Fault Info 1 - HP 8360 L Series User Manual

Swept cw generator (including options 001, 004, and 008)
Table of Contents

Advertisement

F a u l t Menu
See Also

Fault Info 1

Function Group
M e n u M a p 6
Description
F-2 Operating and Programming Reference
Analyzer:
NONE
SERVICE
This softkey displays the latched status of the following fault
messages.
PEAK FAIL
Indicates that the peak algorithm is unable to align
the YTM passband to the frequency of the YO.
This fault indication is possible only if a peaking or
autotrack routine has been initiated.
TRACK FAIL Indicates that the autotrack algorithm is unable to
calculate the calibration constants needed to track
the YTM passband to the frequency of the YO.
This fault indication is possible only if an autotrack
routine has been initiated.
RAMP FAIL
Indicates that the ramp algorithm is unable to adjust
the sweep ramp voltage to 10.00 V at the end of
the sweep. Initiate a full self-test to gather more
information if this fault is indicated.
Indicates that the span algorithm is unable to adjust
SPAN FAIL
the YO to achieve the correct frequency at the end
of a band. This fault indication is possible only if a
sweep span routine has been initiated.
Indicates that the internal YO V/GHz line adjusted
at power-on or at preset is unable to calibrate.
Initiate a full self-test to gather more information if
this fault is indicated.
Indicates that the ADC ( analog-to-digital converter)
ADC FAIL
is not responding to a measurement request within
the time-out period. The ADC is used extensively in
the operations of the swept CW generator. Initiate a
full self-test to gather more information if this fault
is indicated.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 83630lHp 83623lHp 83650lHp 83640l

Table of Contents