Advanced Trigger Configurations; Trigger Keyword Definitions; Abort; Immediate - HP 8360 L Series User Manual

Swept cw generator (including options 001, 004, and 008)
Table of Contents

Advertisement

Trigger Keyword
Definitions
to jump to the next point when the signal becomes TRUE, therefore
the first point in the list or stepped sweeps is produced immediately
upon starting the sweep. Receiving a trigger signal at the last point
causes the IDLE state to be re-entered. Analog sweeps do not use
the trigger signals during the sweep (although the trigger signals are
needed to start the sweep as described).
The

ABORt

command resets any sweep in progress and immediately
returns the instrument to the IDLE state.
The
*WAI,
*OPC and *OPC? commands indicate a complete operation
at the end of the sweep upon re-entry into the IDLE state.

Advanced Trigger Configurations

Because the SCPI layered trigger model is expandable, many more
complex trigger configurations are possible.
The following paragraphs contain condensed definitions of the
keywords used in the command tables. Many of the commands in
trigger related subsystems are
event commands cannot be queried. Similarly, event commands
have no related *RST actions or settings. Event commands cause a
particular action to take place inside the swept CW generator.
ABORt
The ABORt command forces the trigger system to the idle state. Any
measurement or output sequence in process is aborted as quickly as
ABORt
possible.
does not alter the settings programmed by other
commands, unlike *RST.
cannot be queried.

IMMediate

The IMMediate command provides a one-time override of the normal
downward path in an event-detection state. The instrument must be
in the specified event detection state when IMMediate is received, or
an error is generated and the command has no effect. For example,
the instrument must be in the TRIG state for :TRIGger: IMMediate
to work properly. If the instrument is in the idle state, the command
has no effect, and an error would be generated. IMMediate is an
event command and cannot be queried.
The ODELay command specifies the time between the source
settling and the time the trigger out signal is sent. Specifying
CW generator to set the specified time as the delay necessary to
ensure proper settling. Sending *RST sets ODELay to an instrument
dependent value, usually zero.
event commands.
ABORt
is a root level event command and
Remember that

Hide quick links: