HP  8360 L Series User Manual

HP 8360 L Series User Manual

Swept cw generator (including options 001, 004, and 008)
Table of Contents

Advertisement

SERIAL NUMBERS
This manual applies directly to any swept CW generator with the
model and serial number prefix combination shown below. You
may have to modify this manual so that it applies directly to your
instrument version. Refer to the "Instrument History" chapter.
HP 83623L/3OL/4OL/5OL
3844A and below
HEWLETT
PACKARD
HP Part No. 08380-90134
Printed in USA
February 1999 Supersedes: September 1997

Advertisement

Table of Contents
loading

Summary of Contents for HP HP 8360 L Series

  • Page 1 Refer to the “Instrument History” chapter. HP 83623L/3OL/4OL/5OL 3844A and below HEWLETT PACKARD HP Part No. 08380-90134 Printed in USA February 1999 Supersedes: September 1997...
  • Page 2 The information contained in this document is subject to change without notice. Hewlett-Packard makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
  • Page 3 Hewlett-Packard Company certifies that this product met its published specifications at the time of shipment from the factory. Hewlett-Packard further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standards Organization members.
  • Page 4 Product maintenance agreements and other customer assistance agreements are available for Hewlett-Packard products. For any assistance, contact your nearest Hewlett-Packard Sales and Service The following safety notes are used throughout this manual. Familiarize yourself with each of the notes and its meaning before operating this instrument.
  • Page 5 WARNING No operator serviceable parts inside. Refer servicing to qualified personnel. To prevent electrical shock, do not remove covers. For continued protection against fire hazard replace line fuse only with same type and rating (F 5A/250V). The use of other fuses or material is prohibited.
  • Page 6 Before switching on this instrument, make sure that the line CAUTION voltage selector switch is set to the voltage of the power supply and the correct fuse is installed. Always use the three-prong ac power cord supplied with this instrument. Failure to ensure adequate earth grounding by not using this cord may cause instrument damage.
  • Page 7 This manual provides user information for the HP 8360 L-Series Swept CW Generator. This manual applies to instruments having a serial number prefix listed on the title page (behind the “Documentation Map” tab). Some changes may have to be made to this manual so that it applies directly to each instrument;...
  • Page 8 The following conventions are used in the HP 8360 L-Series documentation: Italics Italic type is used for emphasis, and for titles of manuals and other publications. Computer Computer type is used for information displayed on the instrument. For example: In this sequence, POWER LEVEL is displayed.
  • Page 9 Model HP 8360 L-Series Swept CW Generator Hiermit wird bescheinigt, dass dieses Gerat/System in Note Der Deutschen Bundespost wurde das Inverkehrbringen dieses der Serie auf Einhaltung der Bestimmungen eingeraumt.
  • Page 10 Hewlett-Packard Co. Microwave Instruments Division 1400 Fountaingrove Parkway Santa Rosa, CA 95403-1799 Product Name: Synthesized Sweeper HP 83623L, HP 8363OL, HP 836401. Model Numbers: HP 83650L Product Options: This declaration covers all options of the above products. Safety: IEC 348:1978/HD 401 S1:1981...
  • Page 11 This is to declare that this instrument is in conformance with the German Regulation on Noise Declaration for Machines (Laermangabe Acoustic Noise Emmission/Geraeuschemission Operator position am Arbeitsplatz Normal position normaler Betrieb per IS) 7779 The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation.
  • Page 12 Table O-l. Hewlett-Packard Sales and Service Offices UNITED STATES Instrument Support Center Hewlett-Packard Company (800) 403-0801 EUROPEAN FIELD OPERATIONS Germany Headquarters France Hewlett-Packard S.A. Hewlett-Packard France Hewlett-Packard GmbH Hewlett-Packard Strasse 150, Route du Nant-d’Avri1 1 Avenue Du Canada 61352 Bad Homburg v.d.H 1217 Meyrin a/Geneva Zone D’Activite De Courtaboeuf F-91947 Les Ulis Cedex...
  • Page 13: Table Of Contents

    ..Equipment Used In Examples Introducing the HP 8360 L-Series Swept CW Generators ....Display Area ....
  • Page 14 Instrument Addresses ... . . 1-56 HP-IB Instrument Nomenclature ..1-56 Listener ....
  • Page 15 Use of the Command Tables ..l-90 ..HP-IB Check, Example Program 1 l-90 ... .
  • Page 16 The TRIG Configuration ..1-116 Description of Triggering in the HP 8360 L-Series Swept CW Generators ... l-117 Advanced Trigger Configurations ..
  • Page 17 (CENTER) ........Clear Fault .
  • Page 18 HP-IB Address ....HP-IB Menu ....Leveling .
  • Page 19 MI--M2 Sweep ..Manual Sweep ..MARKER Marker Mi ..Marker M2 ..Marker M3 ..Marker M4 ..Marker .
  • Page 20 Ref Osc Menu ........ROTARY KNOB ....
  • Page 21 ... . Auto T - l Freq Std ... T - l ... 10 MHz Freq Std Intrnl .
  • Page 22 2c. Specifications Frequency ....Range ....Resolution .
  • Page 23 Front Panel ... . . How to Prevent a Front Panel Change to an HP-IB ....
  • Page 24 Modules ....3-22 The HP 8970B Noise Figure Meter ..3-23 Remote Operation ... . .
  • Page 25 O-l. Typical Serial Number Label ..l-l. The HP 8360 L-Series Swept CW Generator . . . l-2. Display ....
  • Page 26 A-2. Typical External Leveling Hookup ..C-l. Auxiliary Interface Connector ..c-2. HP-IB Connector and Cable ..c-3. Interface Signals of the Source Module Connector .
  • Page 27 3-6. Rack Flange Kit for Swept CW Generators with Handles Attached Contents ..3-15 3-7. Instrument Preset Conditions for the HP ....3-20 3-24 3-8.
  • Page 28: Getting Started

    HP 8360 L-Series Swept CW Generator. Programming For the user wishing to program an HP 8360 L-Series Swept CW Generator. This section contains an introduction to Standard Commands for Programmable Instruments language (SCPI), Hewlett-Packard’s implementation of...
  • Page 29: How To Use This Chapter

    Equipment Used In Examples Recommended Equipment Model Numbers HP 436A/437B Power Meter Power I HP 8485A Sensor Oscilloscope mm-Wave Source Module HP 83556A HP 8349B Power Amplifier 1-2 Getting Started Introduction...
  • Page 30: Introducing The Hp 8360 L-Series Swept Cw Generators

    Getting Started Basic The HP 8360 L-Series Swept CW Generators are high performance, broadband frequency swept CW generators. H E W L E T T PACKP rR0 SYSTEM USER DrnNED PRESET Figure l-l. The HP 8360 L-Series Swept CW Generator (PRESET) initializes the front panel settings and runs the swept CW generator through a brief self-test.
  • Page 31: Display Area

    ACTIVE ENTRY AND DATA DISPLAY AREA Figure l-2. Display Active Entry and Data Display Area: This area typically displays the frequency and power information of the current instrument state. When data entry is expected, the swept CW generator uses all or part of this area to record the entries. The active entry arrow ( -->) indicates the active entry function and its current value.
  • Page 32: Entry Area

    All function values are changed via the rotary knob and/or keys of the entry area. ENTRY ENTRY ON ON/OFF ARROW KEYS ROTARY KNOB ENTRY KEYS BACKSPACE Figure l-3. Entry Area The following are active only when the swept CW generator expects an input.
  • Page 33: Cw Operation And Start/Stop Frequency Sweep

    CW Operation CW operation is one of the major functions of the swept CW generator, and is easy to do using front panel keys. In CW operation, the swept CW generator produces a single, low-noise, synthesized frequency. Try this example: Press m (i-J @ (J @ @ (?J @ 0 @ Check the active entry area.
  • Page 34 H E W L E T T P A C K A R D STOP SWEEP LED START Figure 1-4. CW Operation and Start/Stop Frequency Sweep s t a r t / s t o p CW Operation Frequency Sweep 1.
  • Page 35 Center Center frequency/span is another way of establishing swept operation. This is just a different way of defining sweep limits. As an example of center frequency/span operation: Press m@ (GHz). Press (SPAN) (iJ IGHz). The swept CW generator is now sweeping from 3.5 to 4.5 GHz (to view these figures, press either (START) or (STOP), then ml).
  • Page 36: Center Frequency/Span Operation

    . ’ CENTER SPAN SWEEP LED Figure 1-5. Center Frequency and Span Operation Span Operation Center Frequency Operation 1. Press @?TiiFj, 1. Press m). 2. Enter value. 2. Enter value. 3. Press terminator key. 3. Press terminator key. Getting Started Basic 1-9...
  • Page 37: Power Level And Sweep Time Operation

    The swept CW generator can produce leveled power for CW, swept Power Level Operation frequency, or power sweep operation. The selected power level can range from -20 dBm (-110 dBm for Option 001 swept CW generators) to -l-25 dBm. For practice: Press I-) @ @ m.
  • Page 38 H E W L E T T P A C K A R O POWER LEVEL SWEEP TIME SWEEP LED Figure 1-6. Power Level and Sweep Time Operation Sweep Time Power Level Operation Operation Press 1. Press S W E E P T I M E 2.
  • Page 39 Continuous sweep is the operation mode set when the swept CW generator is preset. It simply means that when the swept CW generator is performing a swept operation, the sweeps will continuously sweep-retrace-sweep-retrace until a different sweep mode is selected. To choose this sweep mode, press (CONT). To change from continuous sweep to single sweep operation, press (SINGLE).
  • Page 40: Continuous, Single, And Manual Sweep Operation

    H E W L E T T P A C K A R D SWEEP L E D - - - - SINGLE - SWEEP MENU SINGLE CONT Figure 1-7. Continuous, Single, and Manual Sweep Operation Manual Sweep Continuous Sweep Single Sweep 1.
  • Page 41: Marker Operation

    The swept CW generator has five frequency markers that can be used as fixed frequency “landmarks,” or as variable frequency pointers on a CRT display. To view the marker features of the swept CW generator on a CRT, connect the swept CW generator as shown in Figure l-8.
  • Page 42 Marker 1 was chosen because it is selected as the delta marker Delta Mkr Ref . reference. To change reference markers, select Select as the reference. Watch the display change to indicate: You can choose any of the five markers as a reference, but when delta marker is on, if the reference marker has a frequency value higher than the last active marker, the difference between the frequencies is negative and is displayed as such by the swept CW generator.
  • Page 43: Saving And Recalling An Instrument State

    The save/recall registers store and access a previously set instrument state. For example, set the swept CW generator to sweep from 3 to and 2 set at 4.5 and 11.2 GHz. Press (START) (?J LGHz). Press LSTOP) (iJ @ (GHz). Press ) (-) (iJ (TJ IdBo).
  • Page 44 SAVE RECALL H E W L E T T / I I PACK IARC Figure 1-9. Saving and Recalling an Instrument State Save 1. Set up swept CW generator as desired. 1. Press IjZKiJ. 2. Press ISAVE). 2. Press a number 0 through 8. 3.
  • Page 45: Power Sweep And Power Slope Operation

    The power sweep function allows the power output to be swept Power Sweep Operation (positive or negative) when the swept CW generator is in the CW frequency mode. The power output of the swept CW generator determines the maximum leveled power sweep that can be accomplished.
  • Page 46: Power Slope Operation

    Select power Sweep (asterisk on). Press @GE). The swept CW generator performs a power sweep beginning at -20 dBm and ending at +5 dBm. The power meter indicates This function allows for compensation of high frequency system or Power Slope Operation cable losses by linearly increasing the power output as the frequency increases.
  • Page 47: Getting Started Advanced

    Advanced This section of Chapter 1 describes the use of many of the unique features of the HP 8360 L-Series Swept CW Generators. The format used is similar to the one used on the previous pages. When referred to a menu map number, go to the Menu Map tab and unfold the menu map so that you can view it together with the text.
  • Page 48 Advanced Table 1-1. Keys Under Discussion in This Section (continued) Paragraph Heading Keys ‘Optimizing Swept CW Generator Performance” Auto Track Peak RF Always Peak RF Once USER DEFINED MENU ASSIGN Step Sup Menu Jsing Step Sweep List Menu Delete Menu Enter List Freq E n t e r L i s t O f f s e t Enter List Dwell...
  • Page 49: Externally Leveling The Swept Cw Generator

    In externally leveled operations, the output power from the swept CW generator is detected by an external sensor. The output of this detector is returned to the leveling circuitry, and the output power is automatically adjusted to keep power constant at the point of detection.
  • Page 50 Note Figure 1-12 shows the input power versus output voltage characteristics for typical HP diode detectors. From the chart, the leveled power at the diode detector input resulting from any external level voltage setting may be determined. The range of power adjustment is approximately -30 dBm to +18 dBm.
  • Page 51: Typical Diode Detector Response At 25°C

    0 dBV - 1 0 dBV - 2 0 dBV 100 mV - 3 0 dBV - 4 0 dBV 10 mV - 5 0 dBV - 6 0 dBV 1 mV --66 dBV -70 dBV -80 dBV - 3 0 - 2 0 - 1 0 DETECTOR INPUT POWER, dBm...
  • Page 52: External Leveling Used With The Optional Step Attenuator

    External Leveling Used With the Optional Step Attenuator Some external leveling applications require low output power from the swept CW generator. The swept CW generator automatically uncouples the attenuator from the ALC system for all external leveling points. Press Note the display. It shows: POWER LEVEL For example, leveling the output of a 30 dB gain amplifier to a...
  • Page 53: Leveling With Power Meters

    Leveling with Power Leveling with a power meter is similar to leveling with a diode detector. Figure 1-13 shows the setup for power meter leveling. Meters GENERATOR LEVELED OUTPUT POUER SPLITTER Figure 1-13. Leveling with a Power Meter 1. Set up the equipment as shown. Be sure to set the power meter to manual range mode and note the range.
  • Page 54: Leveling With Mm-Wave Source Modules

    Leveling with MM-wave Millimeter-wave source module leveling is similar to power meter leveling. The following figures illustrate the setups for leveling with a Source Modules mm-wave source module. W E P T Cl4 Figure 1-14. MM-wave Source Module Leveling High power model swept CW generators can externally level mm-wave source modules to maximum specified power without a microwave amplifier.
  • Page 55 LEVELED OUTPUT Figure 1-15. MM-wave Source Module Leveling Using a Microwave Amplifier 1. Set up the equipment as shown. 2. Refer to menu map 1. 3. Select Leveling Point Module. where the interface connection is made. All of the ALC data necessary to communicate properly with the swept CW generator is exchanged via the SOURCE MODULE INTERFACE.
  • Page 56: Working With Mixers/Reverse Power Effects

    Note Uncoupled operation applies to Option 001 swept CW generators only. Uncoupled operation is useful when working with mixers. Figure 1-16 shows a hypothetical setup where the swept CW generator is providing a small signal to a mixer. The swept CW generator output is -8 dBm, which in Leveling Mode Normal results in ATTEN = 0 dB, ALC Level = -8 dBm.
  • Page 57: Reverse Power Effects, Coupled Operation With 8Dbm Output

    LO FEED- THROUGH - DETECTOR DETECTOR MEASURES -8 dBm Figure 1-16. Reverse Power Effects, Coupled Operation with -8dBm Output RF OUTPUT MIXa LO LEVEL DETECTOR MEASURES -15 dBm MEASURES +2 dBm REVERSE POWER Figure 1-17. Reverse Power Effects, Uncoupled Operation with -8dBm Output Getting Started Advanced 1-3 1...
  • Page 58: Working With Spectrum Analyzers/Reverse Power Effects

    Reverse power is a problem with spectrum analyzers that do not have preselection capability. Some analyzers have as much as frequencies. The effects of reverse power are less in the heterodyne band (0.01 to 2.0 GHz) where the power amplifier provides some broadband matching.
  • Page 59: Optimizing Swept Cw Generator Performance

    There are two basic front-panel methods of creating a flatness correction array. The first and quickest method is to use an HP 437B power meter. Refer to Figure 1-18 for the setup. The second method is just as accurate, but requires a little more interaction between the operator and the instruments.
  • Page 60: Creating A User Flatness Array Automatically, Example 1

    Creating a User Flatness Array Automatically, Example 1 In this example, a flatness array containing correction frequencies from 4 to 10 GHz at 1 GHz intervals is created. An HP 438B power meter controlled by the swept CW generator through the interface is used to enter the correction data into the flatness array.
  • Page 61 If an HP-IB error message is displayed, verify that the interface connections are correct. Check the HP-IB address of the power meter and ensure that it is the same address the swept CW generator is using (address 13 is assumed).
  • Page 62: Creating A User Flatness Array, Example 2

    Creating a User Flatness Array, Example 2 This example shows how to use the swept CW generator and a power meter in manual entry mode. This example also introduces two features of the swept CW generator. The softkey Freq Follow simplifies the data entry process and the softkey List Mode sets up a list of arbitrary test frequencies.
  • Page 63: Creating A User Flatness Array

    GENERATOR POUER METER , AND OTHER , DEVICES , OUTPUT PORT DEVICE UNDER TEST Figure 1-19. Creating a User Flatness Array For this example, refer to menu map 5, POWER. 1. The equipment setup shown in Figure 1-19 assumes that if your setup has an external leveling configuration, the steps necessary to correctly level have been followed.
  • Page 64: Power Meter

    Access User Flatness Correction Menu 9. Press POWER (E). Select Fitness Menu. flatness array is empty. 11. Press (El. Leave the delete menu and return to the previous soft key menu. 12. Select Copy List This step copies the frequency list into the correction table in sequential order.
  • Page 65: Swept Mm-Wave Measurement With Arbitrary Correction Frequencies, Example 3

    Example 3 The focus of this example is to use user flatness correction to obtain flat power at the output of the HP 83550 series mm-wave source modules. In this case we will use non-sequential correction frequencies in a swept 26.5 to 40 GHz measurement with an HP 83554 source module.
  • Page 66 H P 4 3 7 8 GENERRTOR POUER METER POUER SENSOR INTERFACE SUEPT CU H P 4 3 7 8 GENERATOR POWER flEfER Figure l-20. Creating Arbitrarily Spaced Frequency-Correction Pairs in a Swept mm-wave Environment For this example, refer to menu map 5, POWER. 1.
  • Page 67 If an HP-IB error message is displayed verify that the interface connections are correct. Check the HP-IB address of the power meter and ensure that it is the same address the swept CW...
  • Page 68 generator is using (address 13 is assumed). Refer to the menu map 8, System, for the key sequence necessary to reach softkey Meter Adrs . Enable User Flatness Correction 13. When the operation is complete, (a message is displayed) the flatness correction array is ready to be applied to your setup.
  • Page 69: Scalar Analysis Measurement With User Flatness Corrections, Example 4

    Example 4 The following example demonstrates how to set up a scalar analysis measurement (using an HP 8757 Scalar Network Analyzer) of a 2 to 20 GHz test device such as, an amplifier. User flatness correction is used to compensate for power variations at the test port of a directional bridge.
  • Page 70 The user flatness correction array cannot be stored to a disk. You must make sure that the array is stored in one of the eight internal registers. Recalling a file from an HP 8757 disk will not erase the current array; therefore you may recall an array from an internal register, then recall an associated file from a disk.
  • Page 71 6. Turn off the HP 8757 System Interface. Use the analyzer SYSINTF ON OFF softkey found under the SYSTEM menu to deactivate the system interface. Access User Flatness Correction Menu 7. On the swept CW generator, press POWER [MENU). Select Fltness Menu.
  • Page 72 20. To save the swept CW generator parameters including the correction table in an internal register, press (SAVE) @ (n = number 1 through 8). Reactivate the HP 8757 System Interface 21. Set the analyzer to SYSINTF ON, the analyzer and swept CW generator preset.
  • Page 73: Using Detector Calibration

    3. Enter the appropriate power sensor calibration factors into the power meter. 4. Enable the power meter/sensor cal factor array. For operating information on the HP 437B power meter, refer to its operating and service manual. 5. Connect the power sensor to the output of the coupler (or splitter).
  • Page 74 If an HP-IB error message is displayed verify that the interface connections are correct. Check the HP-IB address of the power meter and ensure that it is the same address the swept CW generator is using (address 13 is assumed). Refer to the menu map 8, System, for the key sequence necessary to reach softkey Meter Adrs.
  • Page 75: Using The Tracking Feature

    Using the Tracking Peaking Feature Peaking is the function that aligns the output filter (YTM) so that its passband is centered on the RF output, in CW or manual-sweep mode. Use peaking to obtain the maximum available power and spectral purity, and best pulse envelopes, at any given frequency above 2.0 GHz.
  • Page 76: Alc Bandwidth Selection

    The ALC bandwidth defaults at factory preset to the auto selection ALC Bandwidth Select Auto which selects the appropriate bandwidth [high or low) for each application. To make the bandwidth sel&tion, thk swept CW generator determines which functions are activated and uses the decision tree shown in Figure l-23.
  • Page 77: Using Step Sweep

    1. Refer to menu map 2. 2. Press FREQUENCY [MENU). Step Swp Menu. Select Step Size. Enter the desired increment value. Dwell and enter a value, or choose the dwell time determined by the ramp mode sweep time, select Dwell Coupled. Step Swp Pt Trig Auto , 8.
  • Page 78: Creating And Using A Frequency List

    Refer to menu map 2. 2. Press FREQUENCY (MENU). To use the frequency points of a frequency list to create the frequency portion of the user flatness correction array: Refer to menu map 5. 2. Press POWER (=I. Menu. 1-52 Getting Started Advanced...
  • Page 79: Using The Security Features

    To access the security menu: 1. Refer to menu map 8. 2. Press SYSTEM (MENU). Getting Started Advanced l-53...
  • Page 80: Changing The Preset Parameters

    1. Set up the swept CW generator in the desired operation state to be used as the preset state. 2. Refer to menu map 8. 3. Press SYSTEM (MENU). Whenever the (w] key is pressed, the swept CW generator will return to the operation state setup and saved in steps 1 and 4.
  • Page 81: Getting Started Programming

    HP-IB, the Hewlett-Packard Interface Bus, is the instrument-to- instrument communication system between the swept CW generator and up to 14 other instruments. Any instrument having HP-IB capability can be interfaced to the swept CW generator, including non-HP instruments that have “GPIB,” “IEEE-488,” “ANSI MC1.l,”...
  • Page 82: Hp-Ib General Information

    Figure C-2, must be observed. Instrument Addresses Each instrument in an HP-IB network must have a unique address, ranging in value from 00-30 (d ecimal). The default address for the swept CW generator is 19, but this can be changed using the My Adrs chapter (Chapter 2) under the “8360 Adrs”...
  • Page 83: Hp-Ib Command Statements

    Consider the following explanations as a starting point, but for detailed information consult the BASIC language reference manual, the I/O programming guide, and the HP-IB manual for the particular computer used. Syntax drawings accompany each statement: All items enclosed by a circle or oval are computer specific terms that must be entered exactly as described;...
  • Page 84: Remote

    REMOTE annunciator is lighted. The syntax is: where the device selector is the address of the instrument appended to the HP-IB port number. Typically, the HP-IB port number is 7, and the default address for the swept CW generator is 19, so the device selector is 719.
  • Page 85: Local

    Related statements used by some computers: RESUME Clear Clear causes all HP-IB instruments, or addressed instruments, to assume a “cleared” condition, with the definition of “cleared” being unique for each device. For the swept CW generator: 1. All pending output-parameter operations are halted.
  • Page 86 Some BASIC examples: CLEAR 7 to clear all HP-IB instruments, or CLEAR 719 to clear an addressed instrument. Related statements used by some computers: RESET CONTROL SEND The preceding statements are primarily management commands that do not incorporate programming codes. The following two statements do incorporate programming codes, and are used for data communication.
  • Page 87: Output

    A BASIC example: 100 OUTPUT719; “programming codes” The many programming codes for the swept CW generator are listed in the “SCPI Command Summary” in Chapter 2. Related statements used by some computers: CONTROL CONVERT IMAGE IOBUFFER TRANSFER Enter Enter is the complement of OUTPUT, and is used to transfer data from the addressed instrument to the controller.
  • Page 88 ASCII CR (carriage return), comma, or semicolon might cause a false termination. Suppression of the EOL causes the computer to accept all bit patterns as data, not commands, and relies on the HP-IB EOI (end or identify) line for correct end-of-data termination.
  • Page 89: Getting Started With Scpi

    An instrument is any device that implements SCPI. Most instruments are electronic measurement or stimulus devices, but this is not a requirement. Similarly, most instruments use an HP-IB interface for communication. The same concepts apply regardless of the instrument function or the type of interface used.
  • Page 90: Standard Notation

    If you encounter problems, study the details of how the output statement handles message terminators such as <new line>. If you are using simple OUTPUT statements in HP BASIC, this is taken care of for you. In HP BASIC, you type: OUTPUT Source ;...
  • Page 91: Response Examples

    <new line> and EOL. To enter the previous response in HP BASIC, you type: ENTER Source;CW-frequency Response examples do not show response message terminators because they are always <new line>...
  • Page 92: Essentials For Beginners

    This subsection discusses elementary concepts critical to first-time users of SCPI. Read and understand this subsection before going on to another. This subsection includes the following topics: Program and Response These paragraphs introduce the Messages basic types of messages sent between instruments and controllers.
  • Page 93: Types Of Commands

    Types of Commands Commands can be separated into two groups, common commands Common commands and subsystem commands. are generally not measurement related. They are used to manage macros, status registers, synchronization, and data storage. Common commands are easy to recognize because they all begin with an asterisk, such as *IDN?, *OPC, and *RST.
  • Page 94: Subsystem Command Trees

    Subsystem Command The Command Tree Structure Trees Most programming tasks involve subsystem commands. SCPI uses a hierarchical structure for subsystem commands similar to the file systems on most computers. In SCPI, this command structure is command tree. called a root level 1 level EE FF GG...
  • Page 95 Colon When it is between two command mnemonics, a colon moves the current path down one level in the command tree. For example, the colon in MEAS:VOLT specifies that VOLT is one level below MEAS. When the colon is the first character of a command, it specifies that the next command mnemonic is a root level command.
  • Page 96 Sets current path to ROOT NO change to current path Set current path DOWN one level :AA:BB:EE; :M:DD:JJ Figure l-26. Proper Use of the Colon and Semicolon In Figure l-26, notice how proper use of the semicolon can save typing. Sending this message: :AA:BB:EE;...
  • Page 97: Subsystem Command Tables

    Subsystem Command These paragraphs introduce a more complete, compact way of documenting subsystems using a tabular format. The command Tables table contains more information than just the command hierarchy shown in a graphical tree. In particular, these tables list command parameters for each command and response data formats for queries.
  • Page 98: More About Commands

    the matching command. The parameter type is listed adjacent to each named parameter. More About Commands Query and Event Commands. Because you can query any value that you can set, the query form of each command is not shown explicitly in the command tables.
  • Page 99: Parameter Types

    Example 1: “FREQuency : CW 5 GHZ ; MULTiplier 2” The command is correct and will not cause errors. It is equivalent to sending: “FREQuency : CW 5 GHZ ; : FREQuency : MULTiplier 2”. Example 2: “FREQuency 5 GHZ; MULTiplier 2” This command results in a command error.
  • Page 100: Numeric Parameters

    Examples of numeric parameters: no decimal point required fractional digits optional 100. leading signs allowed -1.23 space allowed after e in exponents use either E or e in exponentials leading + allowed digits left of decimal point optional Examples of numeric parameters in commands: 100 OUTPUT @Source;": FREQuency:STARt l.OE+09"...
  • Page 101: Discrete Parameters

    Discrete Parameters. Use discrete parameters to program settings that have a finite number of values. Discrete parameters use mnemonics to represent each valid setting. They have a long and a short form, like command mnemonics. You can use mixed upper and lower case letters for discrete parameters.
  • Page 102: Reading Instrument Errors

    When debugging a program, you may want to know if an instrument Reading Instrument error has occurred. Some instruments can display error messages on Errors their front panels. If your instrument cannot do this, you can put the following code segment in your program to read and display error messages.
  • Page 103: Example Programs

    The following is an example program using SCPI compatible Example Programs instruments. The example is written in HP BASIC. This example is a stimulus and response application. It uses a source and counter to test a voltage controlled oscillator. Example Program Description.
  • Page 104: Program Comments

    OUTPUT and ENTER statements, instead of a numeric address. 80 to 100: Assign values to the input test limits in mV. 110 to 130: Clear the instrument HP-IB interfaces. 140 to 160: Reset each instrument to a known measurement state.
  • Page 105 470 to 480: Disconnect output terminals of the instruments from the unit under test, and end the program. All HP BASIC programs must have END as the last statement of the main program. Getting Started Programming 1-79...
  • Page 106: Details Of Commands And Responses

    In This Subsection This subsection describes the syntax of SCPI commands and responses. It provides many examples of the data types used for command parameters and response data. The following topics are explained: Program Message These paragraphs explain how to properly Syntax construct the messages you send from the computer to instruments.
  • Page 107: Subsystem Command Syntax

    The word means that EOI is asserted on the HP-IB interface at the same time the preceding data byte is sent. Most programming languages send these terminators automatically. For example, if you use the HP BASIC...
  • Page 108: Response Message Syntax

    NOTE: Figure 1-31. Simplified Common Command Syntax As with subsystem commands, use a <space> to separate a command mnemonic from subsequent parameters. Separate adjacent parameters with a comma. Parameter types are explained later in this subsection. Figure l-32 shows a simplified view of response message syntax. Response Message Syntax Figure l-32.
  • Page 109: Scpi Data Types

    These paragraphs explain the data types available for parameters and response data. They list the types available and present examples for each type. SCPI defines different data formats for use in program messages and response messages. It does this to accommodate the principle of forgiving listening and precise talking.
  • Page 110: Extended Numeric Parameters

    rounds the parameter. For example, if an instrument has a programmable output impedance of 50 or 75 ohms, you specified instrument setting can only assume integer values, it automatically rounds the value to an integer. For example, sending *ESE 10.123 is the same as sending *ESE 10.
  • Page 111: Discrete Parameters

    discrete parameters Discrete Parameters. Use to program settings that have a finite number of values. Discrete parameters use mnemonics to represent each valid setting. They have a long and a short form, just like command mnemonics. You can used mixed upper and lower case letters for discrete parameters.
  • Page 112: Integer Response Data

    Integer Integer Response Data. response data are decimal representations of integer values including optional signs. Most status register related queries return integer response data. Examples of integer response data: signs are optional leading + sign allowed leading sign allowed -100 never any decimal point Discrete Discrete Response Data.
  • Page 113: Programming Typical Measurements

    “Program Comments” paragraphs to follow the programmed activity. The HP-IB select code is assumed to be preset to 7. All example programs in this section expect the swept CW generator’s HP-IB address to be decimal 19.
  • Page 114: Use Of The Command Tables

    Use of the Command Tables In Table 1-4, notice that a new column titled “Allowed Values” has been added to the command table. This column lists the specific values or range of values allowed for each parameter. A vertical bar The commands listed in the table are only part of all the available SCPI commands of the swept CW generator.
  • Page 115 Table 1-4. Sample Swept CW Generator Commands (continued) Parameters Parameter Type Allowed Values Command extended numeric 0 to 90 [DB] or Boolean coupled atten specified power range output level extended numeric Boolean RF on/off type of sweep discrete 200s to 133 ms or extended numeric sweep time auto sweep...
  • Page 116: Hp-Ib Check, Example Program 1

    Operating and Service Manuals to find the defective device. Program Comments 10: Set up a variable to contain the HP-IB address of the source. 20: Abort any bus activity and return the HP-IB interfaces to their reset states.
  • Page 117: Local Lockout Demonstration, Example Program

    When the swept CW generator is in REMOTE mode, all the front Local Lockout panel keys are disabled except the LOCAL key. But, when the Demonstration, LOCAL LOCKOUT command is set on the bus, even the LOCAL Example Program 2 key is disabled.
  • Page 118 Note that the swept CW generator Ii??%) key produces the same HINT result as programming LOCAL 719 or LOCAL 7. Be careful because the LOCAL 7 command places all instruments on the HP-IB in the local state as opposed to just the swept CW generator. Program Comments 90 to 120: Print a message on the computer’s display, then pause.
  • Page 119: Program Comments

    PRINT "with a sweeptime of 0.5 seconds." Run the program. Program Comments 10: Assign the source’s HP-IB address to a variable. to 50: Abort any HP-IB activity and initialize the HP-IB interface. 60: Set the source to its initial state for programming.
  • Page 120 110: Set the sweeptime to 500 ms. Notice that upper/lower case in commands does not matter. Also spaces before the suffix (“MS”) are not required in SCPI. 120 and 130: Set markers 1 and 2 to a fixed value. Notice that the value for marker 2 does not end with a frequency suffix.
  • Page 121: Program Comments

    Run the program. Program Comments 10: Assign the source’s HP-IB address to a variable. 20 to 50: Abort any HP-IB activity and initialize the HP-IB interface. 60: Clear the computer’s display. 70: Set the source to its initial state for programming.
  • Page 122 130: Enter the query response into a variable ‘W’. Boolean responses are always ‘1’ for ON and ‘0’ for OFF. 140: Print the value of the POWER:STATE on the computer display. 150: Query the value of a discrete function (FREQ:MODE). 160: Dimension a string variable to contain the response.
  • Page 123: Program Comments

    Run the program. Program Comments 10: Assign the source’s HP-IB address to a variable. 20 to 50: Abort any HP-IB activity and initialize the HP-IB interface. 60: Clear the computer’s display. 70: Set up the source for a sweeping state. Note the combination of several commands into a single message.
  • Page 124 90: Clear the computer display. 100: Print a message on the computer display. 110: Set up the source for a CW state. Note the combination of several commands into a single message. This single line is equivalent to the following lines: OUTPUT Source;"*RST"...
  • Page 125: Program Comments

    240 END Run the program. Program Comments 10: Assign the source’s HP-IB address to a variable. 20 to 50: Abort any HP-IB activity and initialize the HP-IB interface. 60: Clear the computer’s display. Set the source to its initial state programming.
  • Page 126 140 and 150: Prompt the operator for the number of sweeps to take. The number of sweeps to take is stored in the variable N. Enter 0 to quit the program. 160: Don’t take any sweeps if N is less than 0. 170: Start a FOR/NEXT loop to take N sweeps.
  • Page 127: Program Comments

    Run the program. Program Comments 10: Assign the source’s HP-IB address to a variable. 20 to 50: Abort any HP-IB activity and initialize the HP-IB interface. 60: Clear the computer’s display. 70: Set the source to its initial state for programming.
  • Page 128 170: Send another *WA1 to the source. Although the ‘WA1 command causes EXECUTION of commands to be held off, it has no effect on the transfer of commands over the HP-IB. The commands continue to be accepted by the source and are buffered until they can be executed.
  • Page 129: Using The User Flatness Correction Commands, Example Program

    2 to 20 with frequency-correction pairs every 100 MHz and path losses do not exceed 5 dBm and that the HP 437B power meter already has its power sensor’s calibration factors stored in sensor data table 0. If another power meter is used, the power sensor’s calibration factors will have to be stored in a look-up table.
  • Page 130 FOR I=1 TO N Freq=Freq+Increment NEXT I B=B-1 OUTPUT QSource;"CORR:FLAT ";B$ ! OUTPUT QSource;"POW:STAT ON" OUTPUT OSource;"CAL:PMET:FLAT:INIT? USER" ENTER OSource;Freq WHILE Freq>O OUTPUT OSource;"CAL:PMET:FLAT:NEXT? ";VAL$(Power);"DBM" ENTER QSource;Freq END WHILE END IF SUB Zero-meter(QMeter,OSource,INTEGER Error-flag) OUTPUT OSource;"Pow:stat off" OUTPUT OMeter;"CS" OUTPUT QMeter;"ZE" Attempts=0 Zeroing=1590 Finished=0...
  • Page 131 REPEAT OUTPUT QMeter;"TR2" ENTER OMeter;Power$ IF Slope2Slope THEN Flips=Flips+l ELSE IF SlopeZ=O THEN Flips=Flips+.2 1000 END IF 1010 1020 UNTIL Flips>=3 1030 1040 RETURN Power 1050 FNEND Getting Started Programming l-105...
  • Page 132: Programming The Status System

    This subsection discusses the structure of the status system used in In This Subsection SCPI instruments, and explains how to program status registers. An important feature of SCPI instruments is that they all implement status registers the same way. The status system is explained in the following paragraphs: General Status These paragraphs explain the way that status Register Model registers are structured in SCPI instruments.
  • Page 133: Transition Filter

    There may or may not be a command to read a particular condition register. Transition Filter transition filter specifies which types of bit state changes in the condition register will set corresponding bits in the event register. Transition filter bits may be set for positive transitions (PTR), negative transitions (NTR), or both.
  • Page 135: Programming The Trigger System

    This subsection discusses the layered trigger model used in SCPI In This Subsection instruments. It also outlines some commonly encountered trigger configurations and programming methods. Trigger system topics are explained in the following paragraphs: Generalized Trigger These paragraphs explain the structure and Model components of the layered trigger model used in all SCPI instruments.
  • Page 136: Details Of Trigger States

    An instrument moves between adjacent states, depending on its internal conditions and the commands that you send. When you first turn on power to an instrument, it is in the idle state. You can force the instrument to the idle state using : ABORt The initiate and event detection trigger states are essentially a list of conditions...
  • Page 137: Inside The Idle State

    idle Inside the Idle State. Figure l-36 illustrates the operation of the state. Figure l-36. Inside the Idle State Turning power on, or sending *RST or :ABORT forces the trigger system to the idle state. The trigger system remains in the idle state until it is initiated by INITiate : IMMediate or initiate the trigger system exits downward to the...
  • Page 138: Inside Event Detection States

    is on an upward path, and INITiate: CONTinuous is ON, it exits downward to an event-detection state. If the trigger system is on an upward path and 1NITiate:CONTinuous is OFF, it exits upward to the idle state. Inside Event Detection States. Figure l-38 illustrates the operation of an arbitrary event detection state named <state-name>.
  • Page 139 qualified event detection etc *’ Figure l-38. Inside an Event Detection State Getting Started Programming...
  • Page 140: Inside The Sequence Operation State

    Inside the Sequence Operation State. Figure l-39 illustrates the operation of the sequence operation state. The downward entrance to the Sequence Operation State signals that some instrument dependent action should begin at once. An upward exit is not allowed until the instrument signals that its action is complete.
  • Page 141: Common Trigger Configurations

    In the previous paragraphs, you learned about the basic building Common Trigger blocks allowed in a SCPI trigger system. Generally, an instrument Configurations implements only a portion of the trigger features available. These paragraphs discuss the simplest configurations: INIT and TRIG. The INIT Configuration The 1NIT configuration is the simplest possible trigger configuration.
  • Page 142: The Trig Configuration

    The TRIG Configuration TRIG Instruments using the configuration include one event detection state named TRIG, and a corresponding TRIGger subsystem. And, all SCPI instruments implement the required Idle EXT >- BUS >- Initiate TRIG Event Detection Sequence Actions Operation Figure 1-41. The TRIG Trigger Configuration l-1 16 Getting Started Programming...
  • Page 143: Description Of Triggering In The Hp 8360 L-Series Swept Cw Generators

    The HP 8360 L-Series Swept CW Generators follow the SCPI model Description of of triggering. It is a layered model with the structure shown in Triggering in the Figure l-42. HP 8360 L-Series Swept CW Generators Idle State Sweep Initiated...
  • Page 144: Advanced Trigger Configurations

    to jump to the next point when the signal becomes TRUE, therefore the first point in the list or stepped sweeps is produced immediately upon starting the sweep. Receiving a trigger signal at the last point causes the IDLE state to be re-entered. Analog sweeps do not use the trigger signals during the sweep (although the trigger signals are needed to start the sweep as described).
  • Page 145: Source

    The most commonly used sources are: The event detector is satisfied by either Group Execute command. <GET> is a low level HP-IB message that can be sent using the TRIGGER command in HP BASIC. An external signal connector is selected as the source.
  • Page 146: Related Documents

    Company This HP BASIC manual contains a good non-technical description of the HP-IB (IEEE 488.1) interface in chapter 12, “The HP-IB Interface”. Subsequent revisions of HP BASIC may use a slightly different title for this manual or chapter. This manual is the best reference on instrument I/O for HP BASIC programmers.
  • Page 147: Operating And Programming Reference

    For operator’s service information, see the Chapter 4, “Operator’s Check and Routine Maintenance”. The operator accessible (SERVICE) HP 8360 B-Series Swept Signal Generator/HP 8360 L-Series Swept CW Generator Service Guide. Programming Language Table 3-9 cross-references the actions that the instrument can...
  • Page 148: Address

    This softkey accesses the HP-IB address menu. Controls the system power meter address. Meter Adrs Can control the swept CW generator’s address, 8360 Adrs depending on the setting of the rear panel HP-IB switch. Printer Adrs Controls the system printer address. Operating and Programming Reference A-l...
  • Page 149 SCPI: NONE, see the individual softkeys listed. Programming Codes Analyzer: NONE HP-IB Menu, softkeys listed above. See Also “Optimizing Swept CW Generator Performance” in Chapter 1 “HP-1B Address Selection” in Chapter 3 A-2 Operating and Programming Reference...
  • Page 150: Alc

    Leveling Point Module Sets the swept CW generator to level power at the output of a millimeter-wave module. Either an HP 8349B or 8355X series millimeter-wave source module must be connected to the SOURCE MODULE INTERFACE. L e v e l i n g P o i n t PwrMtr Sets the swept CW generator to level power at an external power meter.
  • Page 151 Specifies the operating range of an Pwr Mtr Range external power meter used in an external leveling setup. This causes the swept CW generator display to agree with the power meter’s power indication. The following paragraphs explain the power control (leveling) function of the swept CW generator in detail.
  • Page 152 HP 836iliL SERIES SWEPT CW GENERATOR I - - - - e - - m - - m - - m - - m - - - - - DISPLAY KEYBOARD ( F R O T - P A N E L...
  • Page 153 Note Two terms are used in the following discussions: power output and ALC level. Power output means actual output power including the effects of the attenuator. ALC level means power levels before the attenuator. In swept CW generators without attenuators, these two terms are equivalent.
  • Page 154 the ALC level and attenuator are set automatically to POWER LEVEL provide the most accuracy for the power requested. Uncoupled Operation. In some applications it is advantageous to control the ALC level and attenuator separately, using combinations of settings that are not available in coupled operation. In uncoupled mode (Uncoupl Atten ), when the desired power output is set via only the ALC level is changed.
  • Page 155 GENERRTOR RF OUTPUT N E G A T I V E DETECTOR Figure A-2. Typical External Leveling Hookup ALC Disabled Off. In this configuration, the ALC is disabled, power is not sensed at any point, and therefore the absolute power level is uncalibrated (see Figure A-l).
  • Page 156 4. While monitoring the internal detector, the RF modulator level is varied until the detected power is equivalent to the reference power measured in step 2. 5. Modulation is re-enabled if appropriate. These steps are performed in approximately 200 ps and are repeated any time power or frequency is changed.
  • Page 157: Alc Bandwidth Select Auto

    ALC Bandwidth Select Auto Function Group M e n u M a p 1 This softkey sets the swept CW generator to choose the ALC Description bandwidth automatically depending on the current sweep and modulation conditions. An asterisk next to the key label indicates that this feature is active.
  • Page 158: Alc Bandwidth Select Low

    ALC 3W Menu ALC Bandwidth Select Low Function Group Menu Map This softkey sets the swept CW generator to the ALC low bandwidth Description position (10 kHz). In this mode, the ALC bandwidth operates in a narrow bandwidth for all sweep and modulation conditions. An asterisk next to the key label indicates that this feature is active.
  • Page 159: Altrnate Regs

    ALC BW Menu ALC Bandwidth Select Low Sets the ALC bandwidth to the low bandwidth position (10 kHz), and to remain there for all sweep and modulation conditions. S e e A l s o “Optimizing Swept CW Generator Performance” in Chapter 1 Altrnate Regs SYSTEM Function Group...
  • Page 160: Amp1 Markers

    This status structure is the structurally and syntactically the same as on the HP 8340/41. bytes from the swept CW generator. The first status byte concerns the cause of an SRQ (Service Request), while the...
  • Page 161 Bit 4: SRQ on end-of-sweep or mid-sweep update in NA (network analyzer code) mode. Bit 5: SRQ caused by HP-IB syntax error. Bit 6: SERVICE REQUEST; by IEEE-488 convention, the instrument needs service from the controller when this bit is set true.
  • Page 162 ANALYZER STATUS REGISTER Bit 2: Oven for the reference crystal oscillator is not at operating temperature. Bit 3: External reference frequency is selected. Bit 4: RF is unlocked (UNLOCK appears in the message line). Use OF to determine the source of the unlocked output. This bit remains latched until this status byte has been read, or until cleared by the CS or CLEAR 719 commands.
  • Page 163: Arrow Keys

    Function Group ENTRY Menu Map NONE This group of entry keys lets you manipulate numerical values in the Description active entry line. changed. For example, if CW frequency is in the active entry line, and the display indicates: 10005.000000 MHz you may wish to change the 5 to a 6.
  • Page 164 (ASSIGN) (ASSIGN) Function Group USER DEFINED Menu Map NONE Description This hardkey lets you select any softkey and assign its function to 1 of 12 user defined keys in the Menu. The following USER DEFINED message appears on the swept CW generator display: --> Press MENU KEY to be assigned.
  • Page 165: Auto Fill Incr

    Auto Fill Incr Function Group FREQUENCY, POWER Menu Map This softkey is used in two locations: Fltness Menu and Description List Menu. Flatness Menu - When selected, the swept CW generator waits for a frequency increment value to be entered. --> Increment : displayed in the active entry area.
  • Page 166: Auto Fill #Pts

    Auto Fill #Pts Auto Fill #Pts Function Group FREQUENCY, POWER M e n u M a p 2,s Menu This softkey is used in two locations: Fitness Description List Menu. Flatness Menu - When selected, the swept CW generator waits for a numeric value representing the number of correction points to be entered.
  • Page 167: Auto Fill Start

    Auto Fill Start Function Group FREQUENCY, POWER M e n u M a p 2,s This softkey is used in two locations: Fitness Menu and Description List Menu. The operation is the same in both applications. This softkey enables the entry of a start frequency used to determine the beginning frequency of the automatic filling array.
  • Page 168: Auto Fill Stop

    Auto Fill Stop Auto Fill Stop Function Group FREQUENCY, POWER Menu Map Description This softkey is used in two locations: Fltness Menu and List Menu. The operation is the same in both applications. This softkey enables the entry of a stop frequency used to determine the ending frequency of the automatic filling array.
  • Page 169: Auto Track

    Auto Track POWER, USER CAL Function Group M e n u M a p 5,9 This softkey optimizes the tracking of the swept CW generator’s Description output filter to the oscillator. Use it to maximize RF power output. The swept CW generator displays: Peaking At : XXXXX MHz, where XXXXX represents frequency values.
  • Page 170: Blankdisp

    Blank Disp SYSTEM Function Group M e n u M a p 8 When this softkey is selected, it causes the top four lines of the Description display to blank and remain blank until the @ZZi) key is pressed. Blanking the display prevents sensitive information from being displayed.
  • Page 171 Function Group FREQUENCY NONE Menu Map Description This hardkey lets you select the center frequency for center frequency/frequency span swept operation. When you press (CENTER), XXXXX MHz where the swept CW generator displays: XXXXX represents a frequency value. Use the entry area to set the desired value.
  • Page 172: Center=Marker

    MARKER Function Group M e n u M a p 3 This softkey sets the center frequency of the sweep to the frequency Description of the most recently activated marker. Select any marker Ml . . . of the sweep to that of the marker. The frequency span does not change unless the new sweep limits fall outside the frequency range of the swept CW generator.
  • Page 173: Clear Memory

    Clear Memory Clear Memory SYSTEM Function Group Menu Map Description This softkey causes the swept CW generator to return to the factory preset instrument state, after writing alternating ones and zeroes over all state information, frequency lists, and save/recall registers. You can select the number of times to clear memory.
  • Page 174: Clear Point

    Clear Point Function Group POWER M e n u M a p 5 Description This softkey lets you change the correction value for the active frequency point to the “Undefined” state. SCPI: Programming Codes NONE, see Fltness Menu Analyzer: NONE Menu “Optimizing Swept CW Generator Performance”...
  • Page 175 AUXILIARY INTERFACE connector provides control signals to the Multi-pin Connectors HP 8516A S-parameter test set switch doubler. This connector is a 25-pin D-subminiature receptacle located on the rear panel. It is also used for dual swept CW generator measurement systems (two-tone systems).
  • Page 176 CONNECTORS Table C-l. Pin Description of the Auxiliary Interface In/out Signal Function Level No Connection Blank=+5 V o u t Z-Axis Blanking/Markers M a r k e r = - 5 Spare T T L Spare T T L Low Stop Sweep T T L o u t No Connection...
  • Page 177: C-2. Hp-Ib Connector And Cable

    HP-IB connector allows the swept CW generator to be connected to any other instrument or device on the interface bus. All HP-IB instruments can be connected with HP-IB cables and adapters. These cables are shown in the accompanying illustration. The adapters are principally extension devices for instruments that have recessed or crowded HP-IB connectors.
  • Page 178 HP Sales and Service offices can provide additional information on the HP-IB extenders. The codes next to the HP-IB connector, illustrated in Figure C-2, describe the HP-IB electrical capabilities of the swept CW generator, using IEEE Std. 488-1978 mnemonics (HP-IB, GP-IB, IEEE-488, and IEC-625 are all electrically equivalent).
  • Page 179: C-3. Interface Signals Of The Source Module Connector

    CONNECTORS MOD Cl MOD ANLG SENSE SENSE MOD Dl DIG &D DIG -iiND - 1 5 v - 1 5 v Figure C-3. Interface Signals of the Source Module Connector Figure C-3. Interface Signals of the Source Module Connector The codes indicated on the illustration above translate as follows: The codes indicated on the illustration above translate as follows: MOD DO Source module data line zero.
  • Page 180 CONNECTORS The swept CW generator is equipped with a precision 3.5 mm male RF Output Connector connector (2.4 mm male connector on 40 GHz models). The output impedance, SWR, and other electrical characteristics are listed in “Specifications.” When making connections, carefully align the center conductor elements, then rotate the knurled barrel while the mating component remains still.
  • Page 181: Copylist

    This softkey lets you disable the user flatness array (frequency- Description correction pairs) so that the 1601 point flatness array will be applied when is on. The 1601 point flatness array is FLTNESS ON accessible only through the HP-IB interface. SCPI: Programming Codes Analyzer: NONE Fltness Menu See Also “Optimizing Swept CW Generator Performance”...
  • Page 182: Coupling Factor

    Coupling Factor Function Group M e n u M a p 1 Description This softkey allows specification of the coupling factor of an external coupler/detector used to externally level the swept CW generator output power. Negative coupling factor values are required for valid entry.
  • Page 183: Cw/Cf Coupled

    Function Group FREQUENCY Menu Map This softkey couples the CW function to the center frequency Description function. Any change initiated in either one of these parameters causes a change in the other. SCPI: FREQuency:CW:AUTO ONlOFFlllO Programming Codes Analyzer: NONE See Also Operating and Programming Reference C-13...
  • Page 184: Dblr Amp Menu

    Dblr Amp Menu Function Group POWER M e n u M a p 5 Description This softkey accesses the doubler amp mode softkeys. These softkeys are applicable to instrument models with a doubler installed. The doubler has an integral amplifier whose operation is controlled by the instrument firmware.
  • Page 185: Delete Menu

    Delete Menu FREQUENCY, POWER Function Group M e n u M a p %5 In the menu structure there are two occurrences of this softkey. It Description leads to the delete choices for both the frequency list menu and the power flatness menu.
  • Page 186: Delete Current

    D e l e t e Undef Delete Current Function Group FREQUENCY, POWER M e n u M a p 2,s In the menu structure there are two occurrences of this softkey. One Description occurs in the frequency list menu. The other occurs in the power flatness menu.
  • Page 187 Delta Marker MARKER Function Group M e n u M a p 3 This softkey causes the difference in frequency between two markers Description to appear on the swept CW generator display. The frequency difference is indicated in the following format: XXXXX MHz where m= the last marker activated, n= the Im- nl reference marker, and XXXXX represents some frequency value.
  • Page 188: Deltamarker

    Disp Status Delta Mkr Ref M A R K E R Function Group Menu Map Description This softkey displays the five markers available as the delta marker reference. The delta marker frequency is calculated using the equation: where f, is the frequency of the active marker and fmref is the frequency of the reference marker.
  • Page 189: D-L. Mnemonics Used To Indicate Status

    ALC Leveling Point Lvl Internal External Power Meter Source Module ALC Leveling Mode ALC Search Flatness On/Off Automatic Auto Start Sweep Trigger SwpTrig HP-IB External Power Slope Rf Slope Power Sweep Pwr Swp Sweep Mode Ramp Swept Step Step List...
  • Page 190: Doubler Amp Wode Auto

    Doubler Mode Off Doubler Amp Mode AUTO POWER Function Group Menu Map This softkey is applicable to instrument models with a doubler Description installed. The doubler has an integral amplifier whose operation is controlled by the instrument firmware. The use of the amplifier depends on the frequency of operation and on the calibration constants set at the factory.
  • Page 191: Doubler Arap Mode On

    Doubler Mode Off Programming Codes POWer:AMPLifier:STATE ONlOFFlOll POWer:AMPLifier:STATE? Analyzer: NONE Dblr Amp Menu See Also Doubler Amp Mode On Function Group POWER M e n u M a p This softkey is applicable to instrument models with a doubler Description installed.
  • Page 192: Dwell Coupled

    Dwell Coupled Dwell Coupled Function Group FREQUENCY M e n u M a p 2 Description This softkey lets you couple the dwell time for points in the stepped frequency sweep mode to the ramp sweep mode sweep time. The equation to determine the dwell time in the dwell coupled mode is as follows: Coupled Dwell Time = (sweep time) + (number of step points)
  • Page 193: 8360Adrs

    M e n u M a p 8 Description This softkey lets you change the HP-IB address of the swept CW generator. Enter the address desired using the numeric entry keys or the up/down arrow keys. The address value may be set between 0 and 30.
  • Page 194: Enterfreq

    SCPI: NONE, see Fltness Menu Programming Codes Analyzer: NONE Fltness Menu See Also “Optimizing Swept CW Generator Performance” in Chapter I Enter Freq POWER Function Group Menu Map Description This softkey lets you enter a frequency point into the flatness correction array.
  • Page 195: Enter List Freq

    Enter List Freq This softkey lets you enter a dwell time for a frequency point in the Description frequency list array. A frequency point must be entered before a dwell value can be accepted, otherwise the following error message appears: ERROR: Must first enter a List Frequency.
  • Page 196: Enter List Offset

    Enter List Offset FREQUENCY Function Group M e n u M a p 2 Description This softkey lets you enter an offset value for a frequency in the frequency list. A frequency point must be entered before a power value can be accepted, otherwise the following error message appears: ERROR: Must first enter aList Frequency.
  • Page 197: ( Entry On / Off )

    This softkey enables the swept CW generator to act as a controller to an HP 437B power meter. This softkey causes an immediate execute on the interface bus and generates an HP-IB error if no power meter is present on the interface bus or if the swept CW generator is unable to address the power meter.
  • Page 198: Fault Menu

    Fault Menu Function Group SERVICE M e n u M a p Description This softkey accesses the fault information softkeys. Use this softkey if a fault is indicated on the message line. Indicates the latched status of PEAK, TRACK, Fault Info 1 RAMP, SPAN, V/GHZ, and ADC.
  • Page 199: Fault Info 1

    F a u l t Menu Analyzer: NONE See Also Fault Info 1 Function Group SERVICE M e n u M a p 6 Description This softkey displays the latched status of the following fault messages. PEAK FAIL Indicates that the peak algorithm is unable to align the YTM passband to the frequency of the YO.
  • Page 200: Fault Info 2

    F a u l t I n f o 2 SCPI:See Fault Menu. Programming Codes Analyzer: NONE Fault Menu See Also Fault Info 2 SERVICE Function Group M e n u M a p 6 This softkey displays the latched status of the following fault Description messages.
  • Page 201: Fault Info 3

    Fault Info 2 SCPI: Programming Codes NONE Analyzer: NONE Menu See Also Fault Fault Info 3 Function Group SERVICE M e n u M a p 6 Description This softkey displays the latched status of the following fault messages. CALYO FAIL Indicates that the YO adjusted at power-on or at preset is unable to calibrate.
  • Page 202: Fltness Menu

    Copies the frequency list, (see List Menu ), into the frequency parameter of the flatness correction array. Disables the frequency-correction pair array and uses the HP-IB transferred 1601 point correction set to apply correction information. Reveals the delete softkeys. Delete Menu...
  • Page 203 CW generator. H P - 1 6 I CORRECTED D E V I C E UNDER T E S T Figure F-l. Basic User Flatness Configuration Using an HP 4378 Power Meter F-6 Operating and Programming Reference...
  • Page 204: User Flatness Correction Table As Displayed By The Swept Cw Generator

    User Flatness Correction Table as Displayed by the Swept CW Generator Theory of operation The unparalleled leveled output power accuracy and flatness of the HP 8360 L-Series swept CW generator. This is achieved by using a new digital (versus analog) design to control the internal automatic leveling circuitry (ALC).
  • Page 205: The Sources Of Alc Calibration Correction Data

    F l t n e s s Menu 1601 Equodistont Point Arroy Accessible Only F r o m o C o m p u t e r User Flatness Correction Arroy 1601 Points for ALC 1 6 0 1 P o i n t s o f I n t e r n a l Calibration Data Figure F-3.
  • Page 206 (PO max - Ppatn loss). For example, if an HP 83630L has a maximum path loss of 15 dB due to system components between the source output and the test port, the test port power should be set to -5 dBm.
  • Page 207 Fltness Menu command is set to array the array chosen is the 1601 point correction set. CORRection:SOURce[i]? Queries the source of correction. CORRection[:STATe] ON]OFF]l]O Sets the switch on the user flatness correction feature. This is the same as pressing on the front panel. FLTNESS Queries the condition of the internal switch.
  • Page 208: Freq Cal Menu

    Freq Follow Freq Cal Menu Function Group USER CAL M e n u M a p 9 Description This softkey accesses the sweep span calibration menu. Performs a sweep span calibration each time Swp Span Cal Always the frequency span is changed. Performs a sweep span calibration.
  • Page 209: Frequency

    Function Group FREQUENCY M e n u M a p 2 This hardkey allows access to the frequency functions listed below. Description When this feature is on, the center frequency and the CW frequency is kept equal. Changing either the center frequency or the CW frequency causes the other to change to the same value.
  • Page 210: Freq Mult

    Freq Mult Freq Mult FREQUENCY Function Group Menu Map This softkey lets you set a frequency multiplier value and applies it to Description all frequency parameters. Any integer value between and including it does not affect the output of the swept CW generator. For example: 1.
  • Page 211: Freq Offset

    Freq Offset Function Group FREQUENCY M e n u M a p 2 This softkey lets you set a frequency offset value and applies it to Description all frequency parameters. The frequency offset ranges between and including 6110.0 GHz. Changing the frequency offset value changes the display but does not affect the output frequency.
  • Page 212 Programming Codes See the individual types of calibration. Analyzer: NONE Auto Track, Peak RF Always, Peak RF Once, See Also Swp Span Cal Always, Swp Span Cal Once Operating and Programming Reference F-15...
  • Page 213: Global Dwell

    Global Dwell FREQUENCY Function Group Menu Map This softkey is used to set a dwell time value for all points in the Description frequency list array. SCPI: NONE, Programming Codes see List Menu Analyzer:NONE See Also Enter List Dwell, List Menu “Optimizing Swept CW Generator Performance”...
  • Page 214: Hp-Ib Address

    HP-IB address, refer to “Address” in this manual. HP-IB Menu Function Group SYSTEM M e n u M a p 8 This softkey reveals the softkeys in the HP-IB control menu. Description Reveals the softkeys that Adrs Menu allow HP-IB addresses to be changed.
  • Page 215 HP-IB CONNECTORS, HP-IB See Also “Getting Started Programming” in Chapter 1 H-2 Operating and Programming Reference...
  • Page 216: Leveling Modealcof F

    Leveling Mode Function Group Menu Map This softkey lets you open the ALC loop. Direct and separate Description control of the linear modulator circuit (LVL DAC) and attenuator (ATN) is possible (see Figure A-l). The power level must be set using an external indicator (power meter/sensor).
  • Page 217: Leveling Modenorxtal

    Leveling Mode Normal Function Group Menu Map Description This softkey lets you set the leveling mode of the swept CW generator to continuous leveling at the desired leveling point. In this mode, the RF OUTPUT is controlled by the automatic level control (ALC) circuit, otherwise referred to as the leveling loop.
  • Page 218: Leveling Pointextdet

    SCPI: POWer:ALC:STATe SEARch Programming Codes Analyzer: See Also “Working with Spectrum Analyzers/Reverse Power Effects” in Chapter 1 Leveling Point Function Group Menu Map This softkey lets you set the swept CW generator to accept an Description external feedback connection from a negative-output diode detector to level power.
  • Page 219: Leveling Pointmodule

    Description output of an HP 8355X series millimeter-wave source module. All models of the HP 8360 L-Series swept CW generator drive mm-wave source modules. High power models of HP 8360 drive the mm-wave source modules directly and to specified power levels.
  • Page 220: Leveling Pointpwrmtr

    LINE SWITCH Leveling Point Function Group Menu Map This softkey lets you set the swept CW generator to level at the Description power sensor of an external power meter. This mode of operation requires a feedback connection from the power meter to the EXT ALC BNC located on the swept CW generator.
  • Page 221 List Menu FREQUENCY Function Group M e n u M a p 2 This softkey allows access to the frequency list functions. Description Automatically creates a frequency list using the Auto Fill Incr user-specified increment value. Auto Fill XPts Automatically creates a frequency list containing a user-specified number of points.
  • Page 222: Listmenu

    Enter List Dwell or Global Dwell softkey to enter dwell time values. The editing softkeys of this menu are not accessible over HP-IB. Frequency lists to be loaded over HP-IB must first be created in the controlling program and then downloaded in their entirety to the swept CW generator.
  • Page 223: List Mode Pt Trigauto

    List Menu LIST[:POWer]:CORR t’ ec lon:POINts? [MAXimum]MINimum] In the above three commands, the swept CW generator responds with the number of points for the named parameter that are in the list array. If a particular list is shorter than another, an error is generated upon execution.
  • Page 224: List Mode Pt Trigbus

    FREQUENCY Menu Map Description This softkey lets you set the trigger point to be the HP-IB. When the swept CW generator receives an HP-IB trigger, it steps to the next frequency point of the frequency list, provided the swept CW generator is in sweep list mode.
  • Page 225 ( L O C A L ) INSTRUMENT STATE Function Group NONE Menu Map This hardkey lets you cancel remote operation and return the swept Description CW generator to front panel operation. The front panel keys are deactivated when the swept CW generator is operated remotely. If the external controller executes a LOCAL LOCKOUT command, pressing the @%iiJ key does not return the swept CW generator to front panel control.
  • Page 226: Mi--M2 Sweep

    MI--M2 Sweep MARKER Function Group M e n u M a p 3 This softkey lets you set the swept CW generator to start sweeping Description at the frequency of marker 1 (Ml), and stop sweeping at the frequency of marker 2 (M2). M2 must have a higher frequency value than Ml.
  • Page 227 Manual Sweep resolution defined by the a and @ keys. Frequencies in the manual sweep mode are synthesized, just as they are in mode. There are two major differences between manual sweep and a sweep generated by activating the CW function and rotating the rotary knob or pressing the ARROW keys.
  • Page 228: Marker M2

    MENU SELECT Function Group M e n u M a p 3 This hardkey allows access to the marker functions. Description Causes the swept CW generator to display markers as an amplitude pulse. Changes the swept CW generator’s center frequency to the value of the most recently activated marker.
  • Page 229 Marker Ml Function Group MARKER M e n u M a p 3 Description The softkeys labeled Marker Ml through Marker M5 function identically. The softkey turns the marker off/on. When an asterisk appears next to the key label, it indicates that the marker is on, but not necessarily active.
  • Page 230 Marker N3 Marker M2 MARKER Function Group M e n u M a p 3 Description See MARKER Ml Programming Codes Analyzer: M2 function on, MO function off. Menu , See Also “Marker Operation” in Chapter 1 “Programming Typical Measurements” in Chapter 1 Marker M3 MARKER Function Group...
  • Page 231: Marker M4

    Marker M4 Function Group MARKER M e n u M a p 3 See MARKER Ml Description Programming Codes Analyzer: M4 function on, MO function off. See Also “Marker Operation” in Chapter 1 “Programming Typical Measurements” in Chapter 1 Marker M5 MARKER Function Group Menu Map...
  • Page 232: Markers All Off

    M e n u M a p 5 This softkey enables the swept CW generator to act as a controller Description to command an HP 437B power meter to measure flatness correction values at all frequency points defined in the flatness array. SCPI:...
  • Page 233: Measure Corr Current

    Menu Map This softkey lets you enable the swept CW generator to act as a Description controller to command an HP 437B power meter to measure a single flatness correction value at the current flatness array frequency. SCPI: Programming Codes...
  • Page 234: Meter Adss

    Description In cases where the swept CW generator is capable of acting as a controller to an HP 437B power meter, this softkey enables you to set the programming address of the power meter. The address value can be set from 0 to 30, with the factory default address set at 13. The address value is stored in non-volatile memory.
  • Page 235: Module Menu

    Module Menu Function Group IALC) M e n u M a p 1 This softkey accesses the source module selection softkeys. Description Millimeter-wave source modules can be connected to the swept CW generator source module interface connectors (there is one each on the front and rear panels).
  • Page 236: Module Select Auto

    Module Select AUTO POWER and FREQUENCY Function Group 2 and 5 Menu Map This command sets the automatic selection of the millimeter source Description module interface connector. The swept CW generator looks at both front and rear connectors and determines the type of source module (if any) connected.
  • Page 237: Module Select None

    Module Select Front Programming Codes SYSTem:MMHead:SELect FRONtJREARINONE SYSTem:MMHead:SELect? Analyzer: NONE See Also Module Menu Module Select None Function Group POWER and FREQUENCY 2 and 5 Menu Map Description This command disables millimeter source module sensing. The swept CW generator will alter its frequency limits and multiplier even if a source module is connected to either source module interface connector.
  • Page 238: Module Select Rear

    more n/m Module Select Rear Function Group POWER and FREQUENCY 2 and 5 Menu Map This command causes the swept CW generator to examine only Description the rear panel source module interface connector to determine the type of source module (if any) connected. The instrument frequency limits and multiplier are altered according to the source module connected.
  • Page 239: Mtr Meas Menu

    Measures flatness correction values for all frequency points in the flatness correction array that have no correction values assigned. The meter measure function uses an external HP 437B power meter to automatically measure and store power correction values for the frequency points requested.
  • Page 240: Peak Rf Always

    Peak RF Always POWER, USER CAL Function Group M e n u M a p 5,9 This softkey appears in two locations: the POWER Tracking Menu Description and the USER CAL Tracking Menu. operation is the same in both locations. This softkey causes the swept CW generator, when in CW or manual-sweep output mode, to align the output filter (SYTM) so that its passband is centered on the RF output.
  • Page 241 Peak RF Once Function Group POWER, USER CAL Menu Map 5, 9 This softkey appears in two locations: the POWER Tracking Menu Description and the USER CAL Tracking Menu. The operation is the same in both locations. This softkey causes an instantaneous, one-time execution of the peaking function when the swept CW generator is in the CW or manual sweep mode.
  • Page 242 When you press the active entry area displays: POWER LEVEL X. XX dBm, where X represents a numeric value. The data display area indicates: INT : x.xx. Normal, Uncoupled Attenuator, key controls POWER LEVEL Level Level Control Circuits (see Figure A-l) within the ALC level range (-l-25 to -20 dBm).
  • Page 243 When you press the active entry area displays: POWER LEVEL where X represents a numeric value. The data display area indicates: MTR : x . x x . Normal, Module, key controls the output POWER LEVEL power of the swept CW generator as compared to the feedback voltage from a millimeter-wave source module.
  • Page 244 F L T N E S S O F F Set Atten, Tracking Menu , Uncoupl Atten “Programming Typical Measurements” in Chapter 1 POWER Function Group M e n u M a p 5 This hardkey accesses the power function softkeys. Description Accesses the softkeys in the flatness correction menu.
  • Page 245 F L T N E S S O F F P O W E R L E V E L “Introducing the HP 8360 L-Series Swept CW Generators” in Chapter 1 and “Getting Started Advanced” in Chapter 1 “Programming Typical Measurements” in Chapter 1 Power Offset...
  • Page 246 Power Slope Power Slope POWER Function Group Menu Map This softkey lets you compensate for system, cable, and waveguide Description variations due to changes in frequency, by linearly increasing or decreasing power output as the frequency increases. RF slope values may range from -2.50 to j-2.50 dB per GHz.
  • Page 247 Power Sweep POWER Function Group Menu Map Description This softkey enables the power sweep function. RF output power can be swept both positively and negatively over a selected range. The level of the power sweep starting point is the power level programmed.
  • Page 248 If the red LED adjacent to the (PRESET) key (labeled INSTR CHECK) stays on after preset, the swept CW generator failed self-test; refer to HP 8360 B-Series Swept Signal Generator/HP 8360 L-Series Swept CW Generator Troubleshooting Guide.
  • Page 249 Preset Mode Factory SYSTEM Function Group Menu Map Description This softkey sets the standard starting configuration of the swept CW generator when the (PRESET) key is pressed, as set by the manufacturer. An asterisk next to the key label indicates that this feature is active.
  • Page 250 P r i n t e r Adrs Preset Mode User SYSTEM Function Group M e n u M a p 8 This softkey sets the standard starting configuration of the Description swept CW generator when the (PRESET) key is pressed, as set by the user.
  • Page 251 Function Group MENU SELECT Menu Map NONE Description This hardkey lets you view previous menus. All menus visited from the last preset are remembered and displayed in a “last-visited-first-seen” order. Refer to Figure P-l, and follow the arrow paths as indicated. AREA PRIOR Figure P-l.
  • Page 252: Analyzr

    Description This softkey lets you select Analyzer Language as the swept CW generator’s interface language. This language uses HP 8340/8341 mnemonics and provides HP network analyzer compatibility. Any commands issued within 100 ms of a change in language may be ignored or lost.
  • Page 253: Ciil

    SYSTem:LANGuage CIIL Programming Codes Analyzer: CIIL Adrs Menu See Also The M.A.T.E. option (Option 700) is documented in a separate HP 8360 Series Synthesized Sweepers manual supplement called, Option 700 Manual Supplement. Programming Language SCPI SYSTEM Function Group Menu Map...
  • Page 254: Pulse On/Offscal.ar

    (18 ps pulse width, 36 ps pulse period). The rise and fall times of the RF envelope are approximately 2 ps. These pulses allow proper operation with HP scalar network analyzers in ac detection mode. An asterisk next to the key label indicates that this feature is active.
  • Page 255: Pwr Mtr Range

    Pulse On/OffScalar Programming Codes PULSe:SOURce SCALar PULSe[:STATe] ON]OFF]l]O Analyzer: SHPM function on, PM0 function off. See Also Chapter 3 Pwr Mtr Range Function Group Menu Map Description This softkey lets you specify a range of operation (from +20 to -60 dBm) for an external power meter, when a power meter is used to level power externally.
  • Page 256: (Recall)

    Function Group SYSTEM M e n u M a p 8 This hardkey retrieves a front panel setting that was previously Description stored in a SAVE register (1 through 8). SCPI: Programming Codes *RCL <num> The above is an IEEE 488.2 common command. Analyzer: [SAVE), SCPI COMMAND SUMMARY “Saving and Recalling an Instrument State”...
  • Page 257: Rotary Knob

    Ref Osc Menu 10 MHz Freq Standard None Sets the swept CW generator to free-run operation, where no frequency standard is used. SCPI: Programming Codes Analyzer: NONE See Also POWER Function Group NONE Menu Map Description This hardkey turns the RF power output on or off. Press (WON/OFF). If the yellow LED above the hardkey is off, power is off, and RF OFF appears in the message line of the display.
  • Page 258 S A V E SYSTEM Function Group Menu Map This hardkey allows up to eight different front panel settings to Description be stored in memory registers 1 through 8. Swept CW generator settings can then be recalled with the [RECALL) key. A memory register can be alternated with the current front panel setting using the Atrnate Regs soft key.
  • Page 259: Savelock

    Save Lock SYSTEM Function Group Menu Map This softkey lets you disable the save function. It prohibits the Description saving of the present instrument state into a save/recall memory register. If this function is active, an error message is displayed. An asterisk next to the key label indicates that this function is active.
  • Page 260: Scpi Conformance Information

    The HP 8360 L-Series swept CW generators conform to the 1990.0 version of SCPI. The following are the SCPI confirmed commands implemented by the HP 8360 L-Series swept CW generators: :MARKer[n] :AOFF :ALC Operating and Programming Reference S-3...
  • Page 261 :PULM :ROSCillator :NTRansition :NTRansition? :PTRansition :PTRansition? :NTRansition :NTRansition? :PTRansition :PTRansition? S-4 Operating and Programming Reference...
  • Page 262 :LLIMit :LLIMit? :GPIB The following are the SCPI approved commands implemented by the HP 8360L-Series swept CW generators: Instrument-specific diagnostic commands: :ABUS :ABUS? :ADD :PMETer Operating and Programming Reference S-5...
  • Page 263 :IORW :IORW? :LED :IOCHeck :IOCHeck? : o s c :FNCW :FNDN :FNUP :SAMP :YODacs? :YTMDacs :SRECeiver :ASTate :DESC? S-6 Operating and Programming Reference...
  • Page 264 :LOG The following are the commands implemented by the HP 8360 L-Series swept CW generators which are not part of the SCPI definition: :AD Just :PMETer Operating and Programming Reference S-7...
  • Page 265 :ALL :ARRay[i]? S-6 Operating and Programming Reference...
  • Page 266 :SOURce[i]? :MARKer[n] :ALC :CFACtor :CFACtor? Operating and Programming Reference S-9...
  • Page 267 :MSIB :NTRansition :NTRansition? :PTRansition :PTRansition? :SREceiver :NTRansition :NTRansition? :PTRansition :PTRansition? S-10 Operating and Programming Reference...
  • Page 268 a ive ? :XFER :KEY :MMHead :ODELay :ODELay? Operating and Programming Reference S-l 1...
  • Page 269 Sets and queries the value of the Standard Event Status Enable Register. Queries the value of the Standard Event Status Register. This is a destructive read. This returns an identifying string to the HP-IB. The response is in the following format: HEWLETT-PACKARD MMM YY, where the actual model number, serial number, and firmware revision of the swept CW generator queried is passed.
  • Page 270 *OPT? This returns a string identifying any device options. The instrument state is recalled from the specified memory register. The value range is from 0 to 8. The swept CW generator is set to a predefined condition as follows: CALibration:PEAKing:AUTO OFF CALibration:POWer:ATTenationO DBM CALibration:POWer:RANGe 1 CALibration:SPAN:AUTO OFF...
  • Page 271 POWer:ALC:BANDwidth:AUTO ON POWer:ALC:CFACtor -16 DBM POWer:AMPLifier:STATE:AUTO ON POWer:ATTenuation:AUTO ON POWer:CENTerO DBM POWer:MODEFIXed POWer:SLOPe 0 POWer:SLOPe:STATe OFF POWer:SPAN 0 DB POWer:STARt 0 DBM POWer:STATe OFF POWer:STEP:AUTO ON POWer:STOP 0 DBM PULM:SOURce SCALar PULM:STATe OFF SWEep:POINts 11 SWEep:STEP value is (StopMAX-StartMIN)/lO SWEep:TIME MINimum SWEep:TIME:AUTO ON SWEep:TIME:LLIMit 10ms SWEep:GENeration ANALog...
  • Page 272 Queries the Status Byte. This is a non-destructive read. This command performs the same function as the Group Execute Trigger command defined by IEEE 488.1. *TST? A full selftest is performed, without data logging or looping, and returns one of the following error codes: Error Code Definition Test passed.
  • Page 273: S-L. Hp 8360 Scpi Command Summary

    Table S-l. HP 8360 SCPI COMMAND SUMMARY Parameters Parameter C o m m a n d Allowed Values Boolean auto RF peak type of discrete det cal power correction extended numeric value discrete to cal measured power extended numeric auto calibrate...
  • Page 274: Scpi Command Summary

    Table S-l. HP 8380 SCPI COMMAND SUMMARY (continued) Command Parameters Parameter Type1 Allowed Values xtended numeric to 31 winter address xtended numeric to 31 :IORW xtended numeric :DESC? xtended numeric nable listed xtended numeric xtended numeric :LOG tate tate xtended numeric...
  • Page 275 Table S-l. HP 8360 SCPI COMMAND SUMMARY (continued) Command Parameters Parameter Type1 Allowed Values Boolean state center freq extended numeric specified freq range or MAXimumlMINimumlUPlDOWN extended numeric specified freq range cw freq or MAXimumlMINimumlUPlDOWN coupled to Boolean center freq start/stop limits...
  • Page 276 Table S-l. HP 8360 SCPI COMMAND SUMMARY (continued) Command Parameters Parameter Type’ Allowed Values LIST extended numeric (0.1 to 3200 ms}*801 or {specified freq range}*801 or list freq extended numeric extended numeric 1 to maximum defined numeric to lock on...
  • Page 277 Table S-l. HP 8360 SCPI COMMAND SUMMARY (continued) Command Parameters Parameter Type1 Allowed Values :ALC ALC bwidth extended numeric or MAXimumlMINimum bwidth selection Boolean coupling factor extended numeric 0 to -9O[DB] or leveling point discrete state Boolean Boolean Boolean extended numeric...
  • Page 278 Table S-l. HP 8360 SCPI COMMAND SUMMARY (continued) Command Parameters Parameter Type1 Allowed Values PULM discrete tate Boolean discrete ef osc source Boolean tate numeric 0 to 2047 numeric 0 to 2047 ilter 0 to 2047 numeric ilter numeric 0 to 2047...
  • Page 279 Table S-l. HP 8360 SCPI COMMAND SUMMARY (continued) Parameter Type1 Allowed Values Parameters Boolean dual source mode discrete type of sweep control 0.1 to 3200 ins extended numeric settling time or MAXimumlMINimum plus dwell time Boolean dwell calculation state type of sweep...
  • Page 280 Table S-l. HP 8360 SCPI COMMAND SUMMARY (continued) Command Parameters Parameter Type1 Allowed Values save/recall numeric 1 to 8IMAXimumlMINimum register state Boolean :GPIB 1 to 30 address : K E Y key code assign numeric 0 to 511, 1 to 14 excluding 5 and 10...
  • Page 281 Causes the sweep in progress to abort and reset. If ON it immediately restarts the sweep. The pending operation flag (driving *OPC, *WAI, and *OPC?) undergoes a transition once the sweep is reset. CALibration:PEAKing:AUTO ONiOFFlllO CALibration:PEAKing:AUTO? Sets and queries the automatic peaking function. If AUTO is ON, then a peak is done at regular intervals automatically.
  • Page 282 PMETer Initiates a calibration of the power meter flatness. Depends on value of CALibrat ion : PMETer : RANGe. Initiates a calibration of the source module flatness. Depends on value of CALibrat ion : PMETer : RANGe. CALibration:PMETer:FLATness:NEXT? <num>[lvlsuffix] The parameter is the measured power that is currently produced by the swept CW generator.
  • Page 283 Set to 1, if out of range occurred. DIAGnostics:INSTrument:PMETer:ADDRess <num> DIAGnostics:INSTrument:PMETer:ADDRess? Sets and queries the HP-IB address to use for the power meter during swept CW generator calibration routines. Allowable values are 0 through 31. *RST or power on does not effect this value. Default is 13.
  • Page 284 Sets the HP-IB address of the printer to use during some of the calibration procedures when the swept CW generator assumes HP-IB control. *RST and power on do not effect this command. The default is 1. The default value is set at memory initialization only.
  • Page 285 HP 8360 L-Series Swept CW Generator Service Guide to begin the troubleshooting procedure. DIAGnostics:TEST:CONTinue Causes the selftest execution to continue when paused for raw data examination. Does nothing in other conditions. DIAGnostics:TEST:DATA:DESC? Returns the description string of the raw data examined during a selftest.
  • Page 286 DIAGnostics:TINT? <num> A test feature that returns the value passed to it. This is used to test the HP-IB interface. Sets and queries the display ON/OFF switch. After “RST, the value is 1.
  • Page 287 requested value. The start/stop frequencies are updated to reflect the changes based on the coupling equations. The swept CW generator uses “bumping” to move unspecified frequency parameters, but if the final value of any of the frequency headers is the result of bumping, then an error is generated since the user is getting what was specified.
  • Page 288 Sets and queries the manual frequency. This controls the output frequency in swept manual mode. The limits are START and STOP. *RST valueisthe same as FREQ:CENTER.See FREQ:CENTERfor more information. Sets and queries the switch that selects either swept, CW or list operation.
  • Page 289 Queries and turns the frequency offset off and on. After *RST, the setting is OFF. Sets and queries the frequency span. See FREU : CENTER for more information. Sets and queries the START Frequency. See FREU : CENTER for more information.
  • Page 290 take sweep single sweep analyzer compatible language command can be achieved. suffix] IMAXimumlMINimum)l*8Ol Sets and queries the amount of time to dwell at each frequency. The number of dwells can be queried with This dwell is the time after completing phaselock and turning RF on before changing to the next frequency.
  • Page 291 Each new frequency point is stepped to automatically, after waiting the specified A U T O Wait for a <GET> or *TRG over the HP-IB before advancing to the next frequency in the list. A U T O Wait for a signal to be...
  • Page 292 be used, there is really only a single value for all the markers. *RST value is 2 dB. MARKer [n] : AOFF Sets all the markers to OFF at once. While [n] may be used, there is really only a single switch for all the markers. MARKer [n] : DELTa? Qnun>...
  • Page 293 Any place where dBm is accepted as a suffix, any level suffix is Power Subsystem accepted also. In the absence of a suffix, the units are assumed to be set bytheUNIT:POW command. POWer:ALC:BANDwidthl:BWIDth Cnum>[freq suffix] lMAXimumlMINimum POWer:ALC:BANDwidth?l:BWIDth? [MAXimumlMINimuml Sets and queries the ALC bandwidth. This is actually not continuously variable, so the input is rounded to the nearest possible switch position.
  • Page 294 state. Programming a specific value for POWer : AMPLif ier : STATE sets POWer:AMPLifier:STATE:AUTOto OFF. *RSTvalue is ON. POWer:ATTenuation <num>[DB] lMAXimumlMINimumlUPlDOWN POWer : ATTenuat ion? [MAXimum I MINimum] Sets and queries the output attenuation level. Note that when setting the attenuator level to 10 dB the output power is decreased by 10 dB.
  • Page 295 POWer:OFFSet <num>~DBllMAXimumlMINimumlUPlDOWN Sets and queries the power offset,. This function changes mapping of absolute power parameters on input to and output from the swept CW generator. Changing this does not, affect the output power of the swept CW generator. Only the displayed parameters and query responses are changed.
  • Page 296 The *RST value is 0. POWer:SLOPe:STATe ONlOFFlllO POWer:SLOPe:STATe? Sets and queries the power slope state. *RST value is 0. POWer:SPAN <num>[DB] lMAXimumlMINimumlUPlDOWN POWer:SPAN? [MAXimumlMINimum] The coupling equations for power sweep are exactly analogous to those for frequency sweep. Power sweep is allowed to be negative, unlike frequency sweeps.
  • Page 297 Sets and queries the reference oscillator selection switch. The command to set the switch will cause ROSC:SOUR:AUTO OFF to be done also. The *RST value is automatically determined. l ROSCillator:SOURce:AUTO ONIOFFlllO Sets and queries the automatic reference selection switch. The *RST value is 1. STATus:OPERation:CONDition? Queries the Standard Operation Condition register.
  • Page 298 Sets and queries the Data Questionable SRQ Enable register. Queries the Data Questionable Event Register. This is a destructive read. Sets and queries the Negative TRansition Filter for the Data Questionable Status Register. The STAT STATus:QUEStionable:PTRansition? Sets and queries the Positive TRansition Filter for the Data Questionable Status Register.
  • Page 299 Sets and queries the instrument, whether it is in master or slave mode. This applies in a dual source mode. *RST value is MASTer. SWEep : DWELl? [MAXimum I MINimumI Sets and queries the amount of time in seconds that the instrument source settled SRQ stays (dwell) at each step after reporting a pulsing the Trigger...
  • Page 300 or a digitally stepped sweep. In either case, all of the other sweep subsystem functions apply. “RST is ANALog. SWEep:MANual:POINt <num>lMAXimumlMINimum SWEep : MANual : POINt? [MAXimum I MINimum] Sets and queries the step point number to go to and lock. The value is a unitless value that is limited between 1 and the number of points requested.
  • Page 301 SWEep:STEP <num>[freq suffix] IMAXimumlMINimum SWEep:STEP? [MAXimum I MINimum] Sets and queries the size of each frequency step. by the equation: STEP = SPAN/POINTS. If you change step size then the number of points will be changed Parameter Bumped to span/step and a execution error is reported.
  • Page 302 Changes the GPIB’s (General Purpose Interface Bus) address. The *RST value is 19. SYSTem:DUMP:PRINter? Causes a dump of the display contents to be made to the HP-IB. Returns the next message in the error queue. The format of the response is : <error number>,<error string>...
  • Page 303 frequency limits and multiplier will be altered accordingly. However, the leveling point is changed (see POWer:ALC:SOURce). Sets the instrument to its “local operation” state. This is the same as pressing the front panel “green” key. There is no corresponding query. SYSTem:PRESet:SAVE Saves the present state so it can be used whenever the command key is pressed.
  • Page 304 subsystem. This event does not affect any other settings in this subsystem. wait for This command has no effect unless the instrument is in the trig wait trig state. If the swept CW generator is in the state, it performs its trigger action. This is an event and has no “RST condition.
  • Page 305: Scpi Status Register Strubitiri

    STANDARD E ENT STATUS GROUP S-46 Operating and Programming Reference...
  • Page 306 OPERATION STATUS GROUP NEGATIVE TRANSITION FlLTER DATA QUESTIONABLE STATUS GROUP NOTE. OPER. AND OUES. Operating and Programming Reference S-49...
  • Page 307: Security Menu

    ATE environment. A security calibration constant that can be accessed through the service adjustment menu (requires a HP 8360 B-Series password for access) is available also. Refer to the Swept Signal Generator/HP 8360 L-Series Swept CW Generator Service Guide for information on calibration constants. See Also “Using the Security Features”...
  • Page 308: Selftest (Full

    Set Atten SERVICE Function Group M e n u M a p 6 Description This softkey activates the self-test function of the swept CW generator. SCPI: *TST? Programming Codes Analyzer: NONE Also Fault SCPI COMMAND SUMMARY Menu, Chapter 4, “OPERATOR’S CHECK and ROUTINE MAINTENANCE”...
  • Page 309: Software Rev

    “Programming Typical Measurements” in Chapter 1 Software Rev SYSTEM Function Group Menu Map This softkey displays the swept CW generator’s programming Description language, HP-IB address, and firmware date code. SCPI: Programming Codes Analyzer: HP-13 Menu, SCPI COMMAND SUMMARY See Also S-52 Operating and Programming Reference...
  • Page 310 Function Group FREQUENCY M e n u M a p 2 Description This hardkey lets you set a value for the frequency span in the center frequency/frequency span mode of swept frequency operation. Press CW generator sweeps from one half the span below to one half above the center frequency.
  • Page 311: Start=Ml Stop=M2

    Programming Codes FREQuency:STARt <num>[freq suffix] or FREQuency:MODE SWEep Analyzer: FA <num> See Also “CW Operation and Start/Stop Frequency Sweep” in Chapter 1 “Programming Typical Measurements” in Chapter 1 Start=Ml Stop=M2 MARKER Function Group Menu Map This softkey changes the swept CW generator start frequency to the Description frequency value of marker 1, and the stop frequency to the frequency value of marker 2.
  • Page 312: Start Sweep Trigger Auto

    Function Group Menu Map When this softkey is selected, the swept CW generator waits for an Description HP-IB trigger to trigger a sweep. An asterisk next to the key label indicates that this feature is active. SCPI: TRIGger:SOURce BUS Programming Codes...
  • Page 313: Start Sweep Trigger Ext

    Start Sweep Trigger Ext SWEEP Function Group Menu Map When this softkey is selected, the swept CW generator waits for an Description external hardware trigger to trigger a sweep. Connect the trigger pulse to TRIGGER INPUT. It is activated on a TTL rising edge. An asterisk next to the key label indicates that this feature is active.
  • Page 314: Connections Required For A Two-Tone Scalar Network Analyzer Measurement System

    Step Control Master MISTER SLRVE SYNTHESIZER SYNTHESIZER Figure S-l. Connections Required for a Two-Tone Scalar Network Analyzer Measurement System 1. Designate one swept CW generator as the master, the other as the slave. 2. Make the connections. 3. To avoid synchronization problems, always set up the slave (frequency and power) before setting up the master.
  • Page 315: Step Control Slave

    Step Control Slave FREQUENCY Function Group M e n u M a p 2 Description This softkey lets you designate the swept CW generator as the slave in a dual swept CW generator measurement system. A dual swept CW generator system (two-tone measurement system) facilitates accurate device characterizations by providing one timebase reference for both sources.
  • Page 316: Step Dwell

    Step Dwell is active it calibrates the frequency at the end of every frequency band. Programming Codes SWEep:CONTrol:STATe ON]OFFjl]O SWEep:CONTrol:TYPE SLAVe Analyzer: NONE Step Step Sap Menu See Also Control Master, Step Dwell FREQUENCY Function Group Menu Map This softkey lets you set dwell times for points in the stepped Description frequency mode of sweep operation.
  • Page 317: Step Points

    Step Points Function Group FREQUENCY Menu Map This softkey lets you define the number of step points in a stepped Description frequency sweep. The number of points in a stepped sweep can range from 2 to 801. Step Size and Step Points are dependent variables.
  • Page 318: Step Sup Pttrig Bus

    Steps the swept CW generator to Step Sup Pt Trig Bus the next point in a stepped sweep when an HP-IB trigger is received. Steps the swept CW generator to Step Swp Pt Trig Ext the next point in a stepped sweep when an external hardware trigger is received.
  • Page 319 When this softkey is selected, the swept CW generator steps to the Description next point in a stepped frequency sweep when an HP-IB trigger frequency point is reached and continuous sweep is selected, the next trigger causes the step sweep to return to the start frequency.
  • Page 320 (STOP) Step Swp Pt Trig Ext Function Group FREQUENCY Menu Map When this softkey is selected, the swept CW generator steps to Description the next point in the stepped frequency sweep when an external hardware trigger is received. When the last frequency point is reached and continuous sweep is selected, the next trigger causes the step sweep to return to the start frequency.
  • Page 321 Automatically triggers a sweep Start Sweep Trigger Auto when (SINGLE) or (CONT) is pressed. Waits for an HP-IB trigger to Start Sweep Trigger trigger a sweep when (SINGLE) or Waits for an external hardware Start Sweep Trigger...
  • Page 322: Sweep (Menu

    Sweep Mode List SCPI: Programming Codes NONE Analyzer: NONE See Also “Programming Typical Measurements” in Chapter 1 Sweep Mode List SWEEP Function Group Menu Map Description This softkey activates the step frequency list mode. To use this type of sweep, a frequency list must have been entered, otherwise an error message appears.
  • Page 323: Sweep Mode Ramp

    Sweep Mode Ramp SWEEP Function Group M e n u M a p 7 This softkey activates the analog frequency sweep mode. Ramp Description sweep mode is the factory preset state. An asterisk next to the key label indicates that this feature is active. Programming Codes FREQuency:MODE SWEep SWEep[:FREQ uency]:GENeration ANALog...
  • Page 324: Sup Span Calonce

    swp span Cal Always Function Group USER CAL M e n u M a p 9 Description This softkey causes a sweep span calibration each time the frequency span is changed. An asterisk next to the key label indicates this feature is active.
  • Page 325: (Sweep Time)

    Function Group SWEEP M e n u M a p 7 Description This hardkey lets you set a sweep time for frequency sweeps or power sweeps. The sweep time range is 10 ms to 200s but the fastest sweep time is constrained by the frequency span. The fastest possible sweep can be determined automatically: 1.
  • Page 326: System (Menu

    Dims the swept CW generator’s display. Dim Display Displays the present status of the swept Disp Status C W generator. Reveals the HP-IB control menu. HP-IB Menu Sets the preset state, as defined by the Preset Mode Factory manufacturer, to be recalled by the (PRESET) key.
  • Page 327 SYSTEM INIENU_) Activates the USER-DEFINED (MENU) and clears all keys in that menu. SCPI: NONE Programming Codes Analyzer: NONE See Also S-70 Operating and Programming Reference...
  • Page 328 auto SYSTEM Function Group M e n u M a p 8 This softkey sets the swept CW generator to choose its frequency Description standard automatically. If an external standard connected to the 10 MHz REF INPUT BNC, then it is chosen as the reference. If no external standard is connected, the internal standard is chosen as the reference.
  • Page 329 10 MHz Freq Std Extrnl SCPI: Programming Codes ROSCillator[:SOURce] EXTernal Analyzer: NONE Ref Osc Menu See Also 10 MHz Freq Std Intrnl Function Group SYSTEM Menu Map Description This softkey sets the swept CW generator to select the internal 10 MHz signal as the frequency reference. If the internal signal is disconnected or not working properly, UNLOCK appears on the...
  • Page 330 Tracking Menu POWER, USER CAL Function Group Menu Map 5, 9 In the menu structure there are two occurrences of this softkey. One Description occurs in the POWER (MENU), the other occurs in the USER CAL the tracking menu. Realigns the swept CW generator’s output Auto Track filter and oscillator to maximize output power for the swept frequency mode.
  • Page 331 SCPI: Programming Codes TRIGger:ODELay <num>[time suffix] Analyzer: NONE Start Sweep Trigger Auto, Start Sweep Trigger Bus, See Also Start Sweep Trigger Ext T-4 Operating and Programming Reference...
  • Page 332: Service

    Function Group POWER Menu Map This softkey uncouples the attenuator (if there is one) from the ALC Description system. It allows independent control of attenuator settings. An asterisk next to the key label indicates that this feature is active. To set the attenuator after it, is uncoupled, select Set Atten.
  • Page 333 Up/Down Power POWER Function Group Menu Map Description This softkey activates the power step size function. It can be set from 0.01 to 20 dB. In this mode, power is stepped by the up/down arrow keys. An asterisk next to the key label indicates this feature is active.
  • Page 334 SCPI: FREQuency:STEP[:INCR] <num>[freq suffix] or Programming Codes Analyzer: SF or SHCF <num> [Hz]Kz]Mz]Gz] Manual Sweep, Sweep Mode Step, Up/Dn Size Swept See Also FREQUENCY Function Group M e n u M a p 2 This softkey sets the frequency step size in the swept frequency step Description mode.
  • Page 335 Function Group USER CAL M e n u M a p 9 This hardkey accesses the user calibration softkeys. Description Performs a complete alignment as determined by the instrument settings. Accesses the softkeys of the tracking Tracking Menu menu. Freq Cal Menu Accesses the Frequency span calibration menu.
  • Page 336 SCPI: NONE Programming Codes Analyzer: NONE See Also UsrKey Clear Function Group SYSTEM M e n u M a p 8 This Description single 1. Select UsrKey Clear. The user defined menu appears in the 2. Select the softkey you wish to remove from the menu. The active entry area turns off and the softkey is removed from the user defined menu.
  • Page 337 This softkey recalls the user defined and removes all softkeys Description menu assigned to that menu. The empty user defined menu remains in the SCPI: NONE Programming Codes Analyzer: NONE See Also U-6 Operating and Programming Reference...
  • Page 338 The left and right arrows control the resolution with which the center frequency can be changed. This is a front-panel-only feature and is inaccessible over HP-IB. SCPI: NONE...
  • Page 339 DEFAULTING LANGUAGE: This error message is displayed in conjunction with one of the following messages. The HP-IB/Language Invalid Language set on rear panel switch. switch located on the rear panel has been set to an invalid programming language selection. The programming language is defaulted to the previous setting.
  • Page 340 Refer to Chapter 4 and follow the local operator’s check procedures. If you are a qualified service technician and this failure occurs, read the Calibration Constants section in the HP 8360 B-Series Swept Signal Generator/HP 8360 L-Series Swept CW Generator Service Guide. EEROM Failed !!: This error will only occur if the service adjustment menu is accessed.
  • Page 341 Refer to Chapter 4 for instructions on contacting a qualified service technician. INVALID LANGUAGE ON REAR PANEL SWITCH: HP-IB/Language switch located on the rear panel has been set to an invalid programming language selection. Check the rear panel switch. See Chapter 3 for information on language selection.
  • Page 342 SYSTEM CONTROLLER BUS: This error message is generated when an external controller is active on the HP-IB and the swept CW generator has attempted to act as the controller. Disconnect the HP-IB interface or return the swept CW generator to LOCAL operation and repeat the request.
  • Page 343 Qualified service technicians, refer HP 8360 B-Series Swept Signal to “ADJUSTMENTS,” in the Generator/HP 8360 L-Series Swept CW Generator Service Guide more information. 0, No Error: This message indicates that the device has no errors...
  • Page 344 Qualified service technicians, HP 8360 B-Series Swept Signal refer to “ADJUSTMENTS,” in the Generator/HP 8360 L-Series Swept CW Generator Service Guide more information. Universal SCPI Error Error Messages From -499 To -400 Messages...
  • Page 345 Errors that generate Execution Errors do not generate Command Errors, Device-specific Errors, or Query Errors. -240, Hardware error; Rear panel HP-IB switch -224, Illegal parameter value -222, Data out of range;Expected O-l...
  • Page 346 -120, Numeric data error;Bad terminator -113, Undefined Header;Query not allowed -113, Undefined header;Bad mnemonic -109, Missing parameter -108, Parameter not allowed;Too many -105, GET not allowed -104, Data type error -104, Data type error;Block not allowed -104, Data type error;Char not allowed -104, Data type error;Decimal not allowed -104, Data type error;Non-dec not allowed -104, Data type error;String not allowed...
  • Page 347 Menu Maps Menu Maps 2b-1...
  • Page 357 Specifications This section lists the specifications for the HP 8360 L-Series Swept CW Generator. In a effort to improve these swept CW generators, Hewlett-Packard has made changes to this product which are identified with changes in the serial number prefix.
  • Page 358 1 This band is 20 GHz to <‘25.5 GHz on the HP 83640L. 2 This band is 25.5 GHz to < 32 GHz on the HP 83640L. 3 This band is 32 GHz to < 40 GHz on the HP 83640L.
  • Page 359 Accuracy: Same as time base Synthesized Step Minimum Step Size: Same as frequency resolution Sweep Number of Points: 2 to 801 Switching Time: Same as CW Dwell Time: 100 ps to 3.2 s Accuracy: Same as time base Synthesized List Mode Minimum Step Size: Same as frequency resolution Number of Points:...
  • Page 360 Output Power M a x i m u m L e v e l e d Standard HP 83623L H P 8363OL Output Frequencies < 20 GHz Output Frequencies 2 20 GHz H P 83640L Output Frequencies < 26.5 GHz Output Frequencies >...
  • Page 361 Accuracy (dB)4 Specifications apply in CW, step, list, manual sweep, and ramp sweep modes of operation. Frequency (GHz) > 2.0 and 5 40 > 40 Power > +lO dBm f1.2 f0.9 f1.7 > -10 dBm5 f0.6 f0.7 fl.O f1.2 > -60 dBm f0.9 f1.7 f2.5...
  • Page 362 Analog Power Sweep step attenuator. Range External Leveling At External HP 33330D/E Detector: -36 to +4 dBm At External Leveling Input: -200 PV to -0.5 volts Bandwidth External Detector Mode: 10 or 100 kHz (sweep speed and modulation mode dependent), nominal Power Meter Mode: 0.7 Hz, nominal...
  • Page 363 T y p i c a l A L C Linearity ( F r e q u e n c i e s 2 20GHr) - 1 0 A L C Level(dBm) Specifications apply in CW, step, list, and manual sweep modes of operation.
  • Page 364 Non-Harmonically Related Output Frequencies: - 6 0 < 2.0 GHzg - 6 0 > 26.5 and 2 40 GHz - 5 4 - 5 2 > 40 GHz Power-Line Related (< Hz offset from carrier) 10 MHz to < 7 GHz - 5 5 7 GHz to <...
  • Page 365 Single-Sideband Offset from Carrier Phase Noise 1 kHz 10 kHz 100 kHz Band(s) 100 Hz - 7 8 - 8 6 - 1 0 7 10 MHz - 7 0 to < 7 - 8 0 - 6 4 - 7 2 -101 - 6 8 - 7 6...
  • Page 366 Net Weight: 27 kg (60 lb) Shipping Weight: 36 kg (80 lb) Dimensions: 178 H x 425 W x 648 mm D (7.0 x 16.75 x 25.5 inches) HP 83623L, HP 83630L Adapters Supplied Type-N (female) - 3.5 mm (female) Part number 1250-1745 3.5 mm (female) - 3.5 mm (female)
  • Page 367 In CW mode, voltage is proportional to percentage of full instrument frequency range. Minimum load impedance 3 kilohms. Accuracy f0.25%, 510 mV, typical. (BNC female, rear panel.) Stop Sweep Input/Output Sweep will stop when grounded externally. TTL-high while sweeping, TTL-low when HP 8360 stops sweeping. Damage level Specifications 2c-11...
  • Page 368 Provides bias, flatness correction, and leveling connections to HP 83550-series millimeter-wave source modules (Special, front and rear panels.) Auxiliary Interface Provides control signal connections to HP 8516A S-parameter Test Set. (25-pin D-subminiature receptacle, rear panel.) HP 83623L: 10 MHz to 20 GHz High Power Models H P 83630L: 10 MHz to 26.5 GHz...
  • Page 369 Option 910 Extra Operating 81 Service Guides Provides a second copy of operating and service guides. Option 013 Rack Flange Kit Used to rack mount HP 8360 with front handles. Front handles are standard on the HP 8360. Option W30 Two Years Additional Return-To-HP Service Does not include biennial calibration.
  • Page 370: Installation

    Hewlett-Packard office. Keep the shipping material for the carrier’s inspection. The HP office will arrange for repair or replacement without waiting for a claim settlement.
  • Page 371: Equipment Supplied

    All HP 8360 L-Series swept CW generators are sent from the factory with the following basic accessories: Rack handles (mounted) Power cord A set of manuals The following adapters are also shipped with the swept CW generators: Table 3-1. Adapter Descriptions and Part Numbers Shipped with...
  • Page 372: Preparation For Use

    This instrument protects against finger access to hazardous parts within the enclosure. Power Requirements The HP 8360 L-Series swept CW generators require a power source of 115 V (+lO/-257) or 230 V (+10/-15%/o), 48 to 66 Hz, single-phase. Power consumption is 400 VA maximum (30 VA in standby).
  • Page 373: Power Cable

    Figure 3-l shows the styles of plugs available on power cables supplied with Hewlett-Packard instruments. The HP part numbers indicated are part numbers for the complete power cable/plug set. The specific type of power cable/plug shipped with the instrument depends upon the country of shipment destination.
  • Page 374: Ac Power Cables Available

    CABLE CABLE CABLE H P P A R T PLUG DESCRIPTION * LENGTH F O R U S E I N C O U N T R Y P L U G M P E ’ COLOR ( i n c h e s ) 250V 8 1 2 0 - 1 3 5 1 Straight BSI 363A...
  • Page 375: Language Selection

    To set a programming language from the front panel, the instrument Note language on the rear panel HP-IB switch (Ll, L2, and L3 shown in Figure 3-2) must be set to 7 (all 1s). The HP-IB menu provides access to the swept CW generator’s programming language: 1.
  • Page 376: Hp-Ib Address Selection

    Because of this, the address menu Selection provides access not only to the swept CW generator’s HP-IB address, but also to the address at which the swept CW generator expects to see a power meter, and the address at which the swept CW generator expects to see a printer.
  • Page 377: Mating Connectors

    How to View or Change an HP-IB Address from the Front Panel Note To set an HP-IB address from the front panel, the instrument address on the rear panel HP-IB switch (Figure 3-2) must be set to 31 (all 1s). 1. Press SYSTEM (MENU).
  • Page 378: Operating Environment

    Temperature. The swept CW generator may be operated in Operating Environment environments with temperatures from 0 to +55 “C. Humidity. The swept CW generator may be operated in environments with humidity from 5 to 80% relative at +25 to 40 “C. However, protect the swept CW generator from temperature extremes, which can cause condensation within the instrument.
  • Page 379: Chassis Kits

    Slide Assemblies Screws (Inner Slide Assembly) Screws (Outer Slide Assembly) Nuts (Outer Slide Assembly) ilide Adapter Kit includes the following parts) (NON-HP, Adapter Brackets Adapter Bar Screws (Bracket to Bar) Nuts (Bracket to Slide Assembly) Ventilation Requirements: When installing the instrument in a...
  • Page 380: Installation Procedure

    Installation Procedure 1. Refer to Figure 3-3. Remove handle trim strips. 2. Remove four screws per side. 3. Using the screws provided, attach the rack mount flanges to the outside of the handles. 4. Remove the side straps and end caps. 5.
  • Page 381: Chassis Slide Kit

    6. Refer to Figure 3-4. Remove the inner slide assemblies from the outer slide assemblies. 7. To secure the side covers in place, mount the inner slide assemblies to the instrument with the screws provided. 8. With the appropriate hardware, install the outer slide assemblies to the system enclosure.
  • Page 382: Rack Flange Kit For Swept Cw Generators With Handles Removed Contents

    Rack Flange Kit for Option 908 swept CW generators are supplied with rack flanges and the necessary hardware to install them on the swept CW generator Swept CW Generators after removing the instrument handles. The following table itemizes with Handles Removed the parts in this kit.
  • Page 383: Installation Procedure

    Installation Procedure 1. Refer to Figure 3-5. Remove handle trim strips. 2. Remove the four screws on each side that attach the handles to the instrument; remove the handles. 3. Using the screws provided, attach the rack mount flanges to the swept CW generator.
  • Page 384: Rack Flange Kit For Swept Cw Generators With Handles Attached Contents

    Rack Flange Kit for Option 913 swept CW generators are supplied with rack flanges and the necessary hardware to install them on the swept CW generator Swept CW Generators without removing the instrument handles. The following table with Handles Attached itemizes the parts in this kit.
  • Page 385: Installation Procedure

    Installation Procedure 1. Refer to Figure 3-6. Remove handle trim strips. 2. Remove the four screws on each side that attach the handles to the instrument. 3. Using the longer screws provided, attach the rack mount flanges to the outside of the handles. 4.
  • Page 386: Storage And Shipment

    The swept CW generator may be stored or shipped within the Environment following limits: Temperature -40” to f75 “C. Humidity 5% to 95% relative at 0” to +40 “C. Altitude Up to 15240 meters. Pressure approximately 50,000 feet. The swept CW generator should be protected from sudden temperature fluctuations that can cause condensation.
  • Page 387: Package The Swept Cw Generator For Shipment

    Use the following steps to package the swept CW generator for Package the Swept CW shipment to Hewlett-Packard for service: Generator for Shipment 1. Fill in a service tag (available at the end of Chapter 4) and attach it to the instrument. Please be as specific as possible about the nature of the problem.
  • Page 388: Converting Hp 8340/41 Systems To Hp 8360 L-Series

    The following paragraphs are intended to assist you in converting existing HP 8340/8341-based systems to HP 8360 L-Series swept CW generator-based systems. The HP 8360 L-series swept CW generator may be used where no modulation requirements are needed. Both manual and remote operational differences are addressed.
  • Page 389: Manual Operation

    Manual Operation Compatibility The HP 8360 L-Series swept CW generators are designed to be, in all but very few cases, a complete feature superset of the HP 8340/8341 synthesized sweepers. The most notable omissions are that the HP 8360 L-Series does not accept: line triggers (ie.
  • Page 390: System Connections

    HP 8510 network analyzer with firmware revision 4.0 or higher. To upgrade firmware for an existing HP 8510, an HP 11575C Revision 4.0 Upgrade Kit or an HP 11575D Revision 5.0 Upgrade Kit is required. prior to 6.0 (not inclusive)
  • Page 391: The Hp 8757C/E Scalar Network Analyzer

    HP 8340/8341. The HP 8360 L-Series differs from the HP 8340/8341 in one connection only. It is unnecessary to connect the modulator drive signal from the analyzer to the source. The HP 8360 L-Series internally produces the 27.8 kHz modulated signal necessary for AC mode measurements on the analyzer.
  • Page 392: Remote Operation

    Control Interface Intermediate Language CIIL is the instrument control programming language used in Option 700 HP 8360 L-Series. Like the HP 8340/8341 E69, the Option 700 HP 8360 L-Series is M.A.T.E.-compatible. Refer to the HP 8360 Option 700 Manual Supplement for information on this option.
  • Page 393: Numeric Suffixes

    Numeric suffixes consist of 2 or S-character codes that terminate and scale an associated value. The numeric suffixes for network analyzer language on the HP 8360 L-Series and the HP 8340/8341 are identical. Table 3-8 lists the HP 8360 L-Series suffixes. The default unit for each type of suffix is shown in bold type.
  • Page 394: Programming Language Comparison

    Table 3-9. Programming Language Comparison Description Network Analyzer SCPI Language Language Leveling mode, external POW:ALC:SOUR DIOD; : P O W : A T T : A U T O O F F Leveling mode, internal POW:ALC INT Leveling mode, mm module POW:ALC:SOUR MMH;...
  • Page 395 Table 3-9. Programming Language Comparison (continued) Description Network Analyzer SCPI Language Language HP-IB only functions Output status byte *STB? (See SCPI common commands) Status byte mask R M <num> (See SCPI common commands) Extended status byte mask RE <num> *ESE...
  • Page 396 *ESE <num>, *ESE? Reset sweep A B O R Number of steps in SN <num> SWE:POIN <num> Swap network analyzer Test HP-IB interface TI <num> DIAG:TINT? <num> T L <num>time-suffix SWE:TIME:LLIM <num>[time-suffix] T S W ; * W A I SYST:PRES...
  • Page 397 Table 3-9. Programming Language Comparison (continued) Description Network Analyzer SCPI Language Language Modulation SHPM PULS:SOUR SCAL;STAT ON Power jet power level PL <num>DB P O W <num>[DBM] POW:MODE SWE POW:MODE FIX POW:STAT ON POW:STAT OFF SHPS P O W : A T T : A U T O O F F P O W : A T T : A U T O O N P O W : A T T <num>[DB] POW:STEP <num>[DB]I[freqsufiix]...
  • Page 398 (See SCPI common commands) Select an internal (hardware) ROSC INT frequency reference Select an external [hardware) ROSC EXT frequency reference Display/set HP-IB address SYST:COMM:GPIB:ADR (or hardware switch) Select SCPI Select network analyzer language Select CIIL 3YST:KEY:DIS SAVE Lock save/recall registers Unlock save/recall registers...
  • Page 399 CW generator functions prior to use. For delete front panel options of the HP 8360 L-Series, use the “Front Panel Emulator Software” to perform an operator’s check. If the swept CW generator requires service and the routine...
  • Page 400: Description

    STANDBY LED is off, and that the green POWER ON LED is a. Check the display, a cursor will appear in the upper left corner followed by the HP-IB language, HP-IB address, and the date code of the firmware installed in the swept CW generator.
  • Page 401: Main Check

    1. Press (SERVICE). Main Check 3. Press fjm). If the display indicates a user preset was performed, select Factory Preset . Verify that the green SWEEP LED is blinking, the amber RF ON/OFF LED is on, and the red INSTR CHECK LED is off. 4.
  • Page 402: Replacing The Line Fuse

    Routine maintenance consists of replacing a defective line fuse, cleaning the air filter, cleaning the cabinet, and cleaning the display. These items are discussed in the following paragraphs. Table 4-1. Fuse Part Numbers For continued protection against fire hazard, replace line fuse only with same type and rating.
  • Page 403: How To Clean The Fan Filter

    The cooling fan located on the rear panel has a thin foam filter. How to Clean the Fan How often the filter must be cleaned depends on the environment Filter in which the swept CW generator operates. As the filter collects dust, the fan speed increases to maintain airflow (as the fan speed increases, so does the fan noise).
  • Page 404: How To Clean The Cabinet

    How to Clean the Cabinet WARNING To prevent electrical shock, disconnect the HP 8360 L-series swept CW generator from the mains before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
  • Page 405 Instrument History This chapter is left blank until this manual requires changes. Instrument History 5-l...
  • Page 406 L-5 active entry area, l-4 active entry area on/off, E-4 active entry arrow, l-4 adapters, 3-l HP-IB, C-6 adapter, three-prong to two-prong, 3-4 ADC fail, F-2 address swept CW generator, A-l, E-l address changes, no front panel, 3-8...
  • Page 407 select auto, A-9 select high, A-10 select low, A-10 ALC bandwidth selection, l-50 ALC disabled theory of, A-8 ALC leveling internal, L-3 mm-wave module, L-4 normal, L-l power meter, L-4 search, L-2 ALC menu, A-2-4 ALC off, L-l ALC off mode, l-32 ALC open loop, L-l ALC search mode, l-32 align output filter, A-22, P-l...
  • Page 408 B U S trigger source defined, 1-119 cabinet, clean, 4-5 cables HP-IB, C-6 calibrate sweep span always, S-66 calibrate sweep span once, S-67 calibration full user, F-14 sweep span, F-10 calibration failed message, 2a-2...
  • Page 409 SCPI, 3-24 compensation negative diode detectors, l-47 condition register, l-106 connections to HP 83550 series mm-wave modules, 3-22 to HP 8510 network analyzer, 3-21 to HP 8757C/E scalar analyzer, 3-21 to HP 8970B noise figure meter, 3-22 connector...
  • Page 410 correction value enter, E-l correction value entry, F-11 in general programming model, 1-112 coupled attenuator, A-6 coupled frequency, C-12 coupled stepped sweep to sweep time, D-8 coupling factor, C- 11 current path defined, l-68 rules for setting, l-68 custom menus, A-17 CW frequency, C-12 CW frequency, step size, U-2 CW operation, l-6...
  • Page 411 disable interface address changes, 3-8 disable save, S-l disable user flatness array, C-11 discrete parameters discussed in detail, l-85 explained briefly, l-75 discrete response data discussed in detail, l-86 display, l-4 display blank, B-l display, clean, 4-6 display clear, B-l display status, D-5 display status of phase-lock-loops, U-l display zero frequency, Z-l...
  • Page 412 HP-IB check, l-90 local lockout, 1-91 looping and synchronization, l-99 setting up a sweep, l-93 synchronous sweep, l-101 use of queries, l-95...
  • Page 413 C-3 copy frequency list, C-10 frequency increment, A- 18 HP 437B measure at all frequencies, M-7 HP 437B measure at one frequency, M-7, M-8 HP 437B measure functions, M-13 number of points, A-19 start frequency, A-20 stop frequency, A-21...
  • Page 414 step sweep activate, S-65 stop frequency, A-21 trigger external, L-9 trigger functions, P-14 trigger interface bus, L-8 trigger point automatic, L-8 frequency list copy, C-10 frequency list functions, L-5 frequency list, number of points, L-6 frequency markers, l-14 frequency menu, F-l 1 frequency multiplier, F-12 frequency offset, F- 13 frequency softkeys, F-11...
  • Page 415 HP 437B, measure correction, M-7, M-8 HP 437B measure correction functions, M-13 HP 8340 status register, A-13 HP 8340/41 system convert to HP 8360 system, 3-19 HP 8355~series interface connector, C-8 HP 83550 series system connections, 3-22 HP 8360 as controller, 3-7...
  • Page 416 HP-IB address identify, S-52 HP-IB address menu, A-l HP-IB check, example program, l-90 HP-IB connecting cables, l-56 HP-IB connector, C-6 HP-IB connector mnemonics, C-8 HP-IB control functions, H-l HP-IB, definition of, 1-55 HP-IB syntax error message, 2a-3 HP-IB trigger stepped sweep mode, S-62...
  • Page 417 instrument state, save, S-l instrument state save command, S-14 integer response data discussed in detail, l-86 integers rounding, l-84 interface address change, 3-8 factory-set, 3-7 power meter, M-8 printer, P-11 view, 3-8 interface bus trigger, frequency list, L-8 interface bus connector, C-6 interface bus softkeys, H-l interface bus trigger, stepped sweep, S-62 interface bus trigger, sweep mode, S-55...
  • Page 418 normal, L- 1 leveling mode ALC off, A-8, L-l normal, L-l search, A-8, L-2 leveling mode normal, A-5 leveling modes, A-5 leveling point external detector, A-7, L-3 internal, L-3 module, L-4 power meter, A-7, L-4 source module, A-7 leveling points, A-5 line fuse, replacement, 4-4 line switch, L-5 line voltage selection, 3-3...
  • Page 419 3-22 mnemonics, l-63, l-64 conventions for query commands, l-63 long form, l-64 short form, l-64 modify HP 8340/41 program for SCPI, 3-23 MOD key, M-9 modulation pulse, scalar, P-15 module selection, M-10, M-11, M-12 module selection softkeys, M-9...
  • Page 420 HP BASIC statements, 1-81 OUTPUT new line[new line] use as a program message terminator, 1-81...
  • Page 421 P-2 power level step size, U-l power menu functions, P-5 power menu key, P-5 power meter HP 437B, l-34, l-47 power meter leveling, L-4 power meter measure correction functions, M-13 power meter programming address, M-8 power meter range, P-16...
  • Page 422 S-51 power sweep, sweep time, S-67 power switch, L-5 precise talking, l-66, l-83 prefix number, vii preset conditions, HP 8340/41 compared to HP 8360, 3-20 preset key, l-3, P-8 preset mode factory, P-10 user, P-10 preset, save user defined, S-2...
  • Page 423 pwron fail, F-3 defined, l-63 discussed, l-66 queries, example program, l-95 query commands, l-72 query only, l-72 query only, l-72 query status byte, S-14 rack flange kit contents, 3-13 rack flange kit installation, 3-14 rack flange kit, no handles, 3-13 rack flange kit, with handles, 3-15 rack mount slide installation, 3-11 rack mount slide kit contents, 3-10...
  • Page 424 4-l service keys, 2-l service request enable register, S-14 service tags>, 4-6 set attenuator, S-51 setting HP-IB addresses, A-l shipment, 3-17 shipping damage, 3-l single, 1-12 single frequency, C-12 single sweep, 1-12, S-51 slave, step control, S-57...
  • Page 425 in general programming model, l-112 trigger command defined, 1-119 source module interface, L-4 source module interface connector, C-8, M-10, M-11, M-12 source module interface mnemonics, C-9 source module leveling, L-4 source module selection, M-9, M-10, M-11, M-12 space proper use of, l-69 span fail, F-2 span, frequency, S-52 span key, S-52...
  • Page 426 general model, l-106 transition filter, I-107 status register structure, SCPI, S-47 status system overview, l-106 * S T B ? , S - 1 4 step attenuator, A-6 step control master, S-56 step control slave, S-57 step dwell, S-59 stepped frequency mode, dwell time, S-59 stepped mode, number of points, S-59 stepped sweep coupled, D-8 stepped sweep mode, S-66...
  • Page 427 stepped functions, S-60 sweep mode ramp, S-65 sweep modes, l-12 sweep mode step, S-66 sweep mode stepped frequency list, S-65 sweep once, S-51 sweep output connector, C-4 sweep span calibrate always, S-66 sweep span calibrate once, S-67 sweep span calibration, F-10 sweep time, l-10 sweep time coupled to stepped sweep, D-8 sweep time key, S-67...
  • Page 428 S-55 trigger commands defined, l-l 18 trigger functions list mode, P-14 trigger, group execute command, S-15 TRIGGER (HP BASIC), 1-119 trigger input BNC, S-63 trigger input connector, C-5 trigger, interface bus stepped sweep, S-62 trigger out delay, T-3...
  • Page 429: Zoom

    F-4 HP 437B measure, M-7, M-8 power meter measure, M-13 user flatness correction commands, example program, l-103 user preset, P-10 user preset, save, S-2 vector network analyzer connections, 3-21 view interface address, 3-8 view previous menu, P-11...

This manual is also suitable for:

Hp 83630lHp 83623lHp 83650lHp 83640l

Table of Contents