Function Group
M e n u M a p 8
Description
Programming Codes
See
Enter Corr
Function Group
M e n u M a p 5
Description
SYSTEM
This softkey lets you change the HP-IB address of the swept CW
generator. Enter the address desired using the numeric entry keys
or the up/down arrow keys. The address value may be set between
0 and 30. The swept CW generator stores the address value in
non-volatile memory. The default address of the swept CW generator
is 19.
SCPI:
SYSTem:COMMunicate:GPIB:ADDRess
Analyzer:
NONE
Also
Connectors,
HP-IB
"Instrument Addresses" in Chapter 1
"Programming Typical Measurements" in Chapter 1
POWER
This softkey lets you enter a power correction value for a frequency
point in the flatness array. A frequency point must be entered before
a correction value can be accepted, otherwise the following error
message appears:
ERROR - Must first enter correction freq.
The up/down arrow keys let you scroll through the frequency points
available for power correction. If no correction value is entered
for a frequency point, the swept CW generator display indicates
Undefined. The range of acceptable values is -40 to +40 dB. An
asterisk next to the key label indicates that this feature is active.
Menu
Operating and Programming Reference E-l