Replacing Assemblies
To remove and replace the handle
To remove and replace the handle
• Remove the two screws in the endcaps, then lift off the handle.
To remove and replace the feet and tilt stand
1 Remove the screws connecting the four rear feet to the instrument.
2 Separate the rear feet from the instrument to remove them.
3 Press the locking tab on the bottom feet, then remove them.
4 Remove the tilt stand from the bottom front feet by lifting the stand up and out of
the foot.
5 Reverse this procedure to install the feet and tilt stand.
To remove and replace the cover
1
Tum
off the power and unplug the logic analyzer.
2 Remove the probe plate and dIsconnect the logic analyzer cables from the rear panel.
3 Using the previous procedures, remove the handle and the four rear feet.
4 Remove the seven screws from the front moldIng, then slide the molding forward to
remove it.
5 Remove the cover.
To remove the cover, set the instrument upright and facing toward you. Slide the chassis
toward the front, out of the cover, and set it on a static-safe work area.
6 Reverse this procedure to install the cover.
Check that
all
assemblies are properly installed before installing the cover.
When instaJJing the chassis
in
the cover, check that the tabs located at the bottom, rear of the
cover
align
with the holes in the rear panel.