Module Select AUTO
Function Group
Menu Map
Description
Programming Codes
See Also
Module Select Front
Function Group
Menu Map
Description
POWER and FREQUENCY
2 and 5
This command sets the automatic selection of the millimeter source
module interface connector. The swept CW generator looks at both
front and rear connectors and determines the type of source module
(if any) connected. If a source module is present at both connectors,
the swept CW generator selects the front connector as the active
one. After selecting the interface the instrument frequency limits
and multiplier are altered accordingly. However, the leveling point
not
is
changed. See Leveling Point
generator to level at the output of the source module.
An asterisk next to the key label indicates this feature is active. This
feature is the default after preset.
SYSTem:MMHead:SELect:AUTO ONlOFFlllO
SYSTem:MMHead:SELect:AUTO?
Analyzer:
NONE
Menu
Module
POWER and FREQUENCY
2 and 5
This command causes the swept CW generator to examine only
the front panel source module interface connector to determine
the type of source module (if any) connected. The instrument
frequency limits and multiplier are altered according to the source
module connected. However, the leveling point is
Module
Leveling Point
the output of the source module.
An asterisk next to the key label indicates this feature is active.
Module to set the swept
to set the swept CW generator to level at
Operating and Programming Reference M-l 1
CW
not
changed. See