59. Install the collector mount and collector assem-
bly as a unit.
60. Secure the collector mount to the thermal strap
using three screws.
61. Connect the ignitor wire lead connector at the
mating connection adjacent to the ignitor.
62. If it was necessary to remove the rear
panel, install it and secure using four
screws.
63. Install the left side panel and secure using two
screws.
64. Install the right side panel and secure
using four screws.
65. Install the electronics carrier top cover.
66. Install the hinged top cover and secure using a screw and washer.
67. Connect the ground strap to the hinged top cover using a screw.
68. Restore all gas supplies.
69. Restore power to the instrument.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
COLLECTOR
NUT
COLLECTOR
SPANNER
NUT
COLLECTOR
MOUNT
THERMAL
STRAP
SVC 4- 27
DETECTOR PCB
INTERCONNECT
IGNITOR
CLAMPS