Step Swp Pt
Trig Ext
Function Group
Menu Map
Description
Programming Codes
See Also
Function Group
Menu Map
Description
Programming Codes
FREQUENCY
2
When this softkey is selected, the swept CW generator steps to
the next point in the stepped frequency sweep when an external
hardware trigger is received. When the last frequency point is
reached and continuous sweep is selected, the next trigger causes the
step sweep to return to the start frequency. Connect the trigger
signal to the TRIGGER INPUT BNC. An asterisk next to the key
label indicates that this feature is active.
SCPI:
SWEep:TRIGger:SOURce EXT
Analyzer:
TS
Step Sup Menu, Sweep Mode Step
"Using Step Sweep" in Chapter 1
FREQUENCY
NONE
This hardkey activates swept frequency mode and makes the stop
frequency parameter the active function. The start/stop frequency
must be separated by at least 2 Hz in order to remain in the
frequency sweep mode. If start=stop frequency then the zero span
mode is entered.
FREQuency:STOP <num>[freq suffix] or
FREQuency:MODE:SWEep
Analyzer:
FB <num> [Hz]Kz]Mz]Gz]
Operating and Programming Reference S-63
(STOP)