S-28 Operating and Programming Reference
HP 8360 L-Series Swept CW Generator Service Guide
troubleshooting procedure.
DIAGnostics:TEST:CONTinue
l
Causes the selftest execution to continue when paused for raw data
examination. Does nothing in other conditions.
DIAGnostics:TEST:DATA:DESC?
l
Returns the description string of the raw data examined during a
selftest. It returns " " in other conditions.
DIAGnostics:TEST:DATA:MAXimum?
l
Returns the upper test limit for the raw data point examined.
Returns "0" in other conditions.
DIAGnostics:TEST:DATA:MINimum?
l
Returns the lower test limit for the raw data point examined.
Returns "0" in other conditions.
DIAGnostics:TEST:DATA:VALue?
l
Returns the raw data value for the raw data point examined.
Returns a "0" in other conditions.
DIAGnostics:TEST:DISable (<num>)l*?lALL
l
Prevents the listed selftests from being selected. If ALL is sent then
all of the selftests are disabled. *RST causes DIAG :TEST:ENAB ALL to
execute.
DIAGnostics:TEST:ENABle {<num>)l*?lALL
l
Enables the listed selftests to execute. If ALL is sent then all of the
selftests are enabled. *RST causes DIAG : TEST : ENAB ALL to execute.
l
The specified selftest is executed. Normal instrument operation is
suspended and the instrument state is restored upon exiting the
DIAGnostics:TEST:LOG:SOURce ALLIFAIL
l
DIAGnostics:TEST:LOG:SOURce?
l
Sets and queries the source for the raw data logging. ALL specifies
that all raw data points are displayed. FAIL selects only those data
points out of the test limits. Both commands are executable in