
5.3.2

Installation of coaxial cable to guard terminal XE1

CAUTION!
Maximum length of the coaxial cable 10 m
Incorrect measured value due to too long a coaxial cable
Possible consequence: Slight or minor injuries, material damage.
The maximum length of the coaxial cable may not exceed 10 m when using redox
or pH sensors. The measurement signal can otherwise be falsified by the effects
of interference.
Konfektionierung Koaxialkabel zum Anschluss an DCC
oder vorkonfektioniert in den Varianten
Fig. 13: Installation of Coaxial Cable to Guard Terminal XE1
When installing the coaxial cable for the guard terminal XE 1, the lengths of insulation to
be removed from the coaxial cable must be adhered to. The guard terminal
by hand.
34
Minimum cross-sec‐
tion
0.25 mm
2
0.20 mm
2
0.20 mm
2
Maximum cross-
section
1.5 mm
2
1.0 mm
2
1.0 mm
2
Detail Anschluss Koaxialkabel
Stripped insulation
length
8 - 9 mm
10 - 11 mm
A0362
'is tightened'

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dulcometer compact

Table of Contents