S-30 Operating and Programming Reference
requested value. The start/stop frequencies are updated to reflect the
changes based on the coupling equations.
The swept CW generator uses "bumping" to move unspecified
frequency parameters, but if the final value of any of the frequency
headers is the result of bumping, then an error is generated since
not
the user is
getting what was specified. This means, to guarantee
sequence independence requires sending the frequency pairs in a
single message.
Example 1: (present state start = 5 GHz, stop = 6 GHz)
Example 2: (present state start = 5 GHz, stop = 6 GHz)
Example 3: (present state start = 5 GHz stop = 6 GHz)
l
l
Sets and queries the center frequency.
The *RST value is (MAX + MIN)/2.
l
l
l
Sets and queries the CW frequency. This does not change the
See FREQ:CENTER for more information.
l
l
l
l
Sets and queries the CW/ center frequency coupling switch. This
switch keeps the two functions coupled together when ON. Changing
one of them, changes both. *RST setting is
for more information.
an error results since the
stop frequency is bumped.
the final sweep does not
generate an error
(20
22).
no error is generated,
start frequency is unchanged.
still no error.
both are fine,
no errors.
I
OFF. See FREQ:CENTER