HP 8360 L Series User Manual page 155

Swept cw generator (including options 001, 004, and 008)
Table of Contents

Advertisement

A-8 Operating and Programming Reference
GENERRTOR
Figure A-2. Typical External Leveling Hookup
ALC Disabled
ALC
Off. In this configuration, the ALC is disabled, power is
not sensed at any point, and therefore the absolute power level is
uncalibrated (see Figure A-l). Direct and separate control of the
RF Components)
RF modulator (p/o
The swept CW generator's front panel indicates the attenuator
setting and a reference level. The reference level is an approximate
indication of the attenuation provided by the RF modulator.
Typically the RF amplifier that follows the modulator is saturated
for modulation levels near 0 dB. Therefore the actual change in the
RF output power will not track the indicated reference level until the
amplifier is out of saturation.
The ALC off mode is useful for applications that involve pulse
modulation with extremely narrow pulses. If the pulse is narrow
enough, the ALC may be unable to provide accurate leveling due to
bandwidth limitations.
Search.
Search mode is similar to the ALC off mode in that, the
ALC is disabled in order to remove bandwidth limitations. The
essential difference is that, when search mode is enabled, the swept
CW generator searches out the appropriate modulator level such that
the RF output power after the ALC is disabled closely matches the
power prior to search mode being enabled. Specifically, when search
mode is selected the swept CW generator follows this sequence of
steps:
1. All modulation is disabled and the ALC system is closed to
provide a calibrated reference power.
2. The output power is measured using the internal coupler/detector.
3. The ALC system is disabled (opened).
Q
RF OUTPUT
ALC
N E G A T I V E
DETECTOR
and the attenuator is possible.