System Connections; The Hp 8510 Network Analyzer - HP 8360 L Series User Manual

Swept cw generator (including options 001, 004, and 008)
Table of Contents

Advertisement

System Connections

Note

The HP 8510 Network Analyzer

The HP 8360 L-Series swept CW generator is compatible with any
HP 8510 network analyzer with firmware revision 4.0 or higher.
To upgrade firmware for an existing HP 8510, an HP 11575C
Revision 4.0 Upgrade Kit or an HP 11575D Revision 5.0 Upgrade Kit
is required.
prior to 6.0 (not inclusive)
HP 8510 revisions
the following connections:
STOP SWEEP IN/OUT
n
HP-IB INTERFACE
n
HP 8510 revisions 6.0 and greater use the connections as designated
on the rear panel of the swept CW generator. They are:
TRIGGER OUTPUT
n
STOP SWEEP IN/OUT
n
HP-IB INTERFACE
n
AUXILIARY INTERFACE
n
The dedicated HP 8510 versions of the HP 8360 may be configured
to power up to one of two possible system languages, network
analyzer language, or SCPI (Standard Commands for Programmable
Instruments). This configuration is controlled via a switch located
on the rear panel of the instrument. The factory default setting for
this switch is network analyzer language at an HP-IB address of 19.
To interface with a network analyzer the language selected must be
Analyzer language. Refer to earlier paragraphs in this chapter for the
rear panel switch settings.
Models other than the dedicated HP 8510 versions are set at the
factory for SCPI. To interface with a network analyzer the language
selected must be Analyzer language.
require that you use
Installation 3-2 1