Fltness Menu

Function Group
M e n u M a p 5
Description
POWER
This softkey reveals the softkeys in the flatness correction menu that
control user-defined leveling parameters.
A u t o F i l l Incr
Auto
Fill #Pts
Auto
Fill Start
Auto Fill Stop
Clear Point
Copy List
Delete Menu
Enter Corr
Enter Freq
into the flatness correction array.
Freq Follow
Mtr
correction menu.
Automatically creates a frequency list with all
points separated by the specified increment in a
given frequency range.
Automatically creates a frequency list
containing the specified number of points in a
given frequency range.
Sets the start frequency of the flatness
correction array that will load automatically
when either the number of points or the
increment size
is specified.
Sets the stop frequency of the flatness array
that will load automatically when either the
number of points or the increment size is
specified.
Changes the power correction value for the
indicated frequency point to the undefined state.
Copies the frequency list, (see List Menu ),
into the frequency parameter of the flatness
correction array.
Disables the frequency-correction pair array and
uses the HP-IB transferred 1601 point correction
set to apply correction information.
Reveals the delete softkeys.
Enables the entry of a power correction value
for a frequency point.
Enables the entry of a single frequency point
Sets the swept CW generator to CW frequency
mode
so that the corresponding correction
values can
be entered.
Reveals the softkeys in the power meter measure
Operating and Programming Reference F-5
Fltness Menu