S-26 Operating and Programming Reference
pairs. This correction information is used to create a correction
array that can be added to the internal calibration array. The
correction entered is at the associated frequency. Frequencies in
between frequency-correction pair values are determined by linear
interpolation. If a value of START or STOP frequency is specified
that is outside the limits of the specified frequencies, the correction
applied at those points is 0 dB. After "RST, returns a
l
l
Sets and queries the source of correction.
CORRection:FLATness:POINts? CMAXimumIMINimuml
l
Returns the number of frequency-correction pairs entered using the
CORR:FLAT command. After *RST, the value is 2.
l
l
Sets and queries the switch on the users ALC correction system.
The *RST value is OFF.
DIAGnostics:ABUS? <num>
l
Reads the analog bus node number and returns the number of
millivolts.
DIAGnostics:ABUS:AVERage <num>
l
DIAGnostics:ABUS:AVERage?
l
Sets and queries the number of ADC averages to use during the read
ADC query. After *RST, the value is 1.
DIAGnostics:ABUS:STATus?
l
Queries the status of the prior ADC reading. The response is a single
byte that is bit-encoded to mean:
Bit 0
Set to 1, if ADC timed out (hardware fault).
Bit 1
Set to 1, if reading was unsettled.
Bit 2
Set to 1, if out of range occurred.
DIAGnostics:INSTrument:PMETer:ADDRess <num>
l
DIAGnostics:INSTrument:PMETer:ADDRess?
l
Sets and queries the HP-IB address to use for the power meter during
swept CW generator calibration routines. Allowable values are 0
through 31. *RST or power on does not effect this value. Default is
13. It is defaulted only when memory is initialized.
l