Remove/Replace Split/Splitless Capillary Inlet Solenoid Valve - HP 5890 SERIES II Service Manual

Gas chromatograph
Hide thumbs Also See for 5890 SERIES II:
Table of Contents

Advertisement

Remove/Replace Split/Splitless Capillary Inlet Solenoid Valve

HAZARDOUS VOLTAGES ARE PRESENT IN
`
THE INSTRUMENT WHEN THE POWER CORD
IS CONNECTED.
DANGEROUS
DISCONNECTING
BEFORE WORKING ON THE INSTRUMENT.
FLAME IONIZATION (FID) AND NITROGEN
`
PHOSPHOROUS (NPD) DETECTORS USE
HYDROGEN GAS AS FUEL.
HYDROGEN GAS IS TURNED OFF AT ITS
SOURCE BEFORE REPLACING ANY FLOW
COMPONENTS.
1. Set the main power line switch to the off position.
2. Disconnect the power cable from its receptacle.
3. Allow time for the oven and heated zones to cool.
4. When the heated zones are cool, turn off all gas
supplies.
5. Remove the two screws securing the left
side panel along its bottom edge.
6. Slide the left side panel towards the rear
of the instrument and lift.
7. Remove the four screws securing the rear
cover to the instrument.
8. Slide the rear cover towards the rear of the instru-
ment.
9. Remove the electronics carrier top cover
(above the signal cable plugs and receptacles
to expose the top edge of the
TCD detector PCB).
10. Remove the right side panel by
removing four screws: two each
along its top and bottom edges.
SCREWS
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
AVOID A POTENTIALLY
SHOCK
HAZARD
THE
POWER
CORD
BE SURE ALL
SCREWS
SVC 3- 7
BY
ELECTRONICS
CARRIER TOP
COVER
RED O VISIBLE
OFF
ON
SCREWS
SCREWS
RIGHT SIDE PANEL