HP 8360 L Series User Manual page 227

Swept cw generator (including options 001, 004, and 008)
Table of Contents

Advertisement

Manual Sweep
Programming Codes
See Also
M-2 Operating and Programming Reference
resolution defined by the a and @ keys. Frequencies in the
manual sweep mode are synthesized, just as they are in
There are two major differences between manual sweep and a sweep
generated by activating the CW function and rotating the rotary
knob or pressing the ARROW keys.
1. The sweep output voltage ramp is 0 to +lO V in both modes,
but in CW mode, 0 V corresponds to lowest frequency of the
swept CW generator frequency range and +lO V corresponds
to the highest frequency of the range. In manual sweep mode,
0 V corresponds to the start frequency
corresponds to the stop frequency
sweep voltage at intermediate frequencies is a linear interpolation
of the frequency span. For example, a frequency half-way between
the start/stop limits has a sweep voltage of 5 V.
2. The bandcross points in CW mode occur at 2.0, 7, 13.5, 20, 25.5,
and 32 GHz.
In manual sweep mode the bandcrossing points have 200 MHz of
flexibility, that is automatically used by the swept CW generator
for optimum performance. For example, a 2.0 to 7.1 GHz sweep
could be accomplished without any band changes in manual sweep
mode.
SCPI:
SWEep:MODE MANuallAUTO
This is the command for frequency manual sweep.
POWer:MODE SWEep
POWer:SPAN <num>[lvl suffix]~MAXimum~MINimum
This is the command for power manual sweep.
LIST:MODE MANual
This is the command for manual list sweep.
Analyzer: S3
Power Sweep,
Sweep
"Continuous, Single, and Manual Sweep Operation" in Chapter 1
specified
specified.
Mode List
CW
mode.
and +lO V
In both cases, the