Leveling Pointpwrmtr; Line Switch - HP 8360 L Series User Manual

Swept cw generator (including options 001, 004, and 008)
Table of Contents

Advertisement

Leveling Point
Function Group
Menu Map
Description
Programming Codes
See Also
Function Group
Menu Map
Description
Programming Codes
See Also
ALC
1
This softkey lets you set the swept CW generator to level at the
power sensor of an external power meter. This mode of operation
requires a feedback connection from the power meter to the EXT
ALC BNC located on the swept CW generator. An asterisk next to
the key label indicates that this feature is active.
SCPI: POWer:ALC[:SOURce] PMETer
Analyzer: A3
"Externally Leveling the Swept CW Generator" in Chapter
NONE
NONE
The line switch (on/off switch) has two positions, off or standby and
on. If line power is connected to the swept CW generator and the
line switch is set to off, the swept CW generator is in the standby
state (amber LED on). Standby provides power to the internal
frequency standard oven.
When line power is connected and the line switch is set to on, the
swept CW generator power supplies are enabled and a limited
self-test is initiated. The CPU self test is performed; power supplies
and the front panel processor are checked.
NONE
"INSTALLATION" for information on fuses.
"Error Messages" for information on messages displayed at power on.
Operating and Programming Reference L-5

LINE SWITCH