Mating Connectors

10 MHz Frequency
Reference Selection
and Warmup Time
3-8 Installation

How to View or Change an HP-IB Address from the Front Panel

To set an HP-IB address from the front panel, the instrument
address on the rear panel HP-IB switch (Figure 3-2) must be set to
31 (all 1s).
1. Press SYSTEM (MENU).
3. The swept CW generator displays the three address softkeys:
8360 Adrs , Meter Adrs , and Printer Adrs .
4. Select the desired softkey.
5. The swept CW generator displays the address selected for that
instrument.
6. If you want to change the address, use the keypad to enter the
desired address (0 to 30), then press (ENTER).
If the swept CW generator displays Rear panel HP-IB address
must be 31 (11111) in order to change current address (=Xx>,
the address on the rear panel HP-IB switch (Figure 3-2) is set to
something other than 31 (all 1s).
How to Prevent a Front Panel Change to an HP-IB Address
To disable the address softkeys, set the instrument address on the
rear panel HP-IB switch (Figure 3-2) to any address other than 31
(all 1s).
How to Set the HP-IB Address on a Swept CW Generator without a
Front Panel
If your swept CW generator does not have a front panel, set the
address on the rear panel HP-IB switch (Figure 3-2) to the address
you want (factory default is 19).
All of the externally mounted connectors on the instrument
are discussed in the "CONNECTORS" section in Chapter 2.
If you are interested in the HP part number for a connector,
see "Replaceable Parts" in the
Generator/HP 8360 L-Series Swept CW Generator Service Guide.
To keep the internal timebase frequency reference oven at operating
temperature, the swept CW generator must be connected to ac line
power. The swept CW generator requires approximately 30 minutes
to warm up from a cold start before the OVEN display message
goes off. With a stable outside temperature, internal temperature
equilibrium is reached after approximately two hours. For additional
information on warmup times, see "Specifications," in Chapter 2c.