HP 8360 L Series User Manual page 249

Swept cw generator (including options 001, 004, and 008)
Table of Contents

Advertisement

Preset Mode
Factory
Function Group
Menu Map
Description
Programming Codes
See Also
P-10 Operating and Programming Reference
SYSTEM
8
This softkey sets the standard starting configuration of the swept
CW generator when the (PRESET) key is pressed, as set by the
manufacturer. An asterisk next to the key label indicates that this
feature is active. The following is a description of the configuration.
Stop sweep at the maximum specified frequency.
n
Power level set at 0 dBm.
n
Sweep time set to auto.
n
CONT sweep.
n
ALC leveling point internal.
n
ALC leveling mode normal.
n
Markers set to activate at the center frequency of the sweep.
n
All function values stored in memory registers 1 through 9 remain
n
in their previous states.
The checksum of the calibration data is calculated, and if an error
n
is detected, the calibration data in protected memory is used. If
the checksum of the protected data is not correct, then default
values are used an error message (EEROM FAILED, LOST CAL)
is displayed.
SCPI:
SYSTem:PRESet:TYPE FACTory
Analyzer:
IP, which is the same as (ZG%?).
P R E S E T
Preset Mode User
"Changing the Preset Parameters" in Chapter 1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 83630lHp 83623lHp 83650lHp 83640l

Table of Contents