HP 5890 SERIES II Service Manual page 286

Gas chromatograph
Hide thumbs Also See for 5890 SERIES II:
Table of Contents

Advertisement

The message " PASSED SELF TEST"indicates INET, at least with respect
to the HP 5890A, is performing satisfactorily. If FAILED SELF-TEST is
displayed, a bad cable may be indicated; install a different INET cable and
repeat the test. If FAILED SELF-TEST is displayed again for a second
cable, electronic problems within the HP 5890A are indicated; see Section
19, " Electronic Troubleshooting,"for more information.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
The LOOPBACK TEST may be used to check for continuity in an
INET cable; an open cable causes test failure. Verify that the cable
is at fault (rather than the HP 5890A) by testing a second cable.
If an intermittent cable problem is suspected, the test may be re-
peated as necessary while flexing the cable (particularly at its
plugs). An ohmmeter should also be used to test for problems; it is a
reliable method for testing continuity.