Program Comments - HP 8360 L Series User Manual

Swept cw generator (including options 001, 004, and 008)
Table of Contents

Advertisement

Saving and Recalling
States, Example
Program 5
When a typical sweep, like example program 3, is set, up, the
complete front panel state may be saved for later use in non-volatile
memories called registers 1 through 8. This can be done remotely as
a part of a program. Clear and reset the controller and type in the
following program:
10
Source=719
20
ABORT 7
30
LOCAL 7
40
CLEAR Source
50
REMOTE Source
60
70
OUTPUT Source;"*RST;FREQ:MODE SWE;STAR 4GHZ
80
OUTPUT Source;"*SAV 1"
90
100
PRINT "A sweeping state has been saved in REGISTER 1."
110
OUTPUT Source;"*RST;FREQ:CW 1.23456GHZ;:POW:LEV -1DBM"
120
OUTPUT Source;"*SAV 2"
130
PRINT "A CW state has been saved in REGISTER 2."
140
PRINT II..... Press Continue"
150
PAUSE
160
OUTPUT Source;"*RCL 1"
170
PRINT "Register 1 recalled. Verify source is sweeping."
180
PRINT "Press Continue."
190
PAUSE
200
OUTPUT Source;"*RCL 2"
210
PRINT "Register 2 recalled."
220
PRINT "Verify source is in CW mode."
230
END
Run the program.

Program Comments

10: Assign the source's HP-IB address to a variable.
20 to 50: Abort any HP-IB activity and initialize the HP-IB
interface.
60: Clear the computer's display.
70: Set up the source for a sweeping state. Note the combination
of several commands into a single message. This single line is
equivalent to the following lines:
OUTPUT Source;"*RST"
OUTPUT Source;"FREq:MODE SWEep"
OUTPUT Source;"FREq:STARt 4 GHZ"
OUTPUT Source;"FREQ:STOP 5 GHZ"
OUTPUT Source;"INIT:CONT ON"
80: Save this state into storage register 1.
Getting Started Programming l-97

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 83630lHp 83623lHp 83650lHp 83640l