Auto Fill Incr

Function Group
Menu Map
Description
Programming Codes
See Also
A-16 Operating and Programming Reference
FREQUENCY, POWER
This softkey is used in two locations:
List Menu.
Flatness Menu - When selected, the swept CW generator waits for
a frequency increment value to be entered. --> Increment :
displayed in the active entry area. A list of frequencies is created
automatically, beginning at the auto fill start frequency and always
ending with the auto fill stop frequency. The swept CW generator
uses the increment value on all points, but if the stop frequency
requires a different increment to be used to be exact, the swept
CW generator simply ends the frequency list at the stop frequency
disregarding the increment value.
If the
increment value requested creates a list that exceeds the
number of elements available, the following message appears:
TOO MANY CORRECTION PTS REQUESTED
List Menu - When selected, the swept CW generator waits for a
frequency increment value to be entered. --> Increment:
displayed in the active entry area. A list of frequencies is created
automatically, with all points separated by the frequency increment
value. The list begins at the auto fill start frequency and ends at a
frequency less than or equal to the auto fill stop frequency.
If the increment value requested creates a list that exceeds the
number of points available (801), the following message appears:
TOO MANY LIST PTS REQUESTED
SCPI:
NONE,see Fitness Menu or List Menu
Analyzer:
NONE
Fltness Menu, List Menu
"Optimizing Swept CW Generator Performance" in Chapter 1
Fltness
Menu and
is
is