Power Slope Operation

Power Sweep
1. Press POWER (jj).
2. Select Power Sweep.
3. Enter a value.
4. Press terminator key.
Select power Sweep (asterisk on).
Press @GE).
The swept CW generator performs a power sweep beginning at
-20 dBm and ending at +5 dBm. The power meter indicates
This function allows for compensation of high frequency system or
cable losses by linearly increasing the power output as the frequency
increases. For this example, refer to the "Menu Map" section.
Press Power Slope, the active entry area displays:
Power slope is now active, notice that an asterisk is next to the key
label.
Use the entry keys, rotary knob, or arrow keys to enter a value for
the linear slope.
Press Power Slope
SUEPT CU
R F O U T P U T
Figure l-10. Power Sweep and Power Slope Operation
where X is a numeric value.
again to turn this feature off.
POUER HETER
POWER SENSOR
Power Slope
1. Press POWER (&ii@.
2. Select Power Slope.
3. Enter a value.