HP 5890 SERIES II Service Manual page 69

Gas chromatograph
Hide thumbs Also See for 5890 SERIES II:
Table of Contents

Advertisement

12. With a Pozidriv screwdriver, remove the mounting screw from the rear of the block. (The block
must be removed from the rear of the instrument.)
AUXILIARY
INLET FITTING
FID/FPD IGNITION
MICRO SWITCH
REAR CLAMPING PLATE
13. Install the new block on the flow panel from the rear of the instrument. The block fits on the
panel over two standoffs, in only one direction.
14. Replace the mounting screw on the block and tighten it firmly.
15. Connect the plumbing on the rear of the block. The fittings are labelled on the block. In addi-
tion:
On an FID/NPD, replace the hydrogen and air tubes. The hydrogen tube (painted red) con-
D
nects to the lower fitting. The gas tube retainer fits properly only when the hydrogen and
air tubes are installed correctly.
On an FID or FPD block, connect the ignitor leads. (The leads may be reversed without
D
consequence.)
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
PLUG VALVE
COMPRESSION
SPRING
FLOW MANIFOLD
BLOCK (TYPICAL)
RESTRICTOR
INLET FITTING
SVC 3- 26
FRONT CLAMPING
PLATE
FID/FPD IGNITION
VALVE
NEEDLE VALVE
ASSEMBLY
ON/OFF VALVE
WELDMENT
OUTLET FITTING PLATE