Assembly and installation
5.3.1

Cable Cross-Sections and Cable End Sleeves

Without cable end
sleeve
Cable end sleeve
without insulation
Cable end sleeve
with insulation
5.3.2
Electrical connection of the
chlorine sensor
CAUTION!
Length of sensor cable
The sensor is supplied with a fixed
cable.
Possible consequence: Slight or
minor injuries. Material damage.
Modifying the cable (lengthening,
shortening, etc.) is impermissible.
32
Minimum cross-sec‐
tion
0.25 mm
2
0.20 mm
2
0.20 mm
2
Maximum cross-
section
1.5 mm
2
1.0 mm
2
1.0 mm
2
Stripped insulation
length
8 - 9 mm
10 - 11 mm