HP 5890 SERIES II Service Manual page 151

Gas chromatograph
Hide thumbs Also See for 5890 SERIES II:
Table of Contents

Advertisement

10. Remove the photomultiplier tube (PMT) assembly and sulphur filter from the detector assembly
and set it aside.
(CHIMNEY AND BRACKETRY
REMOVED FOR CLARITY)
11. Remove the drip tube from the exhaust tube at the top of the detector.
12. Remove the exhaust tube from the detector using a
9/16- inch wrench.
13. Remove the chimney assembly by removing
the two screws securing it to the chimney back.
14. Loosen the three screws which secure the
clamp which secure the detector to the chim-
ney back.
15. Use a 9/16 inch wrench to loosen the nut hold-
ing the weldment exit tube to the jet assembly.
It will be necessary to hold the jet assembly
with a 1/2- inch wrench to prevent rotation.
16. Pull the heater and sensor from the detector
weldment assembly.
17. Carefully lift the detector, vertically, from the
transfer tube, so as not to damage the fused
silica liner.
18. Inside the oven, remove the column from the
FPD detector base weldment.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
SULPHUR FILTER
PMT ASSEMBLY
CLAMP
SECURING
SCREWS
SVC 4- 81
CHIMNEY
SCREW