Enter List Freq

Function Group
Menu Map
Description
Programming Codes
See Also
This softkey lets you enter a dwell time for a frequency point in the
frequency list array. A frequency point must be entered before a
dwell value can be accepted, otherwise the following error message
appears:
ERROR: Must first enter a List Frequency.
The rotary knob and the up/down arrow keys let you scroll through
the frequency points available to change the default dwell values.
The range of values is 100 ps to 3.2 s. An asterisk next to the key
label indicates that this feature is active.
SCPI:
NONE, see List
List Menu
"Optimizing Swept CW Generator Performance" in Chapter 1
FREQUENCY
2
This softkey lets you enter a frequency point into the frequency list
array. The frequency list may contain as few as one and as many as
801 points. The order frequencies are entered is the order they are
listed. Additions to an existing list are placed as indicated by the
active entry arrow. The rotary knob and the up/down arrow keys let
you scroll through the frequencies points. An asterisk next to the key
label indicates that this feature is active.
SCPI:
NONE, see List Menu
List Menu
"Optimizing Swept CW Generator Performance" in Chapter 1
Menu
Operating and Programming Reference E-3
Enter List Freq