Pt Trig Menu
Function Group
M e n u M a p 2
Description
Programming Codes
See Also
Pulse On/Off
Scalar
Function Group
Menu Map
Description
FREQUENCY
This softkey accesses the list mode point trigger softkeys.
List Mode Pt Trig Auto
List Mode Pt Trig Bus
List Mode Pt Trig Ext
SCPI:
NONE
Analyzer:
NONE
4
This softkey activates pulse modulation mode, and sets the internal
pulse generator to produce 27.778 kHz square wave pulses
(18 ps pulse width, 36 ps pulse period). The rise and fall times of
the RF envelope are approximately 2 ps. These pulses allow proper
operation with HP scalar network analyzers in ac detection mode.
An asterisk next to the key label indicates that this feature is active.
Automatically steps the swept CW
generator to the next point in the
frequency list.
Steps the swept CW generator to
the next point in the frequency list
when an HP-IB trigger is received.
Steps the swept CW generator to
the next point in the frequency list
when an external hardware trigger
is received.
Operating and Programming Reference P-15